Beobachtungsstudie in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2023. Verfasst von Shunmuga Sundaram Ponnusamy und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Velammal Medical College, Madurai, India. Hauptthemen laut MEDLINE: Persistent Left Superior Vena Cava, Cardiac Resynchronization Therapy. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 8 Literaturangaben.
Abstract (Originalsprache)
Introduction
Left bundle branch pacing has gained significant momentum in the last few years. The procedure involves deploying the lead deep inside the interventricular septum through left subclavian vein. We aimed at analyzing the feasibility, efficacy and long‐term outcome of left bundle branch pacing (LBBP) using lumen‐less lead through the right subclavian vein.
Methods
This was a retrospective‐institutional, single center observational study done in consecutive patients who underwent LBBP using 3830 selectsecure tm lead. Left subclavian venous access was the primary strategy for lead implantation. Patients requiring right sided approach due to venous obstruction or persistent left superior‐vena‐cava (PLSVC) for LBBP were included in the study.
Results
Right sided approach was successful in 16 out of 19 (84%) attempted patients. C315‐His catheter was used in all patients without modifying its curvature. PLSVC ( n = 7), left venous obstruction ( n = 7), right sided device upgradation ( n = 1) and left pocket infection ( n = 1) were the reasons for right sided approach. Mean follow‐up duration was 17 ± 12 months. LBBP resulted in reduction in QRS duration from 137.3 ± 37.8 ms to 122.3 ± 9.5 ms ( p −.13) and increase in LV ejection fraction from 46.2 ± 16.3% to 54.4 ± 11.6% ( p −.11). The mean fluoroscopy duration and radiation dose were significantly high in right sided approach ( n = 16) as compared to left sided approach ( n = 293). In patients requiring cardiac‐resynchronization therapy (CRT), right sided LBBP resulted in reduction in QRS duration from 171.8 ± 18.5 to 125.5 ± 11.9 ms ( p −.0001) and increase in LVEF from 29.1 ± 3.8 to 45.1 ± 11.9% ( p −.005).
Conclusion
Right sided LBBP is feasible, safe and effective in patients requiring pacing for symptomatic bradyarrhythmia and CRT. Further development in dedicated tools for right‐sided approach would help in reducing the fluoroscopy‐duration and radiation‐dose.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Velammal Medical College, Madurai, India.
- Department of Microbiology, Velammal Medical College, Madurai, India.
- Geisinger Commonwealth School of Medicine, Geisinger Heart Institute, Wilkes Barre, Pennsylvania, USA.
ORCID-Profile: Shunmuga Sundaram Ponnusamy, Vadivelu Ramalingam, Pugazhendhi Vijayaraman
Zitieren
APA
Ponnusamy, S. S., Ganesan, V., Ramalingam, V., Kumar, M., Rupert, L. J., & Vijayaraman, P. (2023). Right sided approach for left bundle branch pacing using lumen‐less lead: Technical considerations and follow‐up outcome. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16126
Vancouver
Ponnusamy SS, Ganesan V, Ramalingam V, Kumar M, Rupert LJ, Vijayaraman P. Right sided approach for left bundle branch pacing using lumen‐less lead: Technical considerations and follow‐up outcome. Journal of Cardiovascular Electrophysiology. 2023. doi:10.1111/jce.16126
BibTeX
@article{Ponnusamy2023,
author = {Ponnusamy, Shunmuga Sundaram and Ganesan, Vithiya and Ramalingam, Vadivelu and Kumar, Mahesh and Rupert, Lydia Joyce and Vijayaraman, Pugazhendhi},
title = {Right sided approach for left bundle branch pacing using lumen‐less lead: Technical considerations and follow‐up outcome},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2023},
doi = {10.1111/jce.16126}
}
EUVIMED