Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2023. Verfasst von Rohan N Kumthekar und 4 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiology, Nationwide Children's Hospital, Columbus, Ohio, USA. Hauptthemen laut MEDLINE: Pacemaker, Artificial. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
Pacemaker implantation in infants and small children is limited to epicardial lead placement via open chest surgery. We propose a minimally invasive solution using a novel percutaneous access kit.
Objective
To evaluate the acute safety and feasibility of a novel percutaneous pericardial access tool kit to implant pacemaker leads on the epicardium under direct visualization.
Methods
A custom sheath with optical fiber lining the inside wall was built to provide intrathoracic illumination. A Veress needle inside the illumination sheath was inserted through a skin nick just to the left of the xiphoid process and angled toward the thorax. A needle containing a fiberscope within the lumen was inserted through the sheath and used to access the pericardium under direct visualization. A custom dilator and peel‐away sheath with pre‐tunneled fiberscope was passed over a guidewire into the pericardial space via modified Seldinger technique. A side‐biting multipolar pacemaker lead was inserted through the sheath and affixed against the epicardium.
Results
Six piglets (weight 3.7–4.0 kg) had successful lead implantation. The pericardial space could be visualized and entered in all animals. Median time from skin nick to sheath access of the pericardium was 9.5 (interquartile range [IQR] 8–11) min. Median total procedure time was 16 (IQR 14–19) min. Median R wave sensing was 5.4 (IQR 4.0–7.3) mV. Median capture threshold was 2.1 (IQR 1.7–2.4) V at 0.4 ms and 1.3 (IQR 1.2–2.0) V at 1.0 ms. There were no complications.
Conclusion
Percutaneous epicardial lead implantation under direct visualization was successful in six piglets of neonatal size and weight with clinically acceptable acute pacing parameters.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Cardiology, Nationwide Children's Hospital, Columbus, Ohio, USA.
- Department of Pediatrics, The Ohio State University College of Medicine, Columbus, Ohio, USA.
- Department of Mechanical Engineering, Johns Hopkins University, Baltimore, Maryland, USA.
- Sheikh Zayed Institute for Pediatric Surgical Innovation, Washington, District of Columbia, USA.
- Division of Cardiology, Children's National Hospital, Washington, District of Columbia, USA.
- Department of Pediatrics, George Washington University School of Medicine, Washington, District of Columbia, USA.
ORCID-Profile: Paige Mass
Zitieren
APA
Kumthekar, R. N., Opfermann, J. D., Mass, P., Contento, J. M., & Berul, C. I. (2023). Percutaneous epicardial pacing in infants using direct visualization: A feasibility animal study. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.15926
Vancouver
Kumthekar RN, Opfermann JD, Mass P, Contento JM, Berul CI. Percutaneous epicardial pacing in infants using direct visualization: A feasibility animal study. Journal of Cardiovascular Electrophysiology. 2023. doi:10.1111/jce.15926
BibTeX
@article{Kumthekar2023,
author = {Kumthekar, Rohan N and Opfermann, Justin D and Mass, Paige and Contento, Jacqueline M and Berul, Charles I},
title = {Percutaneous epicardial pacing in infants using direct visualization: A feasibility animal study},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2023},
doi = {10.1111/jce.15926}
}
EUVIMED