Diagnostic Performance of Automated Reticulocyte and Red Cell Indices for Detecting Iron Deficiency Anemia in Cirrhotic Patients
International Journal of Laboratory Hematology · 2026-12 · 48 (6) · 1356-1364
DOI: 10.1111/ijlh.70202 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Laboratory Hematology, online erschienen am 20.07.2026. Verfasst von Sahar Ma Ibrahim und 6 weiteren Autor:innen. Beteiligte Einrichtung: Clinical Pathology Department, National Liver Institute, Menoufia University, Shebeen El-Kom, Egypt. Hauptthemen laut MEDLINE: Anemia, Iron-Deficiency, Erythrocyte Indices, Liver Cirrhosis, Reticulocytes. Die Arbeit nennt 39 Literaturangaben.
Abstract (Originalsprache)
Introduction
Iron deficiency anemia (IDA) diagnosis in cirrhotic patients is challenging due to the confounding effects of inflammation and altered iron metabolism. Conventional markers, such as the iron profile, may be unreliable in this population. This study aimed to evaluate the diagnostic performance of reticulocyte hemoglobin content (CHr), percentage of microcytic red blood cells (%MicroR), and percentage of hypochromic red blood cells (%HypoR) for the diagnosis of IDA in cirrhotic patients.
Methods
Four groups were included in this cross‐sectional study: healthy controls, IDA non‐cirrhotic, IDA cirrhotic, and non‐IDA cirrhotic patients. Complete blood counts and reticulocyte indices were analyzed using a Sysmex XN‐ 1000 hematology analyzer, and receiver operating characteristic (ROC) curve analysis was performed to assess the diagnostic performance of the studied indices.
Results
CHr demonstrated the highest diagnostic performance, with an area under the curve (AUC) of 0.925 at a cut‐off value of ≤ 25.7 pg. %MicroR also demonstrated strong discriminative ability (AUC = 0.906) at a cut‐off of > 10.8%, whereas ferritin showed good diagnostic performance (AUC = 0.858) at a cut‐off of ≤ 101 μg/L.
Conclusion
CHr demonstrated excellent diagnostic performance and may serve as a valuable adjunctive marker for assessing iron status in cirrhotic patients, outperforming ferritin and other automated red blood cell indices. These findings support the potential role of reticulocyte indices as valuable tools for diagnosing IDA in cirrhosis, where conventional iron markers may be unreliable.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Clinical Pathology Department, National Liver Institute, Menoufia University, Shebeen El-Kom, Egypt.
- Epidemiology and Preventive Medicine Department, National Liver Institute, Menoufia University, Shebeen El-Kom, Egypt.
- Hepatology and Gastroenterology Department, National Liver Institute, Menoufia University, Shebeen El-Kom, Egypt.
- Hepatology, Gastroenterology and Infectious Disease Department, Al-Azhar University, Cairo, Egypt.
ORCID-Profile: Sally S Mandour
Zitieren
APA
Ibrahim, S. M., Gawad, A. A. A., Hendy, O. M., Dorgham, L. S., Abdelsameea, E., El Deen, N. M. B., & Mandour, S. S. (2026). Diagnostic Performance of Automated Reticulocyte and Red Cell Indices for Detecting Iron Deficiency Anemia in Cirrhotic Patients. International Journal of Laboratory Hematology, 48(6), 1356-1364. https://doi.org/10.1111/ijlh.70202
Vancouver
Ibrahim SM, Gawad AAA, Hendy OM, Dorgham LS, Abdelsameea E, El Deen NMB, et al. Diagnostic Performance of Automated Reticulocyte and Red Cell Indices for Detecting Iron Deficiency Anemia in Cirrhotic Patients. International Journal of Laboratory Hematology. 2026;48(6):1356-1364. doi:10.1111/ijlh.70202
BibTeX
@article{Ibrahim2026,
author = {Ibrahim, Sahar Ma and Gawad, Asmaa Ahmed Abd-El and Hendy, Olfat M and Dorgham, Laila Shehata and Abdelsameea, Eman and El Deen, Nessren Mohammed Bahaa and Mandour, Sally S},
title = {Diagnostic Performance of Automated Reticulocyte and Red Cell Indices for Detecting Iron Deficiency Anemia in Cirrhotic Patients},
journal = {International Journal of Laboratory Hematology},
year = {2026},
volume = {48},
number = {6},
pages = {1356--1364},
doi = {10.1111/ijlh.70202}
}