Pilot‐Scale Evaluation of Partial‐Slide Imaging for Detecting Critical Morphological Features (Excluding Parasites): Initial UK Implementation of a Hybrid Virtual–Light Microscopy Model in Haematology
International Journal of Laboratory Hematology · 2026-12 · 48 (6) · 1330-1340
DOI: 10.1111/ijlh.70185 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Laboratory Hematology, online erschienen am 03.07.2026. Verfasst von Usman Ali und 5 weiteren Autor:innen. Beteiligte Einrichtung: Blood Sciences, Haematology, The Royal London Hospital, London, UK. Hauptthemen laut MEDLINE: Microscopy, Image Processing, Computer-Assisted, Hematology. Die Arbeit nennt 36 Literaturangaben.
Abstract (Originalsprache)
Background
This UK‐based pilot study explores a hybrid approach integrating virtual (VM) and light microscopy (LM), using partial‐slide imaging (PSI) at ×20 magnification (equivalent to approximately ×58 effective output magnification).
Method
This study utilised a 3D Histech Pannoramic MIDI II Slide scanner to digitise 50 peripheral blood smears. PSI was specifically defined by capturing the feathered edge and monolayer regions of the smears. A diverse range of 44 conditions, characterised by distinct morphological features across red blood cell, white blood cell and platelet lineages, was investigated. A comparative analysis between traditional LM and VM was then performed, alongside assessment of scan times and storage requirements, with statistical analyses using Student's t ‐tests, where a p ≤ 0.05 was considered statistically significant.
Results
Excluding parasites, the critical morphological features were reliably assessed, with diagnostic images from VM judged comparable and adequate to LM: the three subcellular features—Döhle bodies, basophilic stippling and Auer rods—were verified side‐by‐side, while overall 44 features were assessed using a three‐point graded scale. Operationally, PSI at ×20 magnification demonstrated substantial efficiency, with a mean scan time of 4 min 17 s and storage of 1.18 GB per slide, representing statistically significant reductions ( p < 0.05) in both scanning time and storage compared to whole‐slide imaging.
Conclusion
This pilot‐scale study supports a hybrid model, demonstrating that PSI at 20× shows promising feasibility for morphological assessment within a hybrid VM–LM model, but further validation is required to confirm diagnostic equivalence with LM. This study offers significantly reduced scan times and storage requirements compared to whole‐slide imaging.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Blood Sciences, Haematology, The Royal London Hospital, London, UK.
ORCID-Profile: Usman Ali
Zitieren
APA
Ali, U., Dadhra, J., Abukar, J., Lee, K., Smith, M., & Butler, T. (2026). Pilot‐Scale Evaluation of Partial‐Slide Imaging for Detecting Critical Morphological Features (Excluding Parasites): Initial UK Implementation of a Hybrid Virtual–Light Microscopy Model in Haematology. International Journal of Laboratory Hematology, 48(6), 1330-1340. https://doi.org/10.1111/ijlh.70185
Vancouver
Ali U, Dadhra J, Abukar J, Lee K, Smith M, Butler T. Pilot‐Scale Evaluation of Partial‐Slide Imaging for Detecting Critical Morphological Features (Excluding Parasites): Initial UK Implementation of a Hybrid Virtual–Light Microscopy Model in Haematology. International Journal of Laboratory Hematology. 2026;48(6):1330-1340. doi:10.1111/ijlh.70185
BibTeX
@article{Ali2026,
author = {Ali, Usman and Dadhra, Juswal and Abukar, Jibril and Lee, Kurtis and Smith, Matthew and Butler, Tom},
title = {Pilot‐Scale Evaluation of Partial‐Slide Imaging for Detecting Critical Morphological Features (Excluding Parasites): Initial UK Implementation of a Hybrid Virtual–Light Microscopy Model in Haematology},
journal = {International Journal of Laboratory Hematology},
year = {2026},
volume = {48},
number = {6},
pages = {1330--1340},
doi = {10.1111/ijlh.70185}
}
EUVIMED