Multicenter Interlaboratory Study of Thrombin Generation Assay in Healthy Adults Using the ST Genesia Analyzer With the STG‐ThromboScreen Reagent (With and Without Thrombomodulin)
International Journal of Laboratory Hematology · 2026
DOI: 10.1111/ijlh.70168 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift International Journal of Laboratory Hematology, 2026. Verfasst von Diego Velasco-Rodríguez und 21 weiteren Autor:innen. Beteiligte Einrichtung: Hospital Universitario Fundación Jiménez Díaz, Madrid, Spain. Hauptthemen laut MEDLINE: Thrombin, Thrombomodulin. Die Arbeit nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
Fully automated thrombin generation (TG) assays facilitate the transition from research to clinical application. We conducted a multicenter inter‐laboratory study using the ST Genesia analyzer to evaluate assay performance and establish reference intervals in healthy adults from Spain and Portugal.
Methods
In this multicenter study, plasma samples from healthy adult volunteers were analyzed across 12 laboratories following a harmonized protocol. TG was measured using the STG‐ThromboScreen reagent in the presence and absence of thrombomodulin (TM). Analytical performance and reference intervals were assessed.
Results
Within‐run and intra‐day precision were excellent (coefficient of variation < 10%) for all TG parameters except velocity index. Normalization reduced interlaboratory variability for amplitude‐based parameters such as endogenous thrombin potential (ETP) and peak height, whereas its effect was less pronounced for velocity index. Satisfactory interlaboratory performance (| z | ≤ 2) was achieved in 96%–98% of results, and normalized parameters showed minimal bias across centers. In 252 healthy adults, median normalized values were: ETP 131% (IQR 118–142), peak height 116% (IQR 104–127), lag time 2.6 min (IQR 2.3–2.8), and time‐to‐peak 5.5 min (IQR 5.0–5.9). Median ETP inhibition in the presence of TM was 47.2% (IQR 35.1–61.8). Age had a modest influence on TG parameters, whereas sex differences were observed.
Conclusions
This multicenter interlaboratory study in the Iberian Peninsula demonstrates high reproducibility of TG measurement using the ST Genesia platform with the STG‐ThromboScreen reagent. These findings support its use for standardized TG assessment and provide reference intervals for clinical and research settings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Hospital Universitario Fundación Jiménez Díaz, Madrid, Spain.
- Clínica Universidad de Navarra, Pamplona, Spain.
- Diagnostica Stago S.A.S, Asnières-sur-Seine, France.
- Hospital Santa Bárbara, Soria, Spain.
- Hospital de la Santa Creu i Sant Pau, Barcelona, Spain.
- Hospital General Universitario Doctor Balmis, Alicante, Spain.
ORCID-Profile: Diego Velasco-Rodríguez, Santiago Bonanad, Susana Gassiot
Zitieren
APA
Velasco-Rodríguez, D., Marcos-Jubilar, M., Martínez-Alfonzo, I., Martos, L., Aguilar, C., Vilalta, N., Llobet, D., Marco-Rico, A., Aragoneses, D., Marco, P., Ricca, L., Ferreira, D., Lopes, S., Mariano-Pego, J., Bernardo, Á., Moscardó, A., Bonanad, S., Gassiot, S., Gutiérrez-Jomarrón, I., …, & Reverter, J. C. (2026). Multicenter Interlaboratory Study of Thrombin Generation Assay in Healthy Adults Using the ST Genesia Analyzer With the STG‐ThromboScreen Reagent (With and Without Thrombomodulin). International Journal of Laboratory Hematology. https://doi.org/10.1111/ijlh.70168
Vancouver
Velasco-Rodríguez D, Marcos-Jubilar M, Martínez-Alfonzo I, Martos L, Aguilar C, Vilalta N, et al. Multicenter Interlaboratory Study of Thrombin Generation Assay in Healthy Adults Using the ST Genesia Analyzer With the STG‐ThromboScreen Reagent (With and Without Thrombomodulin). International Journal of Laboratory Hematology. 2026. doi:10.1111/ijlh.70168
BibTeX
@article{VelascoRodrguez2026,
author = {Velasco-Rodríguez, Diego and Marcos-Jubilar, María and Martínez-Alfonzo, Inés and Martos, Laura and Aguilar, Carlos and Vilalta, Noelia and Llobet, Dolors and Marco-Rico, Ana and Aragoneses, David and Marco, Pascual and Ricca, Luciana and Ferreira, David and Lopes, Sara and Mariano-Pego, João and Bernardo, Ángel and Moscardó, Antonio and Bonanad, Santiago and Gassiot, Susana and Gutiérrez-Jomarrón, Isabel and Páramo, Jose Antonio and Llamas-Sillero, Pilar and Reverter, Joan Carles},
title = {Multicenter Interlaboratory Study of Thrombin Generation Assay in Healthy Adults Using the ST Genesia Analyzer With the STG‐ThromboScreen Reagent (With and Without Thrombomodulin)},
journal = {International Journal of Laboratory Hematology},
year = {2026},
doi = {10.1111/ijlh.70168}
}
EUVIMED