EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Assessment of the Performance of Siemens Scopio Digital Morphology on Bone Marrow Aspirates in Onco‐Hematology

Gina Zini iD, John Marra, Elena Rossi, Silvia Bellesi, Nicoletta Pelliccioni, Giuseppe d'Onofrio, Patrizia Chiusolo

International Journal of Laboratory Hematology · 2026

DOI: 10.1111/ijlh.70138 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Laboratory Hematology, 2026. Verfasst von Gina Zini und 6 weiteren Autor:innen. Beteiligte Einrichtung: Sezione di Ematologia, Dipartimento di Scienze Radiologiche ed Ematologiche, Università Cattolica del Sacro Cuore, Rome, Italy. Hauptthemen laut MEDLINE: Bone Marrow, Bone Marrow Cells. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Siemens Healthineers, Fondazione Policlinico Gemelli.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Digital morphology (DM) systems assisted by artificial intelligence are increasingly being introduced into hematology laboratories; however, data on their performance in routine clinical practice for bone marrow aspirates (BMA) remain limited. We evaluated the automated pre‐classification generated by a full‐field DM system in a large series of BMA samples from patients with onco‐hematological disorders.

Methods

We analyzed 350 BMA samples during routine diagnostic activity; they were evaluated using the Siemens Scopio X100 HT system. Automated results were compared with conventional optical microscopy (OM), which served as the reference method for bone marrow differential counts. Results obtained in pre‐classification and post‐classification were analyzed separately across major cellular lineages and clinically relevant blast thresholds.

Results

Sixteen markedly hypercellular samples could not be quantitatively evaluated. In the 334 evaluable BMA samples, the granulocytic series showed very strong correlation between DM and OM at pre‐classification ( r = 0.85, 95% CI 0.82–0.88), with further improvement after post‐classification ( r = 0.93, 95% CI 0.91–0.94). Erythroblast percentage showed strong correlation at pre‐classification ( r = 0.78, 95% CI 0.74–0.82) and very strong correlation after post‐classification ( r = 0.93, 95% CI 0.92–0.95). Lymphocyte percentage showed moderate correlation at pre‐classification ( r = 0.55, 95% CI 0.47–0.62) and strong correlation after post‐classification ( r = 0.78, 95% CI 0.73–0.82). Blast percentage showed strong correlation overall at pre‐classification ( r = 0.73, 95% CI 0.67–0.77), with very strong correlation in samples with < 5% blasts ( r = 0.91, 95% CI 0.88–0.93), but lower correlation in samples with > 5% blasts ( r = 0.59, 95% CI 0.48–0.68), improving after post‐classification ( r = 0.84, 95% CI 0.78–0.88). Low‐frequency cell populations and challenging hypercellular smears remained more problematic and required expert morphologic review.

Conclusions

In routine BMA evaluation, the Scopio DM system provided good overall performance for the major marrow lineages, with clear improvement after expert post‐classification. Its main value lies in supporting a supervised diagnostic workflow rather than replacing expert microscopic assessment, particularly in challenging samples and in low‐frequency or morphologically heterogeneous cell populations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Bone MarrowBone Marrow CellsHumansFemaleBone Marrow Examination

Stichwörter der Autor:innen

artificial intelligenceblood morphologybone marrow aspiratedigital morphologyfull‐field smear analysisonco‐hematological disorders

Autor:innen und Institutionen

  • Sezione di Ematologia, Dipartimento di Scienze Radiologiche ed Ematologiche, Università Cattolica del Sacro Cuore, Rome, Italy.
  • Dipartimento di Diagnostica per Immagini, Radioterapia Oncologica ed Ematologia, Fondazione Policlinico Universitario "A. Gemelli" IRCCS, Rome, Italy.

ORCID-Profile: Gina Zini

Förderung

  • Siemens Healthineers
  • Fondazione Policlinico Gemelli

Zitieren

APA

Zini, G., Marra, J., Rossi, E., Bellesi, S., Pelliccioni, N., d'Onofrio, G., & Chiusolo, P. (2026). Assessment of the Performance of Siemens Scopio Digital Morphology on Bone Marrow Aspirates in Onco‐Hematology. International Journal of Laboratory Hematology. https://doi.org/10.1111/ijlh.70138

Vancouver

Zini G, Marra J, Rossi E, Bellesi S, Pelliccioni N, d'Onofrio G, et al. Assessment of the Performance of Siemens Scopio Digital Morphology on Bone Marrow Aspirates in Onco‐Hematology. International Journal of Laboratory Hematology. 2026. doi:10.1111/ijlh.70138

BibTeX
@article{Zini2026,
  author = {Zini, Gina and Marra, John and Rossi, Elena and Bellesi, Silvia and Pelliccioni, Nicoletta and d'Onofrio, Giuseppe and Chiusolo, Patrizia},
  title = {Assessment of the Performance of Siemens Scopio Digital Morphology on Bone Marrow Aspirates in Onco‐Hematology},
  journal = {International Journal of Laboratory Hematology},
  year = {2026},
  doi = {10.1111/ijlh.70138}
}
Kostenlos starten →