Laboratory Identification of Lupus Anticoagulant ( LA ) Using Different Activated Partial Thromboplastin Time ( APTT ) Assays

Bárbara G Barion iD, Bianca Stefanello, Maria Luiza S A de Paula, Thaís S Saraiva, Paula R Villaça, Vanderson Rocha, Fernanda A Orsi iD, Tania R F da Rocha

International Journal of Laboratory Hematology · 2026

DOI: 10.1111/ijlh.14549 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Laboratory Hematology, 2026. Verfasst von Bárbara G Barion und 7 weiteren Autor:innen. Beteiligte Einrichtung: Hospital das Clínicas da Faculdade de Medicina de Sao Paulo, Brazil. Hauptthemen laut MEDLINE: Lupus Coagulation Inhibitor. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Siemens Healthineers.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

The International Society of Thrombosis and Hemostasis (ISTH) guidelines suggest a three‐step evaluation for the detection of lupus anticoagulant (LA), including screening, mixing, and confirmation. According to the guidelines, the LA assay based on activated partial thromboplastin time (APTT) should include an initial screening step followed by a confirmatory step that uses a higher concentration of phospholipids in either bilayer or hexagonal form. For the activator, the guidelines recommend using silica, though ellagic acid is also an option. In this context, HemosIL Silica Clotting Time (SCT, Instrumentation Laboratory) is the only assay that fully complies with the guidelines. However, there are other assays available using different reagents, such as Dade Actin FSL/FS (Siemens Healthcare Diagnostics) and PTT‐LA/Staclot LA (Diagnostica Stago), and the relevance of these differences in LA detection is not known.

Methods

This study compared the performance of the three platforms.

Results

Out of 136 samples, the majority were from females (82%) with a median age of 41 years (IQR 32–50); 44 (32%) had a history of thrombosis, and 28 (21%) were on anticoagulants. PTT‐LA/Staclot LA had the highest sensitivity (100%) and specificity (100%). There was an almost perfect agreement between PTT‐LA/Staclot LA and Dade Actin FSL/FS (kappa 0.812). HemosIL SCT sensitivity was 100% and the specificity was 74%, which was increased to 99% by increasing the phospholipid concentration of the screening step.

Conclusion

We observed a good agreement between PTT‐LA/Staclot LA and Dade Actin FSL/FS, and fair to moderate agreement with HemosIL SCT, whose performance improved with increasing phospholipid concentration. These results demonstrate that all three assays are comparable for APTT‐LA detection.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Lupus Coagulation InhibitorHumansPartial Thromboplastin TimeFemaleAdultMaleMiddle AgedSensitivity and SpecificityBlood Coagulation Tests

Stichwörter der Autor:innen

APTTantiphospholipid antibodiesantiphospholipid syndromelupus anticoagulant

Substanzen

Lupus Coagulation Inhibitor

Autor:innen und Institutionen

  • Hospital das Clínicas da Faculdade de Medicina de Sao Paulo, Brazil.
  • University of Sao Paulo Medical School, Brazil.
  • Department of Pathology, School of Medical Sciences, University of Campinas (UNICAMP), Campinas, Brazil.

ORCID-Profile: Bárbara G Barion, Fernanda A Orsi

Förderung

  • Siemens Healthineers

Zitieren

APA

Barion, B. G., Stefanello, B., de Paula, M. L. S. A., Saraiva, T. S., Villaça, P. R., Rocha, V., Orsi, F. A., & da Rocha, T. R. F. (2026). Laboratory Identification of Lupus Anticoagulant ( LA ) Using Different Activated Partial Thromboplastin Time ( APTT ) Assays. International Journal of Laboratory Hematology. https://doi.org/10.1111/ijlh.14549

Vancouver

Barion BG, Stefanello B, de Paula MLSA, Saraiva TS, Villaça PR, Rocha V, et al. Laboratory Identification of Lupus Anticoagulant ( LA ) Using Different Activated Partial Thromboplastin Time ( APTT ) Assays. International Journal of Laboratory Hematology. 2026. doi:10.1111/ijlh.14549

BibTeX
@article{Barion2026,
  author = {Barion, Bárbara G and Stefanello, Bianca and de Paula, Maria Luiza S A and Saraiva, Thaís S and Villaça, Paula R and Rocha, Vanderson and Orsi, Fernanda A and da Rocha, Tania R F},
  title = {Laboratory Identification of Lupus Anticoagulant ( LA ) Using Different Activated Partial Thromboplastin Time ( APTT ) Assays},
  journal = {International Journal of Laboratory Hematology},
  year = {2026},
  doi = {10.1111/ijlh.14549}
}
Kostenlos starten →