Evaluation of 5B9 as a Calibrator or Expression of Results in Absorbance Values for the Diagnosis of Hit With a PF4/Heparin Specific Elisa
International Journal of Laboratory Hematology · 2026
DOI: 10.1111/ijlh.14547 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Laboratory Hematology, 2026. Verfasst von Claire Pouplard und 10 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory of Haematology-Hemostasis, University Hospital of Tours - INSERM UMR 1327 ISCHEMIA, Tours, France. Hauptthemen laut MEDLINE: Heparin, Platelet Factor 4, Thrombocytopenia, Antibodies, Monoclonal. Die Arbeit nennt 24 Literaturangaben. Gefördert von: Stago.
Abstract (Originalsprache)
Background
Immunoassays detecting anti‐PF4/H antibodies must be sensitive to exclude heparin‐induced thrombocytopenia (HIT), and optical density (OD) values are useful for confirming HIT, but no calibration is currently available.
Objectives
To study the impact of OD values on the performance of the Asserachrom HPIA IgG in a cohort of patients with suspected HIT, and the value of a calibration performed with 5B9, a HIT monoclonal antibody.
Methods
The HPIA IgG was performed in 170 patients with a high or intermediate probability of HIT. Results were expressed in OD 450 or ‘5B9 equivalent’ units, using a calibration done with 5B9. HIT was confirmed when HPIA and SRA/PF4 tests were positive.
Results
HIT was excluded in 97 cases because HPIA and SRA/PF4 were negative. The HPIA was positive in 73 cases and HIT confirmed in 43 cases (SRA/PF4+). Applying an OD threshold of 1.05, the NPV and PPV of the test were 98% and 83%, respectively. Calibration of HPIA with 5B9 did not improve its performance, since similar AUC values (ROC curves) were obtained whether results were expressed in OD values or in equivalent units of 5B9. Bayesian analysis showed that in patients with an intermediate pre‐test probability of HIT, the post‐test probability equalled 1% when OD was less than 1, and 100% when OD was over 2.
Conclusion
5B9 as a calibrator failed to improve the performance of HPIA, but this assay can reliably exclude (when negative) or confirm HIT (when OD > 2), without requiring a functional assay.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Laboratory of Haematology-Hemostasis, University Hospital of Tours - INSERM UMR 1327 ISCHEMIA, Tours, France.
- Stago, Asnières sur Seine Cedex, France.
- Institute of Haematology and Transfusion, Lille University Hospital, Lille, France.
- Department of Hematology, CHRU de Brest, Brest, France.
- Haemostais Unit, Dijon University Hospital, Dijon, France.
- Laboratory of Haematology-Hemostasis, University Hospital of Rennes, Rennes, France.
ORCID-Profile: Eve-Anne Guery
Förderung
- Stago
Zitieren
APA
Pouplard, C., Charuel, N., Archer, E., Vayne, C., Bauters, A., Jaouen, S., Savard, P., Maucorps, L., Guery, E. A., Gruel, Y., & Rollin, J. (2026). Evaluation of 5B9 as a Calibrator or Expression of Results in Absorbance Values for the Diagnosis of Hit With a PF4/Heparin Specific Elisa. International Journal of Laboratory Hematology. https://doi.org/10.1111/ijlh.14547
Vancouver
Pouplard C, Charuel N, Archer E, Vayne C, Bauters A, Jaouen S, et al. Evaluation of 5B9 as a Calibrator or Expression of Results in Absorbance Values for the Diagnosis of Hit With a PF4/Heparin Specific Elisa. International Journal of Laboratory Hematology. 2026. doi:10.1111/ijlh.14547
BibTeX
@article{Pouplard2026,
author = {Pouplard, Claire and Charuel, Noémie and Archer, Estelle and Vayne, Caroline and Bauters, Anne and Jaouen, Simon and Savard, Philippe and Maucorps, Laure and Guery, Eve-Anne and Gruel, Yves and Rollin, Jérôme},
title = {Evaluation of 5B9 as a Calibrator or Expression of Results in Absorbance Values for the Diagnosis of Hit With a PF4/Heparin Specific Elisa},
journal = {International Journal of Laboratory Hematology},
year = {2026},
doi = {10.1111/ijlh.14547}
}
EUVIMED