Definition of Significant Platelet Clumping: Should We Review All Samples With a Platelet Clumping Flag From Automated Hematology Analyzer?
International Journal of Laboratory Hematology · 2025
DOI: 10.1111/ijlh.14387 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Laboratory Hematology, 2025. Verfasst von Suhyeon Woo und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, Soonchunhyang University Cheonan Hospital, Soonchunhyang University College of Medicine, Cheonan, Korea. Hauptthemen laut MEDLINE: Platelet Aggregation, Blood Platelets. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Soonchunhyang University Research Fund.
Abstract (Originalsprache)
Objectives
This study aimed to determine a definition for significant platelet clumping (PC) and evaluate the performance of the Sysmex XN instrument for detecting platelet clumps.
Methods
For part 1, 372 specimens with a ‘PLT_clump?’ flag in XN‐9000 were classified into five groups according to the average number of PCs. We compared the initial platelet count (measured by XN‐9000 using impedance method) and corrected platelet count (counted optically or re‐analyzed by XN‐9000 using vortexed or re‐collected sample) of each group. For part 2, 1000 specimens with a PC flag divided into three subgroups {group N (PC = 0), Y (PC ≥ 1), and Z (microscopic fibrin clot)} and additional two groups {group S (PC(+) specimens without any flag and with flags of other categories) and group NC (negative control)} were collected. Positive predictive value (PPV), negative predictive value (NPV), sensitivity and specificity of PC detection of XN‐9000 were obtained and the platelet counts and four indices (PDW, MPV, P_LCR, and PCT) of groups NC, N, Y, Z, and S were compared to detect PC more precisely.
Results
In part 1, all groups showed significant difference between the initial and corrected platelet counts. In part 2, PPV, NPV, prevalence, sensitivity, and specificity were 41.5%, 56.5%, 43.4%, 2.18%, and 98.3%, respectively. The platelet counts and four indices showed statistical differences for detecting PCs, and especially PDW and P_LCR were significantly smaller in group Z than group N or Y.
Conclusions
We suggest the definition of significant PC by the presence of at least three platelets. In addition, utilizing platelet‐related indices should be developed to improve the efficiency of the PC detection.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Laboratory Medicine, Soonchunhyang University Cheonan Hospital, Soonchunhyang University College of Medicine, Cheonan, Korea.
- Clinical Trial Center, Soonchunhyang University Cheonan Hospital, Soonchunhyang University College of Medicine, Cheonan, Korea.
ORCID-Profile: Suhyeon Woo, Bohyun Kim, Nam Hun Heo, Min-Sun Kim, Young Ahn Yoon, Young-Jin Choi
Förderung
- Soonchunhyang University Research Fund
Zitieren
APA
Woo, S., Kim, B., Heo, N. H., Kim, M. S., Yoon, Y. A., & Choi, Y. J. (2025). Definition of Significant Platelet Clumping: Should We Review All Samples With a Platelet Clumping Flag From Automated Hematology Analyzer?. International Journal of Laboratory Hematology. https://doi.org/10.1111/ijlh.14387
Vancouver
Woo S, Kim B, Heo NH, Kim MS, Yoon YA, Choi YJ. Definition of Significant Platelet Clumping: Should We Review All Samples With a Platelet Clumping Flag From Automated Hematology Analyzer?. International Journal of Laboratory Hematology. 2025. doi:10.1111/ijlh.14387