Stable measurement of chemistry, immunochemistry, and hematology analytes in a heparin‐based anticoagulant with iloprost additive: A promising candidate for the polyvalent blood collection tube
International Journal of Laboratory Hematology · 2024
DOI: 10.1111/ijlh.14291 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Laboratory Hematology, 2024. Verfasst von Elias Frost Wiwe und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Biochemistry, Nordsjællands Hospital, Copenhagen University Hospital, Hillerød, Denmark. Hauptthemen laut MEDLINE: Heparin, Anticoagulants, Iloprost, Blood Specimen Collection. Die Arbeit nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Innovationsfonden, Radiometer Medical ApS.
Abstract (Originalsprache)
Introduction
A polyvalent blood collection tube could potentially reduce the number and volume of blood samples drawn from patients and reduce the risk of tube mix‐ups in a point‐of‐care setting in the emergency department and the intensive care unit.
Methods
Four different concentrations of our experimental heparin anticoagulant with iloprost additive (HEP‐ILOP 50 nM, 150 nM, 1000 nM, and 10 μM, respectively) were tested for significant differences and bias performance specifications against EDTA for 29 hematology analytes, and the highest concentration (HEP‐ILOP 10 μM) against lithium heparin for 14 chemistry and immunochemistry analytes. Samples were drawn from 79 consenting subjects from the Oncology Department ( n = 38) and the Intensive and Intermediary Care Unit ( n = 41).
Results
For hematology analytes, the HEP‐ILOP formulation generally provided stable measurement within optimal requirements within 5 h after sampling (mean 104 ± 56 min), with very little difference between the four HEP‐ILOP concentrations. Because of differences in platelet and red blood cell swelling between EDTA and HEP‐ILOP, all size‐dependent analytes required proportional factorization to produce similar results. Platelet count by impedance similarly required factorization, whereas the fluorescent method provided results identical with EDTA. Chemistry and immunochemistry analytes were within optimal requirements except for potassium, lactate dehydrogenase, and glucose, indicating a cytoprotective effect of iloprost reducing cell metabolism and rupture, thereby producing results closer to in vivo conditions.
Conclusions
Our novel dry‐sprayed anticoagulant formulation, HEP‐ILOP, is a promising candidate for a polyvalent blood collection tube, enabling the analysis of hematology, chemistry, and immunochemistry analytes in the same tube.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Clinical Biochemistry, Nordsjællands Hospital, Copenhagen University Hospital, Hillerød, Denmark.
- Department of Anesthesia and Intensive Care, Nordsjællands Hospital, Copenhagen University Hospital, Hillerød, Denmark.
- Department of Oncology and Palliative Care, Nordsjællands Hospital, Copenhagen University Hospital, Hillerød, Denmark.
- Radiometer Medical ApS, Brønshøj, Denmark.
ORCID-Profile: Elias Frost Wiwe
Förderung
- Innovationsfonden
- Radiometer Medical ApS
Zitieren
APA
Wiwe, E. F., Schmidt, L. E., Zeuthen, E., Engelstoft, I. M. S., Christiansen, B. B., Hansen, T. S., Burkhardt, M. A., & Jensen, C. A. J. (2024). Stable measurement of chemistry, immunochemistry, and hematology analytes in a heparin‐based anticoagulant with iloprost additive: A promising candidate for the polyvalent blood collection tube. International Journal of Laboratory Hematology. https://doi.org/10.1111/ijlh.14291
Vancouver
Wiwe EF, Schmidt LE, Zeuthen E, Engelstoft IMS, Christiansen BB, Hansen TS, et al. Stable measurement of chemistry, immunochemistry, and hematology analytes in a heparin‐based anticoagulant with iloprost additive: A promising candidate for the polyvalent blood collection tube. International Journal of Laboratory Hematology. 2024. doi:10.1111/ijlh.14291
BibTeX
@article{Wiwe2024,
author = {Wiwe, Elias Frost and Schmidt, Laura Emilie and Zeuthen, Emilie and Engelstoft, Ida Maria Storm and Christiansen, Buris Brinch and Hansen, Thomas Steen and Burkhardt, Melanie A and Jensen, Claus Antonio Juel},
title = {Stable measurement of chemistry, immunochemistry, and hematology analytes in a heparin‐based anticoagulant with iloprost additive: A promising candidate for the polyvalent blood collection tube},
journal = {International Journal of Laboratory Hematology},
year = {2024},
doi = {10.1111/ijlh.14291}
}
EUVIMED