Photon‐Counting CT for Diagnosing Vertical Root Fractures in Teeth With Metal Posts: An Ex Vivo Comparative Analysis With Four CBCT Devices
International Endodontic Journal · 2026
DOI: 10.1111/iej.70095 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift International Endodontic Journal, 2026. Verfasst von Renata M S Leal und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Restorative Dentistry, Federal University of Paraná, Paraná, Brazil. Hauptthemen laut MEDLINE: Tooth Root, Tooth Fractures, Cone-Beam Computed Tomography, Post and Core Technique. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Coordenação de Aperfeiçoamento de Pessoal de Nível Superior.
Abstract (Originalsprache)
Objective
Photon‐counting computed tomography (PCCT) represents a major innovation in X‐ray detection technology, offering improved signal efficiency and reduced electronic noise compared with cone‐beam computed tomography (CBCT), which can enhance image quality. This study aimed to evaluate the diagnostic performance of PCCT in detecting vertical root fractures (VRF), in comparison with four CBCT devices. Methodology Eighteen single‐rooted teeth were endodontically treated, and VRF was induced in eight of them. Each tooth was individually placed into the mandibular first premolar empty socket of an anthropomorphic phantom and scanned under three conditions: without a metal post, with a nickel‐chromium metal post (Ni‐Cr), and with a cobalt‐chromium metal post (Co‐Cr) in five CT devices: the NAEOTOM Alpha PCCT (Siemens Healthineers) device and four CBCT devices (3D Accuitomo 170—Morita, Veraview X800—Morita, NewTom VGi evo—NewTom, and Carestream 9600—Carestream). The highest‐resolution protocol available on each device was used, resulting in a total of 270 scans. Five experienced dentomaxillofacial radiologists independently and blindly evaluated the scans using a five‐point confidence scale. Diagnostic accuracy was assessed by calculating the area under the ROC curve (AUC), sensitivity, and specificity, with results compared by two‐way ANOVA with post hoc Tukey's test ( α = 0.05).
Results
NewTom VGi and PCCT devices showed significantly higher AUC values than the Veraview X800, regardless of the metal post material ( p < 0.05). CS9600 and PCCT devices exhibited significantly higher sensitivity values in diagnosing with Ni‐Cr posts than the Accuitomo 3D and Veraview X800 devices ( p < 0.05). With the Co‐Cr metal post, the NewTom VGi, CS9600, and PCCT devices showed significantly higher sensitivity values compared to the Veraview X800 device ( p < 0.05). There were no statistically significant differences in specificity, regardless of the CT device or metal post material ( p > 0.05).
Conclusions
The NEAOTOM Alpha PCCT showed high diagnostic accuracy for VRF detection in an ex vivo model, comparable to high‐resolution CBCT devices, highlighting its diagnostic performance under controlled ex vivo conditions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Restorative Dentistry, Federal University of Paraná, Paraná, Brazil.
- Department of Oral Diagnosis, Area of Dental Radiology, Piracicaba Dental School, State University of Campinas (UNICAMP), São Paulo, Brazil.
- Department of Mechanical Engineering, Federal University of Paraná, Paraná, Brazil.
- Department of Radiology, University Hospitals Leuven, Leuven, Belgium.
- OMFS IMPATH Research Group, Department of Imaging and Pathology, Faculty of Medicine, University of Leuven, Leuven, Belgium.
- Department of Oral and Maxillofacial Surgery, University Hospitals Leuven, Leuven, Belgium.
ORCID-Profile: Renata M S Leal, Fernanda B Fagundes, Maria F S A Bortoletto, Samuel C Kluthcovsky, Walter Coudyzer, Bruno C Cavenago, Reinhilde Jacobs, Rocharles Cavalcante Fontenele
Förderung
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior
Zitieren
APA
Leal, R. M. S., Fagundes, F. B., Bortoletto, M. F. S. A., Kluthcovsky, S. C., Coudyzer, W., Cavenago, B. C., Jacobs, R., & Fontenele, R. C. (2026). Photon‐Counting CT for Diagnosing Vertical Root Fractures in Teeth With Metal Posts: An Ex Vivo Comparative Analysis With Four CBCT Devices. International Endodontic Journal. https://doi.org/10.1111/iej.70095
Vancouver
Leal RMS, Fagundes FB, Bortoletto MFSA, Kluthcovsky SC, Coudyzer W, Cavenago BC, et al. Photon‐Counting CT for Diagnosing Vertical Root Fractures in Teeth With Metal Posts: An Ex Vivo Comparative Analysis With Four CBCT Devices. International Endodontic Journal. 2026. doi:10.1111/iej.70095