A Novel Helicobacter pylori Test Based on 16S rRNA Isothermal Amplification Point‐of‐Care Testing of Throat Swab
Helicobacter · 2026-09 · 31 (5)
DOI: 10.1111/hel.70164 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Helicobacter, online erschienen am 08.09.2026. Verfasst von Ziyu Liu und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Endoscopy, State Key Laboratory of Oncology in South China, Collaborative Innovation Center for Cancer Medicine, Sun Yat-Sen University…. Hauptthemen laut MEDLINE: Helicobacter Infections, Helicobacter pylori, Molecular Diagnostic Techniques, Nucleic Acid Amplification Techniques. Die Arbeit nennt 15 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
Helicobacter pylori ( H. pylori ) infection is a major risk factor for gastric cancer (GC) and other gastrointestinal diseases. Current non‐invasive detection methods for H. pylori , such as 13 C‐urea breath test (UBT), are constrained by high cost, radiation concerns, prolonged test duration, and susceptibility to false negative results due to medication use. Thus, a rapid, cost‐effective, accurate, and medication‐resistant detection method for H. pylori infection is urgently needed. Materials and Methods This study developed a novel point‐of‐care testing for H. pylori infection based on 16S rRNA isothermal amplification, using throat swabs as the detection samples (POCT). To systematically evaluate its diagnostic performance, a multi‐center cross‐sectional paired diagnostic study was conducted at Shandong Public Health Clinical Center and East Campus of Qilu Hospital, Shandong University from January 2025 to August 2025. This study enrolled a total of 543 participants, all of whom underwent both POCT and UBT simultaneously on the same day. Regarding UBT as the reference gold standard, the diagnosis metrics including accuracy, sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of POCT for H. pylori detection were analyzed to evaluate its clinical application.
Results
The POCT assay exhibited an overall accuracy of 95%, sensitivity of 89%, and specificity of 97%. PPV and NPV were 90% and 96%, respectively. Its high agreement with UBT was confirmed by a Kappa value of 0.860 ( p < 0.01). Additionally, the diagnostic performance of POCT also demonstrated high consistency between two centers.
Conclusions
The 16S rRNA isothermal amplification point‐of‐care testing exhibits excellent diagnostic accuracy for H. pylori infection. It holds the potential to serve as an economic, efficient, and widely applicable alternative to the traditional UBT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Endoscopy, State Key Laboratory of Oncology in South China, Collaborative Innovation Center for Cancer Medicine, Sun Yat-Sen University Cancer Center, Guangzhou, Guangdong, China.
- Zhongshan School of Medicine, Sun Yat-Sen University, Guangzhou, Guangdong, China.
- Cancer Prevention Center, State Key Laboratory of Oncology in South China, Collaborative Innovation Center for Cancer Medicine, Sun Yat-Sen University Cancer Center, Guangzhou, Guangdong, China.
- Department of Radiology, State Key Laboratory of Oncology in South China, Guangdong Provincial Clinical Research Center for Cancer, Sun Yat-Sen University Cancer Center, Guangzhou, Guangdong, China.
- School of Medicine, Huaqiao University, Quanzhou, Fujian, China.
- Center for Mass Spectrometry Research and Clinical Application, and Center for Integrative and Translational Medicine, Shandong Public Health Clinical Center Affiliated to Shandong University, Jinan, Shandong, China.
ORCID-Profile: Ziyu Liu, Minqing Wu, Yong Diao, Chuanbo Xie
Zitieren
APA
Liu, Z., Wu, M., Li, Y., Huang, Q., Mao, Z., Diao, Y., Yang, H., Liu, Y., Xie, C., & Yu, Z. (2026). A Novel Helicobacter pylori Test Based on 16S rRNA Isothermal Amplification Point‐of‐Care Testing of Throat Swab. Helicobacter, 31(5). https://doi.org/10.1111/hel.70164
Vancouver
Liu Z, Wu M, Li Y, Huang Q, Mao Z, Diao Y, et al. A Novel Helicobacter pylori Test Based on 16S rRNA Isothermal Amplification Point‐of‐Care Testing of Throat Swab. Helicobacter. 2026;31(5). doi:10.1111/hel.70164