EUVIMEDEUVIMED entdecken
ValidierungsstudieOpen Access · CC BY-NC-ND 4.0

Validation of a discrete electrographic seizure detection algorithm for extended‐duration, reduced‐channel wearable EEG

Tyler J Newton iD, Mitchell A Frankel, Zoë Tosi, Avidor B Kazen, Vamshi K Muvvala, Tobias Loddenkemper, Mark C Spitz, Laura Strom, Daniel Friedman, Mark J Lehmkuhle

Epilepsia · 2025-07 · 66 (7) · 2433-2443

DOI: 10.1111/epi.18365 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Epilepsia, online erschienen am 19.03.2025. Verfasst von Tyler J Newton und 9 weiteren Autor:innen. Beteiligte Einrichtung: Epitel, Salt Lake City, Utah, USA. Hauptthemen laut MEDLINE: Algorithms, Electroencephalography, Wearable Electronic Devices, Seizures. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIH HHS, NINDS NIH HHS, National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

Reduced‐channel wearable electroencephalography (EEG) may overcome the accessibility and patient comfort limitations of traditional ambulatory electrographic seizure monitoring during extended‐duration use. Automated algorithms are necessary for review of extended‐duration reduced‐channel EEG, yet current clinical support software is designed only for full‐montage recordings.

Methods

The performance of a novel automated seizure detection algorithm for reduced‐channel EEG (Epitel) was evaluated in a clinical validation study involving 50 participants (31 with seizures) with diverse demographic and seizure representation.

Results

The algorithm demonstrated an event‐level sensitivity of 86.2% (95% confidence interval [CI] = 79.5%–93.2%) and a false detection rate of .162 per hour (95% CI = .116–.221), which is comparable to the performance of current clinical software for full‐montage EEG. Performance varied by electrographic seizure type, with 91.4% sensitivity for focal evolving to generalized seizures, 86.7% for generalized seizures, and 77.3% for focal seizures. The algorithm maintained robust performance in both pediatric participants aged 6–21 years (83% sensitivity) and adults aged 22+ years (90% sensitivity), as well as in ambulatory (80%) and epilepsy monitoring unit (EMU) monitoring environments (87.5%). The false detection rate in ambulatory monitoring environments (.290 false positive [FP] detections/h), all of which involved pediatric participants, was notably higher than in the EMU (.136 FP/h), indicating an area with clear need for improvement for unrestricted at‐home monitoring. The algorithm's supplemental Confidence metric, designed to engender trust in the algorithm, showed a strong correlation with detection precision. Significance These results suggest that this algorithm can provide crucial support for review of extended‐duration reduced‐channel wearable EEG, enabling electrographic seizure monitoring with no restrictions on a person's daily life.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AlgorithmsElectroencephalographyWearable Electronic DevicesSeizuresHumansChildMaleFemaleAdolescentYoung AdultAdultMiddle AgedMonitoring, AmbulatoryAgedSensitivity and SpecificityDetection Algorithms

Stichwörter der Autor:innen

EEGelectroencephalographymachine learningseizure detectionwearables

Autor:innen und Institutionen

  • Epitel, Salt Lake City, Utah, USA.
  • Division of Epilepsy and Clinical Neurophysiology, Department of Neurology, Boston Children's Hospital & Harvard Medical School, Boston, Massachusetts, USA.
  • Epilepsy Section, Department of Neurology, University of Colorado, Aurora, Colorado, USA.
  • Department of Neurology, New York University Grossman School of Medicine, NYU Langone Health, New York, New York, USA.

ORCID-Profile: Tyler J Newton

Förderung

  • NIH HHS
  • NINDS NIH HHS
  • National Institutes of Health

Zitieren

APA

Newton, T. J., Frankel, M. A., Tosi, Z., Kazen, A. B., Muvvala, V. K., Loddenkemper, T., Spitz, M. C., Strom, L., Friedman, D., & Lehmkuhle, M. J. (2025). Validation of a discrete electrographic seizure detection algorithm for extended‐duration, reduced‐channel wearable EEG. Epilepsia, 66(7), 2433-2443. https://doi.org/10.1111/epi.18365

Vancouver

Newton TJ, Frankel MA, Tosi Z, Kazen AB, Muvvala VK, Loddenkemper T, et al. Validation of a discrete electrographic seizure detection algorithm for extended‐duration, reduced‐channel wearable EEG. Epilepsia. 2025;66(7):2433-2443. doi:10.1111/epi.18365

BibTeX
@article{Newton2025,
  author = {Newton, Tyler J and Frankel, Mitchell A and Tosi, Zoë and Kazen, Avidor B and Muvvala, Vamshi K and Loddenkemper, Tobias and Spitz, Mark C and Strom, Laura and Friedman, Daniel and Lehmkuhle, Mark J},
  title = {Validation of a discrete electrographic seizure detection algorithm for extended‐duration, reduced‐channel wearable EEG},
  journal = {Epilepsia},
  year = {2025},
  volume = {66},
  number = {7},
  pages = {2433--2443},
  doi = {10.1111/epi.18365}
}