EUVIMEDEUVIMED entdecken
ValidierungsstudieOpen Access · CC BY-NC 4.0

Generalizability of electroencephalographic interpretation using artificial intelligence: An external validation study

Daniel Mansilla iD, Jesper Tveit, Harald Aurlien, Tamir Avigdor iD, Victoria Ros-Castello, Alyssa Ho, Chifaou Abdallah, Jean Gotman, Sándor Beniczky iD, Birgit Frauscher iD

Epilepsia · 2024-10 · 65 (10) · 3028-3037

DOI: 10.1111/epi.18082 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Epilepsia, online erschienen am 14.08.2024. Verfasst von Daniel Mansilla und 9 weiteren Autor:innen. Beteiligte Einrichtung: Analytical Neurophysiology Lab, Montreal Neurological Institute and Hospital, Montreal, Quebec, Canada. Hauptthemen laut MEDLINE: Electroencephalography, Artificial Intelligence. Die Arbeit wurde bisher 37-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: CIHR, Duke University, Canadian Institutes of Health Research.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

The automated interpretation of clinical electroencephalograms (EEGs) using artificial intelligence (AI) holds the potential to bridge the treatment gap in resource‐limited settings and reduce the workload at specialized centers. However, to facilitate broad clinical implementation, it is essential to establish generalizability across diverse patient populations and equipment. We assessed whether SCORE‐AI demonstrates diagnostic accuracy comparable to that of experts when applied to a geographically different patient population, recorded with distinct EEG equipment and technical settings.

Methods

We assessed the diagnostic accuracy of a “fixed‐and‐frozen” AI model, using an independent dataset and external gold standard, and benchmarked it against three experts blinded to all other data. The dataset comprised 50% normal and 50% abnormal routine EEGs, equally distributed among the four major classes of EEG abnormalities (focal epileptiform, generalized epileptiform, focal nonepileptiform, and diffuse nonepileptiform). To assess diagnostic accuracy, we computed sensitivity, specificity, and accuracy of the AI model and the experts against the external gold standard.

Results

We analyzed EEGs from 104 patients (64 females, median age = 38.6 [range = 16–91] years). SCORE‐AI performed equally well compared to the experts, with an overall accuracy of 92% (95% confidence interval [CI] = 90%–94%) versus 94% (95% CI = 92%–96%). There was no significant difference between SCORE‐AI and the experts for any metric or category. SCORE‐AI performed well independently of the vigilance state (false classification during awake: 5/41 [12.2%], false classification during sleep: 2/11 [18.2%]; p = .63) and normal variants (false classification in presence of normal variants: 4/14 [28.6%], false classification in absence of normal variants: 3/38 [7.9%]; p = .07). Significance SCORE‐AI achieved diagnostic performance equal to human experts in an EEG dataset independent of the development dataset, in a geographically distinct patient population, recorded with different equipment and technical settings than the development dataset.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectroencephalographyArtificial IntelligenceHumansFemaleMaleAdultMiddle AgedAgedAdolescentYoung AdultAged, 80 and overEpilepsySensitivity and SpecificityReproducibility of Results

Stichwörter der Autor:innen

automatic EEG analysisepilepsynormal variantsroutine EEGsleep

Autor:innen und Institutionen

  • Analytical Neurophysiology Lab, Montreal Neurological Institute and Hospital, Montreal, Quebec, Canada.
  • Neurophysiology Unit, Institute of Neurosurgery Dr. Asenjo, Santiago, Chile.
  • Holberg EEG, Bergen, Norway.
  • Department of Neurology, Duke University Medical Center, Durham, North Carolina, USA.
  • Epilepsy Unit, Department of Neurology, Hospital de la Santa Creu i Sant Pau, Barcelona, Spain.
  • Montreal Neurological Institute and Hospital, McGill University, Montreal, Quebec, Canada.

ORCID-Profile: Daniel Mansilla, Tamir Avigdor, Sándor Beniczky, Birgit Frauscher

Förderung

  • CIHR
  • Duke University
  • Canadian Institutes of Health Research

Zitieren

APA

Mansilla, D., Tveit, J., Aurlien, H., Avigdor, T., Ros-Castello, V., Ho, A., Abdallah, C., Gotman, J., Beniczky, S., & Frauscher, B. (2024). Generalizability of electroencephalographic interpretation using artificial intelligence: An external validation study. Epilepsia, 65(10), 3028-3037. https://doi.org/10.1111/epi.18082

Vancouver

Mansilla D, Tveit J, Aurlien H, Avigdor T, Ros-Castello V, Ho A, et al. Generalizability of electroencephalographic interpretation using artificial intelligence: An external validation study. Epilepsia. 2024;65(10):3028-3037. doi:10.1111/epi.18082

BibTeX
@article{Mansilla2024,
  author = {Mansilla, Daniel and Tveit, Jesper and Aurlien, Harald and Avigdor, Tamir and Ros-Castello, Victoria and Ho, Alyssa and Abdallah, Chifaou and Gotman, Jean and Beniczky, Sándor and Frauscher, Birgit},
  title = {Generalizability of electroencephalographic interpretation using artificial intelligence: An external validation study},
  journal = {Epilepsia},
  year = {2024},
  volume = {65},
  number = {10},
  pages = {3028--3037},
  doi = {10.1111/epi.18082}
}