Performance of two‐dimensional shear wave elastography and transient elastography compared to liver biopsy for staging of liver fibrosis
European Journal of Clinical Investigation · 2023
DOI: 10.1111/eci.13980 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Journal of Clinical Investigation, 2023. Verfasst von Audrey Kovatsch und 16 weiteren Autor:innen. Beteiligte Einrichtung: Department of Gastroenterology and Hepatology, University Hospital Zurich, Zurich, Switzerland. Hauptthemen laut MEDLINE: Elasticity Imaging Techniques, Liver Diseases. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
Staging of liver fibrosis traditionally relied on liver histology; however, transient elastography (TE) and more recently two‐dimensional shear wave elastography (2D‐SWE) evolved to noninvasive alternatives. Hence, we evaluated the diagnostic accuracy of 2D‐SWE assessed by the Canon Aplio i800 ultrasound system using liver biopsy as reference and compared the performance to TE.
Methods
In total, 108 adult patients with chronic liver disease undergoing liver biopsy, 2D‐SWE and TE were enrolled prospectively at the University Hospital Zurich. Diagnostic accuracies were evaluated using the area under the receiver operating characteristic (AUROC) analysis, and optimal cut‐off values by Youden's index.
Results
Diagnostic accuracy of 2D‐SWE was good for significant (≥F2; AUROC 85.2%, 95% confidence interval (95%CI):76.2–91.2%) as well as severe fibrosis (≥F3; AUROC 86.8%, 95%CI: 78.1–92.4%) and excellent for cirrhosis (AUROC 95.6%, 95%CI: 89.9–98.1%), compared to histology. TE performed equally well (significant fibrosis: 87.5%, 95%CI: 77.7–93.3%; severe fibrosis: 89.7%, 95%CI: 82.0–94.3%; cirrhosis: 96%, 95%CI: 90.4–98.4%), and accuracy was not statistically different to 2D‐SWE. 2D‐SWE optimal cut‐off values were 6.5, 9.8 and 13.1 kPa for significant fibrosis, severe fibrosis and cirrhosis, respectively.
Conclusions
Performance of 2D‐SWE was good to excellent and well comparable with TE, supporting the application of this 2D‐SWE system in the diagnostic workup of chronic liver disease.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Gastroenterology and Hepatology, University Hospital Zurich, Zurich, Switzerland.
- Department of Pathology, University Hospital Zurich, Zurich, Switzerland.
- Department of Visceral Surgery and Medicine, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
- Institute of Epidemiology, Biostatistics and Prevention, University of Zurich, Zurich, Switzerland.
- Department of Radiology, Brigham and Women's Hospital and Harvard Medical School, Boston, Massachusetts, USA.
- Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, Zurich, Switzerland.
ORCID-Profile: Christoph Jüngst
Zitieren
APA
Kovatsch, A., Honcharova-Biletska, H., Segna, D., Steigmiller, K., Blümel, S., Deibel, R. A., Kühlewindt, T., Leinenkugel, G., Müller, S., Furrer, E., Schawkat, K., Reiner, C. S., Weber, A., Müllhaupt, B., Scharl, M., Gubler, C., & Jüngst, C. (2023). Performance of two‐dimensional shear wave elastography and transient elastography compared to liver biopsy for staging of liver fibrosis. European Journal of Clinical Investigation. https://doi.org/10.1111/eci.13980
Vancouver
Kovatsch A, Honcharova-Biletska H, Segna D, Steigmiller K, Blümel S, Deibel RA, et al. Performance of two‐dimensional shear wave elastography and transient elastography compared to liver biopsy for staging of liver fibrosis. European Journal of Clinical Investigation. 2023. doi:10.1111/eci.13980
BibTeX
@article{Kovatsch2023,
author = {Kovatsch, Audrey and Honcharova-Biletska, Hanna and Segna, Daniel and Steigmiller, Klaus and Blümel, Sena and Deibel, Rudolf A and Kühlewindt, Tobias and Leinenkugel, Georg and Müller, Sandra and Furrer, Eva and Schawkat, Khoschy and Reiner, Cäcilia S and Weber, Achim and Müllhaupt, Beat and Scharl, Michael and Gubler, Christoph and Jüngst, Christoph},
title = {Performance of two‐dimensional shear wave elastography and transient elastography compared to liver biopsy for staging of liver fibrosis},
journal = {European Journal of Clinical Investigation},
year = {2023},
doi = {10.1111/eci.13980}
}
EUVIMED