EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY-NC 4.0

Comparison of Diluted Insulin (U10) Versus Standard (U100) Insulin in Automated Insulin Delivery Systems: Reducing Time Below Range in Young Children

Heidi Falk iD, Ada Aaltio, Mari-Anne Pulkkinen iD, Tero Varimo, Anna-Kaisa Tuomaala iD

Diabetes, Obesity and Metabolism · 2026

DOI: 10.1111/dom.71102 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Diabetes, Obesity and Metabolism, 2026. Verfasst von Heidi Falk und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pediatrics and Adolescence, University of Helsinki and Helsinki University Hospital, Helsinki, Uusimaa, Finland. Hauptthemen laut MEDLINE: Diabetes Mellitus, Type 1, Insulin Infusion Systems, Insulin, Hypoglycemic Agents. Die Arbeit nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Lastentautien Tutkimussäätiö.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Treating young children with Type 1 diabetes (CwT1D) is challenging due to their very small need for insulin. We investigated the safety of diluted insulin (U10) in the AID system and how its use affected glycaemic outcomes compared to standard insulin.

Methods

Patients aged 0–4 years who initiated treatment with diluted insulin in an AID system during the years 2020–2025 were identified from the diabetes registry ( N = 19). They had a maximum daily insulin dose of ≤ 6 units. An age and gender‐matched control group ( N = 27) using standard insulin (U100) in an AID system was determined. Longitudinal CGM outcomes were analysed within the treatment group from the start of diluted insulin and between groups from diagnosis over 12 months to assess time‐by‐treatment effects.

Results

After 3 months of diluted insulin use, TBR < 3.9 and TBR < 3.0 decreased significantly (from 4.6% ± 4.1% to 2.7% ± 2.5%, p = 0.008; from 1.2% ± 1.8% to 0.4% ± 0.6%, p = 0.01, respectively) and remained low through 12 months. Compared with controls, children using diluted insulin had lower TBR < 3.9 mmol/L at later follow‐up visits. In the HbA1c‐adjusted mixed‐effects model, the between‐group difference in change from 3 months was −1.85 percentage points at 9 months (95% CI: −3.26 to −0.44, p = 0.010) and −1.52 percentage points at 12 months (95% CI: −2.92 to −0.13, p = 0.032). No significant between‐group changes in other CGM measures over time were observed.

Conclusion

The use of diluted insulin in young CwT1D, who require low insulin doses and use an automated insulin delivery system, reduces time below range with no evidence of increased hyperglycaemia.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Diabetes Mellitus, Type 1Insulin Infusion SystemsInsulinHypoglycemic AgentsHumansFemaleMaleChild, PreschoolInfantBlood GlucoseGlycated HemoglobinContinuous Glucose MonitoringHypoglycemiaInfant, NewbornTreatment OutcomeRegistries

Stichwörter der Autor:innen

AID settingsAID systemchildren with Type 1 diabetesdiluted insulin

Substanzen

InsulinHypoglycemic AgentsBlood GlucoseGlycated Hemoglobin

Autor:innen und Institutionen

  • Department of Pediatrics and Adolescence, University of Helsinki and Helsinki University Hospital, Helsinki, Uusimaa, Finland.

ORCID-Profile: Heidi Falk, Mari-Anne Pulkkinen, Anna-Kaisa Tuomaala

Förderung

  • Lastentautien Tutkimussäätiö

Zitieren

APA

Falk, H., Aaltio, A., Pulkkinen, M. A., Varimo, T., & Tuomaala, A. K. (2026). Comparison of Diluted Insulin (U10) Versus Standard (U100) Insulin in Automated Insulin Delivery Systems: Reducing Time Below Range in Young Children. Diabetes, Obesity and Metabolism. https://doi.org/10.1111/dom.71102

Vancouver

Falk H, Aaltio A, Pulkkinen MA, Varimo T, Tuomaala AK. Comparison of Diluted Insulin (U10) Versus Standard (U100) Insulin in Automated Insulin Delivery Systems: Reducing Time Below Range in Young Children. Diabetes, Obesity and Metabolism. 2026. doi:10.1111/dom.71102

BibTeX
@article{Falk2026,
  author = {Falk, Heidi and Aaltio, Ada and Pulkkinen, Mari-Anne and Varimo, Tero and Tuomaala, Anna-Kaisa},
  title = {Comparison of Diluted Insulin (U10) Versus Standard (U100) Insulin in Automated Insulin Delivery Systems: Reducing Time Below Range in Young Children},
  journal = {Diabetes, Obesity and Metabolism},
  year = {2026},
  doi = {10.1111/dom.71102}
}
Kostenlos starten →