EUVIMEDEUVIMED entdecken
Crossref

Bench‐Top Evaluation of Pancreatobiliary Guidewires for Endoscopic Ultrasound‐Guided Drainage

Takehiko Koga iD, Yusuke Ishida iD, Naoaki Tsuchiya iD, Takanori Kitaguchi, Keisuke Matsumoto, Makoto Fukuyama, Tatsuro Nii, Norihiro Kojima iD, Sara Haruki, Hiromichi Iwaya iD, Keiji Yokoyama iD, Shinya Ashizuka iD und 15 weitere

Digestive Endoscopy · 2026-11 · 38 (11)

DOI: 10.1111/den.70348 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Digestive Endoscopy, online erschienen am 18.11.2026. Verfasst von Takehiko Koga und 26 weiteren Autor:innen. Beteiligte Einrichtung: Department of Gastroenterology and Medicine, Fukuoka University Faculty of Medicine, Fukuoka, Japan. Hauptthemen laut MEDLINE: Drainage, Endosonography, Ultrasonography, Interventional. Die Arbeit nennt 21 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Guidewire manipulation is a key determinant of procedural success in endoscopic ultrasound‐guided drainage/anastomosis (EUS‐D/A). However, most guidewires were originally designed for endoscopic retrograde cholangiopancreatography, and objective selection criteria for EUS‐D/A have not been established. This study aimed to clarify the relationships between the structural features and the mechanical performance of guidewires used in EUS‐D/A.

Methods

Twenty‐one commercially available guidewires (0.018–0.035 in.) were evaluated. Structural parameters, including the polyurethane length, tip deflection height, tip and shaft core wire diameters, outer diameter, covering texture, and covering adhesion, were measured. Mechanical performance was assessed using bench‐top experiments including insertion and withdrawal forces, torque response and whip motion within an endoscopic ultrasound‐guided tissue acquisition (EUS‐TA) needle, bending force, tip load, loop performance, and the turning point of loop formation. Correlations between the structural and performance parameters were analyzed.

Results

Core wire diameter exhibited a strong positive correlation with the bending force, which was strongly correlated with the insertion and withdrawal forces. Tip core diameter was strongly correlated with the tip load, and loop formation consistently occurred at the initiation of the steeply tapered core wire segment. Torque response rate was positively correlated with the core wire diameter and core‐to‐outer diameter ratio. Guidewires with a core wire diameter ≥ 0.52 mm and polytetrafluoroethylene‐based gentle‐pitch uneven coverings demonstrated a higher torque response rate.

Conclusions

This bench‐top study suggests that structural parameters related to the shaft and tip are associated with guidewire behavior during EUS‐D/A.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

DrainageEndosonographyUltrasonography, InterventionalHumansEquipment Design

Stichwörter der Autor:innen

biliary tractendoscopic retrograde cholangiopancreatography (ERCP)endoscopic ultrasound‐guided biliary drainage (EUS‐BD)endoscopic ultrasound‐guided pancreatic duct drainage (EUS‐PDD)guidewire

Autor:innen und Institutionen

  • Department of Gastroenterology and Medicine, Fukuoka University Faculty of Medicine, Fukuoka, Japan.
  • Department of Gastroenterology, Taragi Municipal Hospital, Kumamoto, Japan.
  • Internal Medicine, Taragi Municipal Hospital, Kumamoto, Japan.
  • Iwaya Internal Medicine and Endoscopy Clinic, Kagoshima, Japan.

ORCID-Profile: Takehiko Koga, Yusuke Ishida, Naoaki Tsuchiya, Norihiro Kojima, Hiromichi Iwaya, Keiji Yokoyama, Shinya Ashizuka, Takashi Tanaka, Kazuhide Takata, Fumihito Hirai

Zitieren

APA

Koga, T., Ishida, Y., Tsuchiya, N., Kitaguchi, T., Matsumoto, K., Fukuyama, M., Nii, T., Kojima, N., Haruki, S., Iwaya, H., Yokoyama, K., Ashizuka, S., Funakoshi, S., Tanaka, T., Takata, K., Kuno, N., Sakisaka, H., Shibata, K., Matsuoka, H., …, & Hirai, F. (2026). Bench‐Top Evaluation of Pancreatobiliary Guidewires for Endoscopic Ultrasound‐Guided Drainage. Digestive Endoscopy, 38(11). https://doi.org/10.1111/den.70348

Vancouver

Koga T, Ishida Y, Tsuchiya N, Kitaguchi T, Matsumoto K, Fukuyama M, et al. Bench‐Top Evaluation of Pancreatobiliary Guidewires for Endoscopic Ultrasound‐Guided Drainage. Digestive Endoscopy. 2026;38(11). doi:10.1111/den.70348

BibTeX
@article{Koga2026,
  author = {Koga, Takehiko and Ishida, Yusuke and Tsuchiya, Naoaki and Kitaguchi, Takanori and Matsumoto, Keisuke and Fukuyama, Makoto and Nii, Tatsuro and Kojima, Norihiro and Haruki, Sara and Iwaya, Hiromichi and Yokoyama, Keiji and Ashizuka, Shinya and Funakoshi, Sadahiro and Tanaka, Takashi and Takata, Kazuhide and Kuno, Nobuaki and Sakisaka, Hideto and Shibata, Kumiko and Matsuoka, Hiroki and Uchida, Yotaro and Miyayama, Takashi and Shibata, Mamoru and Nomaru, Ryohei and Tokushige, Hiroaki and Ko, Yi-Ling and Shimoji, Yasuharu and Hirai, Fumihito},
  title = {Bench‐Top Evaluation of Pancreatobiliary Guidewires for Endoscopic Ultrasound‐Guided Drainage},
  journal = {Digestive Endoscopy},
  year = {2026},
  volume = {38},
  number = {11},
  doi = {10.1111/den.70348}
}
Kostenlos starten →