Evaluation of a novel disposable esophagogastroduodenoscopy system in emergency, bedside, and intraoperative settings: Pilot study (with videos)

Ze-Long Han, Bi-Tao Lin, Zhen-Jiang Wang, Xiang Chen, Yi-Yi Xi, Jun-Fen Wang, Wei-Guang Qiao, Ying Huang, Zhi-Zhao Lin, Shao-Hui Huang, Tiffany Y Chua, Si-De Liu iD und 1 weitere

Digestive Endoscopy · 2023

DOI: 10.1111/den.14548 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Digestive Endoscopy, 2023. Verfasst von Ze-Long Han und 12 weiteren Autor:innen. Beteiligte Einrichtung: Guangdong Provincial Key Laboratory of Gastroenterology, Department of Gastroenterology, Nanfang Hospital, Southern Medical University, Guangzhou, Ch…. Hauptthemen laut MEDLINE: Endoscopy, Digestive System, Endoscopy. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 17 Literaturangaben. Gefördert von: Guangdong Science and Technology Plan Project, National Natural Science Funds of China, President Foundation of Nanfang Hospital, Southern Medical University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

The disposable esophagogastroduodenoscopy (EGD) system is a novel endoscopic device which is highly portable and is designed to eliminate the risk of cross‐infection caused by reusable EGD. This study aimed to investigate the feasibility and safety of disposable EGD in emergency, bedside, and intraoperative settings.

Methods

This was a prospective, single‐center, noncomparative study. Disposable EGD was used for emergency, bedside, and intraoperative endoscopies in 30 patients. The primary end‐point was the technical success rate of the disposable EGD. Secondary end‐points included technical performance indicators including clinical operability, image quality score, procedure time, the incidence of device malfunction and/or failure, and the incidence of adverse events.

Results

A total of 30 patients underwent diagnosis and/or treatment with disposable EGD. Therapeutic EGD was performed on 13/30 patients, including hemostasis ( n = 3), foreign body retrieval ( n = 6), nasoenteric tube placement ( n = 3), and percutaneous endoscopic gastrostomy ( n = 1). The technical success rate was 100%: all procedures and indicated interventions were completed without changing to a conventional upper endoscope. The mean image quality score obtained immediately after procedure completion was 3.72 ± 0.56. The mean (± SD) procedure time was 7.4 (± 7.6) min. There were no device malfunctions or failures, device‐related adverse events, or overall adverse events.

Conclusion

The disposable EGD may be a feasible alternative to the traditional EGD in emergency, bedside, and intraoperative settings. Preliminary data show that it is a safe and effective tool for diagnosis and treatment in emergency and bedside upper gastrointestinal cases.

Trial registration

Chinese Clinical Trial Registry (Trial ID: ChiCTR2100051452, https://www.chictr.org.cn/showprojen.aspx?proj=134284 ).

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Endoscopy, Digestive SystemEndoscopyHumansPilot ProjectsProspective StudiesIntubation, Gastrointestinal

Stichwörter der Autor:innen

bedside and intraoperative settingsdisposable esophagogastroduodenoscopyemergency endoscopy

Autor:innen und Institutionen

  • Guangdong Provincial Key Laboratory of Gastroenterology, Department of Gastroenterology, Nanfang Hospital, Southern Medical University, Guangzhou, China.
  • Department of Gastroenterology, Zhuhai People's Hospital (Zhuhai Hospital Affiliated to Jinan University), Zhuhai, China.
  • Prince of Wales Hospital, Chinese University of Hong Kong, Shatin, Hong Kong SAR, China.
  • Pazhou Lab, Guangzhou, China.

ORCID-Profile: Si-De Liu

Förderung

  • Guangdong Science and Technology Plan Project
  • National Natural Science Funds of China
  • President Foundation of Nanfang Hospital, Southern Medical University

Zitieren

APA

Han, Z. L., Lin, B. T., Wang, Z. J., Chen, X., Xi, Y. Y., Wang, J. F., Qiao, W. G., Huang, Y., Lin, Z. Z., Huang, S. H., Chua, T. Y., Liu, S. D., & Luo, X. B. (2023). Evaluation of a novel disposable esophagogastroduodenoscopy system in emergency, bedside, and intraoperative settings: Pilot study (with videos). Digestive Endoscopy. https://doi.org/10.1111/den.14548

Vancouver

Han ZL, Lin BT, Wang ZJ, Chen X, Xi YY, Wang JF, et al. Evaluation of a novel disposable esophagogastroduodenoscopy system in emergency, bedside, and intraoperative settings: Pilot study (with videos). Digestive Endoscopy. 2023. doi:10.1111/den.14548

BibTeX
@article{Han2023,
  author = {Han, Ze-Long and Lin, Bi-Tao and Wang, Zhen-Jiang and Chen, Xiang and Xi, Yi-Yi and Wang, Jun-Fen and Qiao, Wei-Guang and Huang, Ying and Lin, Zhi-Zhao and Huang, Shao-Hui and Chua, Tiffany Y and Liu, Si-De and Luo, Xiao-Bei},
  title = {Evaluation of a novel disposable esophagogastroduodenoscopy system in emergency, bedside, and intraoperative settings: Pilot study (with videos)},
  journal = {Digestive Endoscopy},
  year = {2023},
  doi = {10.1111/den.14548}
}
Kostenlos starten →