Device‐related differences and interchangeability of forced expiratory volume in 1 s measurements: A comparison of three spirometers
Clinical Physiology and Functional Imaging · 2026-11 · 46 (6)
DOI: 10.1111/cpf.70105 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Clinical Physiology and Functional Imaging, online erschienen am 05.11.2026. Verfasst von Leon L Csonka und 3 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Medicine and Health Technology, Tampere University, Tampere, Finland. Hauptthemen laut MEDLINE: Spirometry, Lung. Die Arbeit nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Tampereen Tuberkuloosisäätiö, Suomen Tuberkuloosin Vastustamisyhdistyksen Säätiö, Hengityssairauksien Tutkimussäätiö.
Abstract (Originalsprache)
Background
Forced expiratory volume in 1 s (FEV 1 ) obtained using different types of laboratory spirometers and handheld microspirometers is often used interchangeably in research, despite substantial differences between devices. The aim of this study was to assess device‐related differences and the interchangeability of FEV 1 measurements among three commonly used spirometers.
Methods
We recruited 20 healthy adults experienced in performing spirometry. Each participant performed three measurements with each of the three spirometers: the handheld microspirometers Medikro Duo and MIR Spirobank, and the laboratory spirometer Vyntus Pneumo.
Results
In the linear mixed‐effects model, there was a significant main effect of device on FEV 1 (F(2,38) = 25.25, p < 0.001), indicating that FEV 1 values differed between devices. Spirobank yielded the highest FEV 1 values, whereas the Medikro Duo yielded the lowest. The mean difference in FEV 1 between Medikro Duo and Spirobank was −0.25 L (−0.34 to −0.16, p < 0.001), and between Medikro Duo and Vyntus Pneumo −0.16 L (−0.25 to −0.07, p < 0.001). The comparison between Spirobank and Vyntus Pneumo showed a smaller difference of 0.09 L (0.00 to 0.18), which narrowly missed statistical significance ( p = 0.050). Bland–Altman plots showed wide limits of agreement between spirometers, suggesting poor interchangeability.
Conclusions
Systematic bias and varying agreement between the laboratory spirometer and the handheld microspirometers indicate that FEV 1 measurements obtained with different devices should not be assumed to be fully interchangeable. This does not imply that handheld microspirometry is inferior, but rather that device‐specific characteristics must be considered when interpreting and comparing FEV 1 values.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Faculty of Medicine and Health Technology, Tampere University, Tampere, Finland.
- Department of Clinical Physiology and Nuclear Medicine, Tampere University Hospital, Tampere, Finland.
- Allergy Centre, Tampere University Hospital, Tampere, Finland.
ORCID-Profile: Leon L Csonka, Lauri Lehtimäki
Förderung
- Tampereen Tuberkuloosisäätiö
- Suomen Tuberkuloosin Vastustamisyhdistyksen Säätiö
- Hengityssairauksien Tutkimussäätiö
- Tays tukisäätiö
- Valtion tutkimusrahoitus
Zitieren
APA
Csonka, L. L., Tikkakoski, A., Karjalainen, J., & Lehtimäki, L. (2026). Device‐related differences and interchangeability of forced expiratory volume in 1 s measurements: A comparison of three spirometers. Clinical Physiology and Functional Imaging, 46(6). https://doi.org/10.1111/cpf.70105
Vancouver
Csonka LL, Tikkakoski A, Karjalainen J, Lehtimäki L. Device‐related differences and interchangeability of forced expiratory volume in 1 s measurements: A comparison of three spirometers. Clinical Physiology and Functional Imaging. 2026;46(6). doi:10.1111/cpf.70105
BibTeX
@article{Csonka2026,
author = {Csonka, Leon L and Tikkakoski, Antti and Karjalainen, Jussi and Lehtimäki, Lauri},
title = {Device‐related differences and interchangeability of forced expiratory volume in 1 s measurements: A comparison of three spirometers},
journal = {Clinical Physiology and Functional Imaging},
year = {2026},
volume = {46},
number = {6},
doi = {10.1111/cpf.70105}
}
EUVIMED