Effect of device and expiratory maneuver technique on peak expiratory flow measurements
Clinical Physiology and Functional Imaging · 2026-05 · 46 (3)
DOI: 10.1111/cpf.70067 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Clinical Physiology and Functional Imaging, online erschienen am 15.05.2026. Verfasst von Leon L Csonka und 3 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Medicine and Health Technology, Tampere University, Tampere, Finland. Hauptthemen laut MEDLINE: Spirometry, Lung, Exhalation, Flowmeters. Die Arbeit nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Tampereen Tuberkuloosisäätiö, Suomen Tuberkuloosin Vastustamisyhdistyksen Säätiö, Hengityssairauksien Tutkimussäätiö.
Abstract (Originalsprache)
Background
Peak expiratory flow (PEF) obtained from dedicated PEF meters and various types of spirometers is often used interchangeably in research, despite differences in both devices and expiratory maneuver technique. The aim of this study was to assess the effects of expiratory maneuver type and measurement device on PEF.
Methods
We recruited 20 healthy adults experienced in performing spirometry. Each subject performed three measurements using two expiratory maneuvers (a short, explosive maneuver and a long, spirometry‐style maneuver) with four devices: the handheld microspirometers Medikro Duo and MIR Spirobank, the laboratory spirometer Vyntus Pneumo, and the Mini‐Wright PEF meter.
Results
PEF differed significantly between devices (F(3,133) = 171.8, p < 0.001), but not between short and long expiratory maneuver techniques (F(1,133) = 2.24, p = 0.137). There was no significant interaction between device and technique (F(3,133) = 0.72, p = 0.543). Compared with the Mini‐Wright, Spirobank yielded PEF values that were on average 87.1 L/min higher (74.6–99.7, p < 0.001), and Vyntus Pneumo 26.5 L/min higher (14.0–39.1, p < 0.001), whereas Medikro did not differ significantly from the Mini‐Wright (mean difference −9.5 L/min, −22.1–3.1, p = 0.269). The pairwise differences between the three spirometers were all statistically significant (all p < 0.001).
Conclusions
Because short and long expiratory maneuver techniques yield similar results on PEF, future studies of home spirometry may compare the diagnostic performance of PEF and forced expiratory volume in 1 s (FEV 1 ) using the same long, spirometry‐style expiratory maneuvers. However, comparison of absolute PEF values between devices must be done cautiously.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Faculty of Medicine and Health Technology, Tampere University, Tampere, Finland.
- Department of Clinical Physiology and Nuclear Medicine, Tampere University Hospital, Tampere, Finland.
- Allergy Centre, Tampere University Hospital, Tampere, Finland.
ORCID-Profile: Leon L Csonka, Lauri Lehtimäki
Förderung
- Tampereen Tuberkuloosisäätiö
- Suomen Tuberkuloosin Vastustamisyhdistyksen Säätiö
- Hengityssairauksien Tutkimussäätiö
- Tays tukisäätiö
- Valtion tutkimusrahoitus
Zitieren
APA
Csonka, L. L., Tikkakoski, A., Karjalainen, J., & Lehtimäki, L. (2026). Effect of device and expiratory maneuver technique on peak expiratory flow measurements. Clinical Physiology and Functional Imaging, 46(3). https://doi.org/10.1111/cpf.70067
Vancouver
Csonka LL, Tikkakoski A, Karjalainen J, Lehtimäki L. Effect of device and expiratory maneuver technique on peak expiratory flow measurements. Clinical Physiology and Functional Imaging. 2026;46(3). doi:10.1111/cpf.70067
BibTeX
@article{Csonka2026,
author = {Csonka, Leon L and Tikkakoski, Antti and Karjalainen, Jussi and Lehtimäki, Lauri},
title = {Effect of device and expiratory maneuver technique on peak expiratory flow measurements},
journal = {Clinical Physiology and Functional Imaging},
year = {2026},
volume = {46},
number = {3},
doi = {10.1111/cpf.70067}
}
EUVIMED