EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Adjustable Vocal Implant System ( VOIS ) for Unilateral Vocal Fold Paralysis: Short‐Term Outcomes From a 12‐Patient Case Series

Sam Kebbe iD, Michael Keane, Chuanyu Gao, Yadsan Devabalan iD, Guan-Min Ho, Rohan Bidaye

Clinical Otolaryngology · 2026

DOI: 10.1111/coa.70142 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Otolaryngology, 2026. Verfasst von Sam Kebbe und 5 weiteren Autor:innen. Beteiligte Einrichtung: Leicester Medical School, George Davies Centre, Leicester, UK. Hauptthemen laut MEDLINE: Vocal Cord Paralysis, Voice Quality, Laryngoplasty, Prosthesis Implantation. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Unilateral vocal cord palsy (UVCP) significantly affects phonation and swallowing. While injection laryngoplasty and medialisation thyroplasty are effective, both may require revision. Vocal Implant System (VOIS; APrevent Biotech GmbH) is an adjustable titanium‐silicone thyroplasty implant that allows postoperative modification, potentially reducing the need for further surgery.

Methods

Twelve patients, seven male and five female, underwent VOIS implantation. Voice outcomes were assessed pre‐ and postoperatively using the GRBAS scale and the Voice Handicap Index‐10 (VHI‐10).

Results

The composite GRBAS score improved significantly from 9.0 ± 1.87 to 3.75 ± 2.49 ( p = 0.0024). The VHI‐10 decreased from 25.75 ± 9.36 to 14.92 ± 9.55 ( p = 0.0093). No intra‐ or postoperative complications occurred. Two patients required postoperative implant volume adjustments—one performed in clinic and one under local anaesthetic in theatre.

Conclusion

The VOIS implant thyroplasty demonstrated significant improvement in both perceptual and patient‐reported voice outcomes with a favourable safety profile. Larger, long‐term studies are warranted to confirm its comparative efficacy and durability.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Vocal Cord ParalysisVoice QualityLaryngoplastyProsthesis ImplantationLarynx, ArtificialHumansFemaleMaleMiddle AgedTreatment OutcomeAdultProsthesis DesignAgedRetrospective Studies

Stichwörter der Autor:innen

adjustable VOIS implantmedialisation thyroplastypatient reported outcome measurephonation outcomesunilateral vocal cord palsy

Autor:innen und Institutionen

  • Leicester Medical School, George Davies Centre, Leicester, UK.
  • Northampton General Hospital, Northampton, UK.
  • Mackay Memorial Hospital, Taipei, Taiwan.
  • Yomin ENT & Pediatric Clinic, New Taipei City, Taiwan.
  • APrevent Medical, Taipei, Taiwan.

ORCID-Profile: Sam Kebbe, Yadsan Devabalan

Zitieren

APA

Kebbe, S., Keane, M., Gao, C., Devabalan, Y., Ho, G. M., & Bidaye, R. (2026). Adjustable Vocal Implant System ( VOIS ) for Unilateral Vocal Fold Paralysis: Short‐Term Outcomes From a 12‐Patient Case Series. Clinical Otolaryngology. https://doi.org/10.1111/coa.70142

Vancouver

Kebbe S, Keane M, Gao C, Devabalan Y, Ho GM, Bidaye R. Adjustable Vocal Implant System ( VOIS ) for Unilateral Vocal Fold Paralysis: Short‐Term Outcomes From a 12‐Patient Case Series. Clinical Otolaryngology. 2026. doi:10.1111/coa.70142

BibTeX
@article{Kebbe2026,
  author = {Kebbe, Sam and Keane, Michael and Gao, Chuanyu and Devabalan, Yadsan and Ho, Guan-Min and Bidaye, Rohan},
  title = {Adjustable Vocal Implant System ( VOIS ) for Unilateral Vocal Fold Paralysis: Short‐Term Outcomes From a 12‐Patient Case Series},
  journal = {Clinical Otolaryngology},
  year = {2026},
  doi = {10.1111/coa.70142}
}
Kostenlos starten →