EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Investigations of Intraoral Scan Body Extensions. Effect of Design on the Accuracy, Usability, and Feasibility for Digital Full‐Arch Implant Scans In Vitro

Julian Hessel Baranowski iD, Michael Braian, Victoria Franke Stenport iD, Ann Wennerberg

Clinical Oral Implants Research · 2026

DOI: 10.1111/clr.70158 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Oral Implants Research, 2026. Verfasst von Julian Hessel Baranowski und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Prosthodontics/Dental Materials Science, The Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden. Hauptthemen laut MEDLINE: Dental Impression Technique, Computer-Aided Design, Dental Implants, Dental Prosthesis Design. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Region Östergötland, Västra Götalandsregionen.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To evaluate the effect of different scan body extension designs on the accuracy, usability, and feasibility of digital full‐arch implant impressions in an edentulous mandibular model. Accuracy was expressed as trueness and precision, usability as scan time, and feasibility as the number of total scan failures. Material and Methods An edentulous mandibular model with four implants was fabricated with minimal surface morphology. Five scan body extension configurations were tested: no extensions (control), commercially available extensions, conventional impression copings used as extensions, and two morphological 3D‐printed extensions oriented between adjacent implants or toward the center of the arch. Two intraoral scanners were used. For each scanner and configuration, 30 scans were acquired. Inter‐implant distances were compared against coordinate measuring machine reference values. Trueness was assessed as mean absolute deviation, precision as standard deviation, and differences were analyzed with non‐parametric tests.

Results

The highly morphological scan body extensions produced the most consistent improvements. Combined across scanners, trueness improved by 40–57 μm relative to control for the longest inter‐implant span. Precision improved significantly for the center‐oriented configuration. All scan body extension techniques reduced scan time by 18%–41% and reduced scan failures from 13% (control) to 3% for the morphological designs. Improvements were most pronounced for longer spans and varied between scanners.

Conclusions

Highly morphological 3D‐printed scan body extensions significantly improved the accuracy, usability, and feasibility of complete‐arch digital implant impressions compared to an unmodified control workflow. Clinical validation is required to confirm these findings for broader use.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Dental Impression TechniqueComputer-Aided DesignDental ImplantsDental Prosthesis DesignHumansFeasibility StudiesJaw, EdentulousPrinting, Three-DimensionalIn Vitro TechniquesModels, DentalImaging, Three-DimensionalMandible

Stichwörter der Autor:innen

computer‐aided designdental implantsdental impression techniquedental prosthesisdimensional measurement accuracyedentulousimplant‐supportedjawprintingthree‐dimensional

Substanzen

Dental Implants

Autor:innen und Institutionen

  • Department of Prosthodontics/Dental Materials Science, The Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden.
  • Center for Oral Rehabilitation, Public Dental Service Östergötland, Linköping, Sweden.
  • Swedish Organisation for Computer Aided Digital Dentistry, SWECADD, Malmö, Sweden.
  • The Brånemark Clinic, Gothenburg, Sweden.

ORCID-Profile: Julian Hessel Baranowski, Victoria Franke Stenport

Förderung

  • Region Östergötland
  • Västra Götalandsregionen

Zitieren

APA

Baranowski, J. H., Braian, M., Stenport, V. F., & Wennerberg, A. (2026). Investigations of Intraoral Scan Body Extensions. Effect of Design on the Accuracy, Usability, and Feasibility for Digital Full‐Arch Implant Scans In Vitro. Clinical Oral Implants Research. https://doi.org/10.1111/clr.70158

Vancouver

Baranowski JH, Braian M, Stenport VF, Wennerberg A. Investigations of Intraoral Scan Body Extensions. Effect of Design on the Accuracy, Usability, and Feasibility for Digital Full‐Arch Implant Scans In Vitro. Clinical Oral Implants Research. 2026. doi:10.1111/clr.70158

BibTeX
@article{Baranowski2026,
  author = {Baranowski, Julian Hessel and Braian, Michael and Stenport, Victoria Franke and Wennerberg, Ann},
  title = {Investigations of Intraoral Scan Body Extensions. Effect of Design on the Accuracy, Usability, and Feasibility for Digital Full‐Arch Implant Scans In Vitro},
  journal = {Clinical Oral Implants Research},
  year = {2026},
  doi = {10.1111/clr.70158}
}
Kostenlos starten →