Influence of Bone Density and Guide Protocol on the Accuracy of Self‐Cutting Implants Using Static Guided Implant Placement—An In Vitro Study
Clinical Oral Implants Research · 2025-10 · 36 (10) · 1248-1260
DOI: 10.1111/clr.14470 ↗Kurz gefasst
Studienprotokoll in der Zeitschrift Clinical Oral Implants Research, online erschienen am 03.07.2025. Verfasst von Caroline Streichfuss und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Prosthodontics and Biomaterials, Centre for Implantology, RWTH Aachen University Hospital, Aachen, Germany. Hauptthemen laut MEDLINE: Dental Implantation, Endosseous, Bone Density, Dental Implants, Surgery, Computer-Assisted. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
To investigate the influence of bone density and guide protocol on the accuracy of static guided implant placement of self‐cutting implants in vitro.
Methods
A total of 242 implant replicas of self‐cutting implants and implants with a tapered design as a control were placed in 40 maxilla replica models with healed posterior ridges differing in bone density and anterior extraction sockets. After full‐guided osteotomy, the implants in the healed ridge were either placed through the template or freehand. In anterior extraction sockets, a pilot‐guided group was additionally investigated. Postoperative scans were superimposed with the planning and investigated for angular, coronal, and apical deviation.
Results
For 3D coronal deviation, the study indicates significantly lower deviations in soft bone ( p < 0.01) and for guided implant placement ( p < 0.01) in posterior sites. Apical and angular deviations are significantly influenced by the placement protocol ( p ≤ 0.03) and the presence of reference points from neighbouring teeth ( p < 0.01). In the anterior site, the coronal deviation seems to be improved by a direct view of the surgical field, while 3D apical and angular accuracy appears to benefit from a high degree of guidance ( p < 0.01).
Conclusion
Within the limitations of this in vitro study, for low bone density, immediate implant placement, and in clinically challenging situations without reference to the adjacent teeth, the use of template guidance for insertion statistically significantly reduced deviation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Prosthodontics and Biomaterials, Centre for Implantology, RWTH Aachen University Hospital, Aachen, Germany.
ORCID-Profile: Caroline Streichfuss
Zitieren
APA
Streichfuss, C., Wolfart, S., & Waltenberger, L. (2025). Influence of Bone Density and Guide Protocol on the Accuracy of Self‐Cutting Implants Using Static Guided Implant Placement—An In Vitro Study. Clinical Oral Implants Research, 36(10), 1248-1260. https://doi.org/10.1111/clr.14470
Vancouver
Streichfuss C, Wolfart S, Waltenberger L. Influence of Bone Density and Guide Protocol on the Accuracy of Self‐Cutting Implants Using Static Guided Implant Placement—An In Vitro Study. Clinical Oral Implants Research. 2025;36(10):1248-1260. doi:10.1111/clr.14470
BibTeX
@article{Streichfuss2025,
author = {Streichfuss, Caroline and Wolfart, Stefan and Waltenberger, Lukas},
title = {Influence of Bone Density and Guide Protocol on the Accuracy of Self‐Cutting Implants Using Static Guided Implant Placement—An In Vitro Study},
journal = {Clinical Oral Implants Research},
year = {2025},
volume = {36},
number = {10},
pages = {1248--1260},
doi = {10.1111/clr.14470}
}