Are Sleeves Necessary in Static Computer‐Assisted Implant Surgery? A Comparative In Vitro Analysis
Clinical Oral Implants Research · 2025
DOI: 10.1111/clr.14368 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Clinical Oral Implants Research, 2025. Verfasst von Jenna Hang und 1 weiteren Autor:innen. Beteiligte Einrichtung: Advanced Educational Program in Periodontics, Department of Diagnostic and Surgical Sciences, Marquette University School of Dentistry, Milwaukee, Wi…. Hauptthemen laut MEDLINE: Surgery, Computer-Assisted, Mandible, Cone-Beam Computed Tomography, Dental Implantation, Endosseous. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 39 Literaturangaben.
Abstract (Originalsprache)
Objectives
This study aims to examine differences in trueness and precision between surgical guides with (S) and without sleeves (SL). A secondary aim was to assess the impact of the sleeve‐to‐bone distance. Materials and Methods Mandible replicas ( n = 120) were printed from an STL file obtained from a clinical CBCT. The mandibles were divided into sleeved (S, n = 60) and sleeveless (SL, n = 60) groups, each further divided into three categories ( n = 20 each) with different heights from the guide to the implant platform: 2 mm (H2), 4 mm (H4), or 6 mm (H6). Digital planning and surgical guide design were done for a 4.1 × 10 mm implant for site #30. Post‐op positions were captured using a scan body and lab scanner. Angular deviation was the primary outcome, with 3D and 2D deviations as secondary parameters. Statistical analysis included two‐sample t‐tests, and one‐way and two‐way ANOVA.
Results
Group S (2.41 ± 1.41°) had significantly greater angular deviation than Group SL (1.65 ± 0.93°; p = 0.0001). Angular deviation increased with sleeve‐to‐bone distance. H2 deviations were 1.48 ± 0.80° (S) vs. 1.02 ± 0.45° (SL; p < 0.05), H4: 2.36 ± 1.04° (S) vs. 1.48 ± 0.79° (SL; p < 0.05), H6: 3.37 ± 0.67° (S) vs. 2.46 ± 0.89° (SL; p < 0.05). 3D deviation at the implant platform was 0.36 ± 0.17 mm (S) vs. 0.30 ± 0.15 mm (SL; p < 0.05) and at the apex 0.74 ± 0.34 mm (S) vs. 0.53 ± 0.31 mm (SL; p < 0.01). Group SL at H2 had the smallest inter‐implant distance (0.53 ± 0.37°), while Group S at H4 had the largest (1.20 ± 0.84°; p < 0.05).
Conclusions
Sleeveless guides are more accurate than sleeved guides, and angular deviation is influenced by the distance from the guide to the implant platform.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Advanced Educational Program in Periodontics, Department of Diagnostic and Surgical Sciences, Marquette University School of Dentistry, Milwaukee, Wisconsin, USA.
- Department of Diagnostic and Surgical Sciences, Marquette University School of Dentistry, Milwaukee, Wisconsin, USA.
ORCID-Profile: Arndt Guentsch
Zitieren
APA
Hang, J., & Guentsch, A. (2025). Are Sleeves Necessary in Static Computer‐Assisted Implant Surgery? A Comparative In Vitro Analysis. Clinical Oral Implants Research. https://doi.org/10.1111/clr.14368
Vancouver
Hang J, Guentsch A. Are Sleeves Necessary in Static Computer‐Assisted Implant Surgery? A Comparative In Vitro Analysis. Clinical Oral Implants Research. 2025. doi:10.1111/clr.14368
BibTeX
@article{Hang2025,
author = {Hang, Jenna and Guentsch, Arndt},
title = {Are Sleeves Necessary in Static Computer‐Assisted Implant Surgery? A Comparative In Vitro Analysis},
journal = {Clinical Oral Implants Research},
year = {2025},
doi = {10.1111/clr.14368}
}
EUVIMED