EUVIMEDEUVIMED entdecken
Crossref

Accuracy of manual and artificial intelligence‐based superimposition of cone‐beam computed tomography with digital scan data, utilizing an implant planning software: A randomized clinical study

Genauigkeit der manuellen und KI-basierten Überlagerung von Cone-Beam-CT mit digitalen Scandaten mithilfe einer Implantatplanungssoftware: randomisierte klinische Studie

Panagiotis Ntovas iD, Laurent Marchand, Matthew Finkelman, Marta Revilla-León iD, Wael Att

Clinical Oral Implants Research · 2024-10 · 35 (10) · 1262-1272

DOI: 10.1111/clr.14313 ↗

Auf einen Blick

Bei 150 Patient:innen war eine vollautomatische KI-Registrierung ohne Restaurationsartefakte ähnlich genau wie konventionelle Methoden. Bei mehreren Restaurationsartefakten spielte die Erfahrung der Anwender:innen eine Rolle.

Fragestellung

Wie genau sind konventionelle und automatische KI-basierte Registrierungen von CBCT mit intraoralen Scans, und welchen Einfluss haben Erfahrung, Restaurationsartefakte, fehlende Zähne und Freiend-Situationen?

Methode

Randomisierte klinische Studie; 150 zufällig ausgewählte Patient:innen; Registrierungen in einer Implantatplanungssoftware; Genauigkeitsvergleich mit metrologischer Software anhand anatomischer Landmarken.

Ergebnisse

Vollautomatische KI-Registrierung war bei Patient:innen ohne Restaurationen nicht signifikant anders als die getesteten konventionellen Methoden. Bei mehreren Restaurationsartefakten war die Erfahrung der Anwender:innen wichtig; die Genauigkeit wurde von Freiend-Situationen beeinflusst, nicht von der absoluten Zahl fehlender Zähne (p < .0083).

Schlussfolgerung der Autor:innen

Ohne Bildartefakte kann eine automatisierte KI-basierte Registrierung von CBCT- und Modellscandaten so genau sein wie konventionelle Überlagerungsmethoden. Zahl und Größe der Überlagerungsbereiche sollten situationsabhängig gewählt werden.

Kernaussagen

  • Ohne Restaurationsartefakte war KI-Registrierung ähnlich genau wie konventionelle Methoden.
  • Bei mehreren Restaurationsartefakten beeinflusste Erfahrung die Genauigkeit.
  • Freiend-Situationen beeinflussten die Genauigkeit; fehlende Zähne nicht.

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Oral Implants Research, online erschienen am 10.06.2024. Verfasst von Panagiotis Ntovas und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Prosthodontics, School of Dental Medicine, Tufts University School of Dental Medicine, Boston, Massachusetts, USA. Hauptthemen laut MEDLINE: Artificial Intelligence, Cone-Beam Computed Tomography, Software. Die Arbeit wurde bisher 24-mal zitiert (Crossref) und nennt 33 Literaturangaben.

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To investigate the accuracy of conventional and automatic artificial intelligence (AI)‐based registration of cone‐beam computed tomography (CBCT) with intraoral scans and to evaluate the impact of user's experience, restoration artifact, number of missing teeth, and free‐ended edentulous area. Materials and Methods Three initial registrations were performed for each of the 150 randomly selected patients, in an implant planning software: one from an experienced user, one from an inexperienced operator, and one from a randomly selected post‐graduate student of implant dentistry. Six more registrations were performed for each dataset by the experienced clinician: implementing a manual or an automatic refinement, selecting 3 small or 3 large in‐diameter surface areas and using multiple small or multiple large in‐diameter surface areas. Finally, an automatic AI‐driven registration was performed, using the AI tools that were integrated into the utilized implant planning software. The accuracy between each type of registration was measured using linear measurements between anatomical landmarks in metrology software.

Results

Fully automatic‐based AI registration was not significantly different from the conventional methods tested for patients without restorations. In the presence of multiple restoration artifacts, user's experience was important for an accurate registration. Registrations' accuracy was affected by the number of free‐ended edentulous areas, but not by the absolute number of missing teeth ( p < .0083).

Conclusions

In the absence of imaging artifacts, automated AI‐based registration of CBCT data and model scan data can be as accurate as conventional superimposition methods. The number and size of selected superimposition areas should be individually chosen depending on each clinical situation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Artificial IntelligenceCone-Beam Computed TomographySoftwareAdultAgedFemaleHumansMaleMiddle AgedArtifactsDental Implantation, EndosseousPatient Care Planning

Stichwörter der Autor:innen

CBCT matchingartifactsdata fusiondeep learningguided implant surgerymodel scan dataregistration accuracyvirtual implant planning

Autor:innen und Institutionen

  • Department of Prosthodontics, School of Dental Medicine, Tufts University School of Dental Medicine, Boston, Massachusetts, USA.
  • Department of Public Health and Community Service, Tufts University School of Dental Medicine, Boston, Massachusetts, USA.
  • Department of Restorative Dentistry, School of Dentistry, University of Washington, Seattle, Washington, USA.
  • Faculty and Director of Research and Digital Dentistry, Kois Center, Seattle, Washington, USA.
  • Center for Dental Medicine, Department of Prosthetic Dentistry, ,Faculty of Medicine, University of Freiburg, Freiburg, Germany.
  • Private Practice, The Face Dental Group, Boston, Massachusetts, USA.

ORCID-Profile: Panagiotis Ntovas, Marta Revilla-León

Zitieren

APA

Ntovas, P., Marchand, L., Finkelman, M., Revilla-León, M., & Att, W. (2024). Accuracy of manual and artificial intelligence‐based superimposition of cone‐beam computed tomography with digital scan data, utilizing an implant planning software: A randomized clinical study. Clinical Oral Implants Research, 35(10), 1262-1272. https://doi.org/10.1111/clr.14313

Vancouver

Ntovas P, Marchand L, Finkelman M, Revilla-León M, Att W. Accuracy of manual and artificial intelligence‐based superimposition of cone‐beam computed tomography with digital scan data, utilizing an implant planning software: A randomized clinical study. Clinical Oral Implants Research. 2024;35(10):1262-1272. doi:10.1111/clr.14313

BibTeX
@article{Ntovas2024,
  author = {Ntovas, Panagiotis and Marchand, Laurent and Finkelman, Matthew and Revilla-León, Marta and Att, Wael},
  title = {Accuracy of manual and artificial intelligence‐based superimposition of cone‐beam computed tomography with digital scan data, utilizing an implant planning software: A randomized clinical study},
  journal = {Clinical Oral Implants Research},
  year = {2024},
  volume = {35},
  number = {10},
  pages = {1262--1272},
  doi = {10.1111/clr.14313}
}
Kostenlos starten →