Sequential osseointegration of a novel implant system based on 3D printing in comparison with conventional titanium implants
Clinical Oral Implants Research · 2023
DOI: 10.1111/clr.14072 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Oral Implants Research, 2023. Verfasst von Niklaus P Lang und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Periodontology, School of Dental Medicine, University of Bern, Bern, Switzerland. Hauptthemen laut MEDLINE: Osseointegration, Dental Implants. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Ruetschi Technology (Muntelier, Switzerland).
Abstract (Originalsprache)
Objectives
To evaluate the sequential osseointegration of a novel titanium implant system based on a 3D printing technology in comparison with conventional titanium implants. Material and methods Two novel titanium implants based on 3D printing were tested in the mandible of eight Beagle dogs. As a control, two different commercially available titanium implants were used. The implants were staged to accommodate healing periods of 2 and 6 weeks. The primary outcome variable was bone‐to‐implant contact (BIC) in non‐decalcified tissue sections and micro‐CT analysis.
Results
Histomorphometrically, the proportions of tissues adjacent to the implant surfaces were similar for all implants, whereas the BIC percentage of new mineralized bone was greater for the control implants after both 2 and 6 weeks ( p < .05). Micro‐CT analysis revealed increasing osseous volume and BIC from 2 to 6 weeks. In contrast to the histomorphometry, the BIC evaluation with the micro‐CT data revealed a significantly higher BIC for the two test implants compared with controls ( p < .001). The analysis of the total implant surface area disclosed a value that was approximately double as high for the test compared to the control implants.
Conclusions
The novel titanium implant system based on 3D printing yielded values for osseointegration that were adequate and satisfactory. The higher percentage of new mineralized bone in the control implants is explained by the fact of a completely different three‐dimensional surface area.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Periodontology, School of Dental Medicine, University of Bern, Bern, Switzerland.
- Department of Cranio-Maxillofacial Surgery, University Hospital, University of Bern, Bern, Switzerland.
- Private Clinic ZahnArt-Belp AG, Bern, Switzerland.
- Department of Veterinary Clinical Sciences, University of Santiago de Compostela, Lugo, Spain.
ORCID-Profile: Niklaus P Lang, Jean-Claude Imber, Kiri N Lang, Bruno Schmid, Fernando Muñoz, Dieter D Bosshardt, Nikola Saulacic
Förderung
- Ruetschi Technology (Muntelier, Switzerland)
Zitieren
APA
Lang, N. P., Imber, J. C., Lang, K. N., Schmid, B., Muñoz, F., Bosshardt, D. D., & Saulacic, N. (2023). Sequential osseointegration of a novel implant system based on 3D printing in comparison with conventional titanium implants. Clinical Oral Implants Research. https://doi.org/10.1111/clr.14072
Vancouver
Lang NP, Imber JC, Lang KN, Schmid B, Muñoz F, Bosshardt DD, et al. Sequential osseointegration of a novel implant system based on 3D printing in comparison with conventional titanium implants. Clinical Oral Implants Research. 2023. doi:10.1111/clr.14072
BibTeX
@article{Lang2023,
author = {Lang, Niklaus P and Imber, Jean-Claude and Lang, Kiri N and Schmid, Bruno and Muñoz, Fernando and Bosshardt, Dieter D and Saulacic, Nikola},
title = {Sequential osseointegration of a novel implant system based on 3D printing in comparison with conventional titanium implants},
journal = {Clinical Oral Implants Research},
year = {2023},
doi = {10.1111/clr.14072}
}
EUVIMED