EUVIMEDEUVIMED entdecken
Crossref

Accuracy of a dynamic navigation system for dental implantation with two different workflows and intraoral markers compared to static‐guided implant surgery: An in‐vitro study

Maximilian Struwe, Wadim Leontiev iD, Thomas Connert, Sebastian Kühl iD, Andreas Filippi, Valentin Herber, Dorothea Dagassan-Berndt

Clinical Oral Implants Research · 2023-03 · 34 (3) · 196-208

DOI: 10.1111/clr.14030 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Oral Implants Research, online erschienen am 22.01.2023. Verfasst von Maximilian Struwe und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oral Surgery, University Center for Dental Medicine, University of Basel, Basel, Switzerland. Hauptthemen laut MEDLINE: Dental Implants, Surgery, Computer-Assisted. Die Arbeit wurde bisher 20-mal zitiert (Crossref) und nennt 22 Literaturangaben. Gefördert von: Center of Dental Traumatology Basel (grant no. AZ 20-01 ZUZ).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To investigate the accuracy of a miniaturized dynamic navigation system with intraoral markers and two different workflows for dental implantation and to compare with static computer‐assisted implant surgery (sCAIS) surgery. Materials and Methods Two operators performed a total of 270 implant insertions in polyurethane mandibular models under simulated clinical conditions. Implants were placed after CBCT‐based virtual planning in three different groups: two workflows utilizing dynamic computer‐assisted implant surgery (dCAIS; DG1: marker in CBCT; DG2: 3D‐printed marker) and the others with sCAIS (TG: template guided). Postoperative surface scans were matched to the planning data and allowed an evaluation of the angular and spatial deviation between the planned and the actually achieved implant position. Descriptive statistics were followed by a Mixed Model Analysis to determine the influence of the operator, the method, and operating area on different accuracy parameters and the random effect of the model number.

Results

The mean angular deviation ranged from 2.26° (DG1) to 2.96° (TG). The mean 3D deviation at the implant's tip ranged from 1.08 mm (TG) to 1.51 mm (DG2) and at the implant's base from 0.69 mm (TG) to 1.49 mm (DG2). The operator showed no significant influence on the accuracy. The method showed significant influence on singular parameters and the operating area on all spatial accuracy parameters.

Conclusions

Dynamic navigation systems with intraoral markers enable accurate implant positioning, which is comparable to the static‐guided implant surgery. 3D‐printed markers provide less accurate results compared to prefabricated markers, attached before CBCT scan.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Dental ImplantsSurgery, Computer-AssistedHumansDental Implantation, EndosseousWorkflowCone-Beam Computed TomographyComputer-Aided DesignImaging, Three-Dimensional

Stichwörter der Autor:innen

dynamic navigation systemguided implant surgeryoral implant surgery

Substanzen

Dental Implants

Autor:innen und Institutionen

  • Department of Oral Surgery, University Center for Dental Medicine, University of Basel, Basel, Switzerland.
  • Department of Periodontology, Endodontology and Cariology, University Center for Dental Medicine Basel UZB, University of Basel, Basel, Switzerland.
  • Center for Dental Imaging, University Center for Dental Medicine, University of Basel, Basel, Switzerland.

ORCID-Profile: Wadim Leontiev, Sebastian Kühl

Förderung

  • Center of Dental Traumatology Basel (grant no. AZ 20-01 ZUZ)

Zitieren

APA

Struwe, M., Leontiev, W., Connert, T., Kühl, S., Filippi, A., Herber, V., & Dagassan-Berndt, D. (2023). Accuracy of a dynamic navigation system for dental implantation with two different workflows and intraoral markers compared to static‐guided implant surgery: An in‐vitro study. Clinical Oral Implants Research, 34(3), 196-208. https://doi.org/10.1111/clr.14030

Vancouver

Struwe M, Leontiev W, Connert T, Kühl S, Filippi A, Herber V, et al. Accuracy of a dynamic navigation system for dental implantation with two different workflows and intraoral markers compared to static‐guided implant surgery: An in‐vitro study. Clinical Oral Implants Research. 2023;34(3):196-208. doi:10.1111/clr.14030

BibTeX
@article{Struwe2023,
  author = {Struwe, Maximilian and Leontiev, Wadim and Connert, Thomas and Kühl, Sebastian and Filippi, Andreas and Herber, Valentin and Dagassan-Berndt, Dorothea},
  title = {Accuracy of a dynamic navigation system for dental implantation with two different workflows and intraoral markers compared to static‐guided implant surgery: An in‐vitro study},
  journal = {Clinical Oral Implants Research},
  year = {2023},
  volume = {34},
  number = {3},
  pages = {196--208},
  doi = {10.1111/clr.14030}
}
Kostenlos starten →