EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Robotic Quad Zygomatic Implant Surgery Accuracy Ex Vivo With a New Semi‐Autonomous Optically‐Driven Haptic System

Alessandro Pozzi iD, James Chow iD, Andrea Laureti iD, Lorenzo Arcuri iD, Ilaria Cammarota, Alessandra Nardi, Hom-Lay Wang iD

Clinical Implant Dentistry and Related Research · 2026-06 · 28 (3)

DOI: 10.1111/cid.70155 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Implant Dentistry and Related Research, online erschienen am 28.04.2026. Verfasst von Alessandro Pozzi und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Science and Translational Medicine, University of Rome Tor Vergata, Rome, Italy. Hauptthemen laut MEDLINE: Zygoma, Surgery, Computer-Assisted, Dental Implantation, Endosseous, Robotic Surgical Procedures. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To evaluate ex vivo accuracy of robotic computer‐assisted implant surgery (r‐CAIS) for quad zygomatic implant (ZI) placement with a novel semi‐autonomous, optically‐driven, haptic‐assisted system.

Methods

Two expert operators placed ZIs in human cadavers. Accuracy was evaluated by superimposing pre‐ and post‐operative cone‐beam computed tomography (CBCT) scans, measuring linear (mm) on the x ‐, y ‐, and z ‐ axes and angular (degrees) deviations. Sample size was calculated. The main analysis focused on composite outcomes (global platform, global apical, and angular deviation), while extended analyses examined deviations on individual spatial axes (buccal–lingual, mesial–distal, depth, and non‐depth) for both implant platform and apex. Three multivariate models were executed on global platform, global apical, and angular deviation considering implant position (anterior vs. posterior) and quadrant (left vs. right).

Results

Forty ZIs were placed in 10 human cadavers. The mean global deviations at implant platform and apex were 1.16 mm (standard deviation (SD) 0.47) and 1.51 mm (SD 0.70), with mean angular deviation of 1.04° (SD 0.65°). In the multivariate models, implant position was found to be significant for global platform deviation ( p = 0.0218), with posterior implants performing better, and for angular deviation ( p = 0.0465), with anterior implants performing better.

Conclusion

The optically driven haptic‐assisted robotic system demonstrated high accuracy in quad ZI surgery, with all implants placed without intraoperative complications or significant errors impacting adjacent anatomical structures. Mean global linear deviations of 1 and 1.5 mm at platform and apex, and 1° of angular deviation were experienced. Further research is necessary to validate these promising results in vivo.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ZygomaSurgery, Computer-AssistedDental Implantation, EndosseousRobotic Surgical ProceduresHumansCadaverCone-Beam Computed TomographyFemale

Autor:innen und Institutionen

  • Department of Clinical Science and Translational Medicine, University of Rome Tor Vergata, Rome, Italy.
  • Department of Restorative Sciences Augusta University, Augusta, Georgia, USA.
  • Department of Restorative Dentistry and Biomaterials Sciences, Harvard School of Dental Medicine, Boston, Massachusetts, USA.
  • Brånemark Osseointegration Centre, Hong Kong, People's Republic of China.
  • Department of Chemical Science and Technologies, University of Rome Tor Vergata, Rome, Italy.
  • Division of Fixed Prosthodontics and Biomaterials, University Clinics for Dental Medicine, University of Geneva, Geneva, Switzerland.

ORCID-Profile: Alessandro Pozzi, James Chow, Andrea Laureti, Lorenzo Arcuri, Hom-Lay Wang

Zitieren

APA

Pozzi, A., Chow, J., Laureti, A., Arcuri, L., Cammarota, I., Nardi, A., & Wang, H. L. (2026). Robotic Quad Zygomatic Implant Surgery Accuracy Ex Vivo With a New Semi‐Autonomous Optically‐Driven Haptic System. Clinical Implant Dentistry and Related Research, 28(3). https://doi.org/10.1111/cid.70155

Vancouver

Pozzi A, Chow J, Laureti A, Arcuri L, Cammarota I, Nardi A, et al. Robotic Quad Zygomatic Implant Surgery Accuracy Ex Vivo With a New Semi‐Autonomous Optically‐Driven Haptic System. Clinical Implant Dentistry and Related Research. 2026;28(3). doi:10.1111/cid.70155

BibTeX
@article{Pozzi2026,
  author = {Pozzi, Alessandro and Chow, James and Laureti, Andrea and Arcuri, Lorenzo and Cammarota, Ilaria and Nardi, Alessandra and Wang, Hom-Lay},
  title = {Robotic Quad Zygomatic Implant Surgery Accuracy Ex Vivo With a New Semi‐Autonomous Optically‐Driven Haptic System},
  journal = {Clinical Implant Dentistry and Related Research},
  year = {2026},
  volume = {28},
  number = {3},
  doi = {10.1111/cid.70155}
}
Kostenlos starten →