Repeatability and agreement of total corneal astigmatism measured in keratoconic eyes using four current devices

Yue Feng, Tore Nitter, Geir Bertelsen, Aleksandar Stojanovic iD

Clinical & Experimental Ophthalmology · 2024

DOI: 10.1111/ceo.14423 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Clinical & Experimental Ophthalmology, 2024. Verfasst von Yue Feng und 3 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Community Medicine, Faculty of Health Sciences, University in Tromsø, Tromsø, Norway. Hauptthemen laut MEDLINE: Tomography, Optical Coherence, Astigmatism, Keratoconus, Cornea. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 54 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Norwegian research council, Norges Forskningsråd.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

To evaluate repeatability and agreement in measurements of total corneal astigmatism (TCA) in keratoconic eyes, using four optical coherence tomography (OCT)‐based devices: Anterion, Casia SS‐1000, IOLMaster 700, and MS‐39.

Methods

Three consecutive measurements were taken with each device in 136 eyes. TCA values were converted into components J 0 and J 45 . The Anterion and the IOLMaster 700 also provided axial length (AL) measurements. The repeatability was calculated using pooled within‐subject standard deviation ( S w ). The agreement among the four devices was assessed by pairwise comparisons and Bland–Altman plots.

Results

For all devices, the repeatability of TCA measurements showed S w ≤0.23 D for TCA magnitude, ≤0.14 D for J 0 , and ≤0.12 D for J 45 . There were statistically significant differences in TCA magnitude for each pair, except for IOLMaster 700 with MS‐39, and Anterion with MS‐39. The repeatability ( S w ) of axis measurements had a statistically significant negative correlation with the TCA magnitude ( p < 0.001 for all devices). Both Anterion and IOLMaster 700 had high repeatability in AL measurements ( S w : 0.007 mm for Anterion and 0.009 mm for IOLMaster 700). The difference in AL between the two was 0.015 ± 0.033 mm ( p < 0.001).

Conclusions

All four devices showed good repeatability in TCA measurements in keratoconic eyes, the agreement for TCA measurements between the tested devices was generally low. Anterion and IOLMaster 700 showed good repeatability and agreement in AL measurements.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, Optical CoherenceAstigmatismKeratoconusCorneaHumansReproducibility of ResultsMaleFemaleAdultYoung AdultAxial Length, EyeMiddle AgedProspective StudiesCorneal TopographyAdolescent

Stichwörter der Autor:innen

axial lengthkeratoconusoptical coherence tomographytotal corneal astigmatism

Autor:innen und Institutionen

  • Institute of Community Medicine, Faculty of Health Sciences, University in Tromsø, Tromsø, Norway.
  • Department of Ophthalmology, University Hospital North Norway, Tromsø, Norway.
  • Institute of Clinical Medicine, Faculty of Health Sciences, University in Tromsø, Tromsø, Norway.

ORCID-Profile: Aleksandar Stojanovic

Förderung

  • Norwegian research council
  • Norges Forskningsråd

Zitieren

APA

Feng, Y., Nitter, T., Bertelsen, G., & Stojanovic, A. (2024). Repeatability and agreement of total corneal astigmatism measured in keratoconic eyes using four current devices. Clinical & Experimental Ophthalmology. https://doi.org/10.1111/ceo.14423

Vancouver

Feng Y, Nitter T, Bertelsen G, Stojanovic A. Repeatability and agreement of total corneal astigmatism measured in keratoconic eyes using four current devices. Clinical & Experimental Ophthalmology. 2024. doi:10.1111/ceo.14423

BibTeX
@article{Feng2024,
  author = {Feng, Yue and Nitter, Tore and Bertelsen, Geir and Stojanovic, Aleksandar},
  title = {Repeatability and agreement of total corneal astigmatism measured in keratoconic eyes using four current devices},
  journal = {Clinical & Experimental Ophthalmology},
  year = {2024},
  doi = {10.1111/ceo.14423}
}
Kostenlos starten →