Influence of the CleaRing intraocular open capsule device on refractive predictability in cataract surgery
Clinical & Experimental Ophthalmology · 2023
DOI: 10.1111/ceo.14282 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical & Experimental Ophthalmology, 2023. Verfasst von Elinor Megiddo-Barnir und 1 weiteren Autor:innen. Beteiligte Einrichtung: Ophthalmology Department, The Edith Wolfson Medical Center, Holon, Israel. Hauptthemen laut MEDLINE: Refraction, Ocular, Lens Implantation, Intraocular, Capsule Opacification, Phacoemulsification. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 43 Literaturangaben. Gefördert von: Hanita IOLs.
Abstract (Originalsprache)
Background
The marked improvement in cataract surgery and intraocular lens (IOL) quality has led to a decline in posterior capsular opacification (PCO) incidence; however, PCO remains a common complication of cataract surgery. The CleaRing intraocular capsule open device (IOCD) decreases PCO incidence. We aimed to investigate the influence of the CleaRing IOCD on refractive predictability in cataract surgery.
Methods
We conducted this prospective pilot study at the Wolfson Medical Center, Holon, Israel. Ten eyes of patients who underwent cataract surgery and insertion of an IOL after IOCD implantation into the capsular bag were included. All patients completed 12 months of follow‐up, including refraction, measurement of uncorrected (UDVA) and best‐corrected distance visual acuity (CDVA), slit‐lamp biomicroscopy, and ultrasound biomicroscopy.
Results
All the surgeries were uneventful, with no postoperative complications. The IOL was centred in the device and bag in all cases. The mean prediction error at 1 and 12 months postoperatively was +0.28 ± 0.32 D and +0.50 ± 0.32 D, respectively. The mean UDVA was 0.17 ± 0.13 and 0.15 ± 0.11 logMAR, and the mean CDVA was 0.04 ± 0.10 and 0.04 ± 0.06 logMAR, respectively. The manifest refractive cylinders at 12 months postoperatively were compatible with corneal astigmatism.
Conclusions
Implantation of the IOCD resulted in a slight, predicted, and stable hyperopic shift with a low standard deviation. The standard deviation of the prediction error demonstrated excellent refractive accuracy and predictability using the IOCD, which was as low as 0.32 D at the 12‐month follow‐up.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Ophthalmology Department, The Edith Wolfson Medical Center, Holon, Israel.
- Sackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel.
ORCID-Profile: Elinor Megiddo-Barnir, Guy Kleinmann
Förderung
- Hanita IOLs
Zitieren
APA
Megiddo-Barnir, E., & Kleinmann, G. (2023). Influence of the CleaRing intraocular open capsule device on refractive predictability in cataract surgery. Clinical & Experimental Ophthalmology. https://doi.org/10.1111/ceo.14282
Vancouver
Megiddo-Barnir E, Kleinmann G. Influence of the CleaRing intraocular open capsule device on refractive predictability in cataract surgery. Clinical & Experimental Ophthalmology. 2023. doi:10.1111/ceo.14282
BibTeX
@article{MegiddoBarnir2023,
author = {Megiddo-Barnir, Elinor and Kleinmann, Guy},
title = {Influence of the CleaRing intraocular open capsule device on refractive predictability in cataract surgery},
journal = {Clinical & Experimental Ophthalmology},
year = {2023},
doi = {10.1111/ceo.14282}
}
EUVIMED