Flexible and navigable suction access sheaths: balancing sheath and scope size for desired flows
BJU International · 2025
DOI: 10.1111/bju.16865 ↗Kurz gefasst
Fachartikel in der Zeitschrift BJU International, 2025. Verfasst von Richard Menzies-Wilson und 4 weiteren Autor:innen. Beteiligte Einrichtung: Nuffield Department of Surgery, University of Oxford, Oxford, UK. Hauptthemen laut MEDLINE: Ureteroscopes, Ureteroscopy, Kidney. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Boston Scientific Corporation.
Abstract (Originalsprache)
Objective
To study the effects of ureteroscope diameter, ‘flexible and navigable suction’ access sheaths (FANS) diameter, and irrigation pressure on intrarenal pressure (IRP) and irrigation fluid flow rates in ex vivo porcine kidneys; these benchtop data were compared against mathematical modelling results. Materials and Methods Fresh ex vivo porcine kidneys and ureters were used. The tip of an 11/13‐ or 10/12‐F ClearPetra ® FANS was placed within the renal pelvis. Either a 9.5‐F Boston Scientific LithoVue™, 7.5‐F PUSEN PU3033, or 6.3‐F HugeMed HU30M ureteroscope was inserted through the sheath. Irrigation pressures of 0, 50, 100, 150 and 200 mmHg were applied. No suction was applied. The corresponding steady state IRP and flow rate was recorded. This was replicated in three separate porcine kidneys. Benchtop data were compared to mathematical model predictions. The maximum flow rates that could be achieved with an IRP ≤30 mmHg were extrapolated.
Results
The FANS and ureteroscope geometries affect outflow resistance. Lower outflow resistances, allow higher irrigation pressures and flow rates to be used whilst maintaining an IRP ≤30 mmHg (without suction). Whilst keeping IRP ≤30 mmHg with an 11/13‐F FANS: ~700 mmHg irrigation (~120 mL/min) could be used with a 6.3‐F ureteroscope; ~300 mmHg irrigation (flow rate ~65 mL/min) could be used with a 7.5‐F ureteroscope; ~100 mmHg irrigation (~15 mL/min) could be used with a 9.5‐F ureteroscope.
Conclusions
Using a FANS in the renal pelvis potentially allows for higher irrigation pressures whilst maintaining a low IRP (without the need for suction). With intraoperative monitoring of IRP, irrigation pressure could be increased to achieve flow rates up to 24 times higher than those achievable with gravity irrigation (120 vs ~5 mL/min).
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Nuffield Department of Surgery, University of Oxford, Oxford, UK.
- Boston Scientific Corporation, Marlborough, MA, USA.
ORCID-Profile: Richard Menzies-Wilson, Candace Rhodes, Ben Turney
Förderung
- Boston Scientific Corporation
Zitieren
APA
Menzies-Wilson, R., Williams, J., Ruiken, T., Rhodes, C., & Turney, B. (2025). Flexible and navigable suction access sheaths: balancing sheath and scope size for desired flows. BJU International. https://doi.org/10.1111/bju.16865
Vancouver
Menzies-Wilson R, Williams J, Ruiken T, Rhodes C, Turney B. Flexible and navigable suction access sheaths: balancing sheath and scope size for desired flows. BJU International. 2025. doi:10.1111/bju.16865
BibTeX
@article{MenziesWilson2025,
author = {Menzies-Wilson, Richard and Williams, Jessica and Ruiken, Thijs and Rhodes, Candace and Turney, Ben},
title = {Flexible and navigable suction access sheaths: balancing sheath and scope size for desired flows},
journal = {BJU International},
year = {2025},
doi = {10.1111/bju.16865}
}
EUVIMED