Comparison of holmium:yttrium‐aluminium‐garnet ( YAG) , thulium fiber laser, and pulsed thulium: YAG lasers on soft tissue: an ex vivo study
BJU International · 2024-11 · 134 (5) · 763-772
DOI: 10.1111/bju.16447 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift BJU International, online erschienen am 12.07.2024. Verfasst von Stessy Kutchukian und 9 weiteren Autor:innen. Beteiligte Einrichtung: Service d'Urologie, AP-HP, Hôpital Tenon, Sorbonne Université, Paris, France. Hauptthemen laut MEDLINE: Thulium, Lasers, Solid-State. Die Arbeit wurde bisher 16-mal zitiert (Crossref) und nennt 27 Literaturangaben.
Abstract (Originalsprache)
Objectives
To assess laser–tissue interactions through ablation, coagulation, and carbonisation characteristics in a non‐perfused porcine kidney model between three pulsed lasers: holmium (Ho): yttrium‐aluminium‐garnet (YAG), thulium fiber laser (TFL), and pulsed thulium (p‐Tm):YAG. Materials and Methods A 150‐W Ho:YAG, a 60‐W TFL, and a 100‐W p‐Tm:YAG lasers were compared. The laser settings that can be set identically between the three lasers and be clinically relevant for prostate laser enucleation were identified and used on fresh, unfrozen porcine kidneys. Laser incisions were performed using stripped laser fibers of 365 and 550 μm, set at distances of 0 and 1 mm from the tissue surface at a constant speed of 2 mm/s. Histological analysis evaluated shape, depth, width of the incision, axial coagulation depth, and presence of carbonisation.
Results
Incision depths, widths, and coagulation zones were greater with Ho:YAG and p‐Tm:YAG lasers than TFL. Although no carbonisation was found with the Ho:YAG and p‐Tm:YAG lasers, it was common with TFL, especially at high frequencies. The shapes of the incisions and coagulation zones were more regular and homogeneous with the p‐Tm:YAG laser and TFL than with Ho:YAG laser. Regardless of the laser used, short pulse durations resulted in deeper incisions than long pulse durations. Concerning the distance, we found that to be effective, TFL had to be used in contact with the tissue. Finally, 365‐μm fibers resulted in deeper incisions, while 550‐μm fibers led to wider incisions and larger coagulation zones.
Conclusion
Histological analysis revealed greater tissue penetration with the p‐Tm:YAG laser compared to the TFL, while remaining less than with Ho:YAG. Its coagulation properties seem interesting insofar as it provides homogeneous coagulation without carbonisation, while incisions remained uniform without tissue laceration. Thus, the p‐Tm:YAG laser appears to be an effective alternative to Ho:YAG and TFL lasers in prostate surgery.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Service d'Urologie, AP-HP, Hôpital Tenon, Sorbonne Université, Paris, France.
- GRC n°20, Groupe de Recherche Clinique sur la Lithiase Urinaire, Hôpital Tenon, Sorbonne Université, Paris, France.
- PIMM, UMR 8006 CNRS, Arts et Métiers ParisTech, Paris, France.
- Service d'Urologie, CHU de Poitiers, Poitiers, France.
- Service d'Urologie, CHU de Limoges, Limoges, France.
- Service d'Anatomopathologie, AP-HP, Hôpital Tenon, Sorbonne Université, Paris, France.
Zitieren
APA
Kutchukian, S., Chicaud, M., Berthe, L., Coste, F., Lapouge, P., Alshehhi, H., Buob, D., Traxer, O., Panthier, F., & Doizi, S. (2024). Comparison of holmium:yttrium‐aluminium‐garnet ( YAG) , thulium fiber laser, and pulsed thulium: YAG lasers on soft tissue: an ex vivo study. BJU International, 134(5), 763-772. https://doi.org/10.1111/bju.16447
Vancouver
Kutchukian S, Chicaud M, Berthe L, Coste F, Lapouge P, Alshehhi H, et al. Comparison of holmium:yttrium‐aluminium‐garnet ( YAG) , thulium fiber laser, and pulsed thulium: YAG lasers on soft tissue: an ex vivo study. BJU International. 2024;134(5):763-772. doi:10.1111/bju.16447
BibTeX
@article{Kutchukian2024,
author = {Kutchukian, Stessy and Chicaud, Marie and Berthe, Laurent and Coste, Frédéric and Lapouge, Pierre and Alshehhi, Hussa and Buob, David and Traxer, Olivier and Panthier, Frédéric and Doizi, Steeve},
title = {Comparison of holmium:yttrium‐aluminium‐garnet ( YAG) , thulium fiber laser, and pulsed thulium: YAG lasers on soft tissue: an ex vivo study},
journal = {BJU International},
year = {2024},
volume = {134},
number = {5},
pages = {763--772},
doi = {10.1111/bju.16447}
}