Fachartikel in der Zeitschrift BJU International, 2023. Verfasst von Jayson P Kemble und 6 weiteren Autor:innen. Beteiligte Einrichtung: Mayo Clinic, Rochester, MN, USA. Hauptthemen laut MEDLINE: Cystoscopes, Carbon Dioxide. Die Arbeit wurde bisher 60-mal zitiert (Crossref) und nennt 27 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To compare the carbon footprint and environmental impact of single‐use and reusable flexible cystoscopes. Materials and Methods We analysed the expected clinical lifecycle of single‐use (Ambu aScope™ 4 Cysto) and reusable (Olympus CYF‐V2) flexible cystoscopes, from manufacture to disposal. Performance data on cumulative procedures between repairs and before decommissioning were derived from a high‐volume multispecialty practice. We estimated carbon expenditures per‐case using published data on endoscope manufacturing, energy consumption during transportation and reprocessing, and solid waste disposal.

Results

A fleet of 16 reusable cystoscopes in service for up to 135 months averaged 207 cases between repairs and 3920 cases per lifecycle. Based on a manufacturing carbon footprint of 11.49 kg CO 2 /kg device for reusable flexible endoscopes and 8.54 kg CO 2 /kg device for single‐use endoscopes, the per‐case manufacturing cost was 1.37 kg CO 2 for single‐use devices and 0.0017 kg CO 2 for reusable devices. The solid mass of single‐use and reusable devices was 0.16 and 0.57 kg, respectively. For reusable devices, the energy consumption of reusable device reprocessing using an automated endoscope reprocessor was 0.20 kg CO 2 , and per‐case costs of device repackaging and repair were 0.005 and 0.02 kg CO 2 , respectively. The total estimated per‐case carbon footprint of single‐use and reusable devices was 2.40 and 0.53 kg CO 2 , respectively, favouring reusable devices.

Conclusion

In this lifecycle analysis, the environmental impact of reusable flexible cystoscopes is markedly less than single‐use cystoscopes. The primary contributor to the per‐case carbon cost of reusable devices is energy consumption of reprocessing.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

CystoscopesCarbon DioxideHumansCystoscopyCarbon FootprintHealth Expenditures

Stichwörter der Autor:innen

carbon footprintcystoscopyendoscopyenvironmental impactflexible cystoscope

Substanzen

Carbon Dioxide

Autor:innen und Institutionen

  • Mayo Clinic, Rochester, MN, USA.
  • Lenox Hill Hospital Northwell Health, New York, NY, USA.
  • Johns Hopkins University School of Medicine, Baltimore, MD, USA.

ORCID-Profile: Jayson P Kemble, Kevin Koo

Zitieren

APA

Kemble, J. P., Winoker, J. S., Patel, S. H., Su, Z. T., Matlaga, B. R., Potretzke, A. M., & Koo, K. (2023). Environmental impact of single‐use and reusable flexible cystoscopes. BJU International. https://doi.org/10.1111/bju.15949

Vancouver

Kemble JP, Winoker JS, Patel SH, Su ZT, Matlaga BR, Potretzke AM, et al. Environmental impact of single‐use and reusable flexible cystoscopes. BJU International. 2023. doi:10.1111/bju.15949

BibTeX
@article{Kemble2023,
  author = {Kemble, Jayson P and Winoker, Jared S and Patel, Sunil H and Su, Zhuo T and Matlaga, Brian R and Potretzke, Aaron M and Koo, Kevin},
  title = {Environmental impact of single‐use and reusable flexible cystoscopes},
  journal = {BJU International},
  year = {2023},
  doi = {10.1111/bju.15949}
}
Kostenlos starten →