Retention of a Detached Robotic Scissors Tip Cover in the Abdominal Wall: A Case Report
Asian Journal of Endoscopic Surgery · 2026-01 · 19 (1)
DOI: 10.1111/ases.70242 ↗Kurz gefasst
Fallbericht in der Zeitschrift Asian Journal of Endoscopic Surgery, online erschienen am 14.01.2026. Verfasst von Masatsugu Kojima und 14 weiteren Autor:innen. Beteiligte Einrichtung: Department of Surgery, Shiga University of Medical Science, Otsu, Shiga, Japan. Hauptthemen laut MEDLINE: Robotic Surgical Procedures, Abdominal Wall, Rectal Neoplasms, Foreign Bodies. Die Arbeit nennt 10 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Robotic surgery has become increasingly widespread; however, device‐related complications specific to robotic platforms are rarely reported. We describe a case of rectal cancer treated with robotic‐assisted abdominoperineal resection using the da Vinci Surgical System. At the end of the procedure, the tip cover of the robotic scissors detached and was inadvertently retained within the abdominal wall. It was invisible on plain radiography but was detected on postoperative computed tomography as a cylindrical structure beneath the rectus abdominis muscle. The patient underwent reoperation to retrieve the tip cover and recovered uneventfully, being discharged without further complications. This case highlights a rare but important complication of robotic surgery—detachment and retention of the tip cover of robotic scissors in the abdominal wall. Because tip covers may not be reliably detected on plain radiography, computed tomography is crucial for their identification. Strict counting protocols and heightened awareness are essential to prevent such events.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Surgery, Shiga University of Medical Science, Otsu, Shiga, Japan.
- Department of Surgery, Nagahama Red Cross Hospital, Nagahama, Shiga, Japan.
- Department of Nursing, Shiga University of Medical Science Hospital, Otsu, Shiga, Japan.
- Department of Radiology, Shiga University of Medical Science Hospital, Otsu, Shiga, Japan.
- Medical Safety Section, Shiga University of Medical Science Hospital, Otsu, Shiga, Japan.
ORCID-Profile: Masatsugu Kojima, Toru Miyake, Sachiko Kaida
Zitieren
APA
Kojima, M., Miyake, T., Tani, S., Muramoto, K., Nishina, Y., Kaida, S., Takebayashi, K., Maehira, H., Otake, R., Mori, H., Nitta, N., Kimura, M., Ushio, N., Shimizu, T., & Tani, M. (2026). Retention of a Detached Robotic Scissors Tip Cover in the Abdominal Wall: A Case Report. Asian Journal of Endoscopic Surgery, 19(1). https://doi.org/10.1111/ases.70242
Vancouver
Kojima M, Miyake T, Tani S, Muramoto K, Nishina Y, Kaida S, et al. Retention of a Detached Robotic Scissors Tip Cover in the Abdominal Wall: A Case Report. Asian Journal of Endoscopic Surgery. 2026;19(1). doi:10.1111/ases.70242
BibTeX
@article{Kojima2026,
author = {Kojima, Masatsugu and Miyake, Toru and Tani, Soichiro and Muramoto, Keiji and Nishina, Yusuke and Kaida, Sachiko and Takebayashi, Katsushi and Maehira, Hiromitsu and Otake, Reiko and Mori, Haruki and Nitta, Nobuhito and Kimura, Miyuki and Ushio, Noritoshi and Shimizu, Tomoharu and Tani, Masaji},
title = {Retention of a Detached Robotic Scissors Tip Cover in the Abdominal Wall: A Case Report},
journal = {Asian Journal of Endoscopic Surgery},
year = {2026},
volume = {19},
number = {1},
doi = {10.1111/ases.70242}
}
EUVIMED