The Physiological Effects of an Adjusted Alarm Architecture on a Neonatal Intensive Care Unit
Acta Paediatrica · 2025-08 · 114 (8) · 1910-1917
DOI: 10.1111/apa.70054 ↗Kurz gefasst
Fachartikel in der Zeitschrift Acta Paediatrica, online erschienen am 06.03.2025. Verfasst von Melissa A C M Kalden und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neonatal and Paediatric Intensive Care, Division of Neonatology, Erasmus MC, University Medical Center - Sophia Children's Hospital, Ro…. Hauptthemen laut MEDLINE: Clinical Alarms, Intensive Care Units, Neonatal. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Themen: Neonatologie, Intensivpflege
Abstract (Originalsprache)
Aim
The effects of using handheld devices in combination with filtering and delaying alarms were investigated. Effects on the number of alarms, patient safety, and nurses' experience were evaluated.
Methods
Alarm and physiological trend data were collected over two periods of three months for a control ( n = 54) and intervention ( n = 47) group. During the intervention period, an adapted alarm architecture, filtering and delaying alarms, was implemented, and the number of alarms, critical cardiorespiratory events, and episodes of decreased oxygen saturation and heart rate were compared to the contemporary alarm architecture. Nurses filled out a survey on their experiences.
Results
The adapted alarm architecture reduced the number of alarms by 84%. This reduction did not result in significant differences in the number of critical events. Additionally, the duration and depth of the patient's episodes of mildly decreased oxygen saturation and heart rate were unaffected. Nurses reported that they continue to receive too many alarms and occasionally miss alarms.
Conclusion
Alarms can be filtered and delayed, reducing the number of alarms and preventing alarm fatigue. Patient safety is not at risk since the number of critical events and the decreases in oxygen saturation and heart rate do not differ significantly between the groups.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Neonatal and Paediatric Intensive Care, Division of Neonatology, Erasmus MC, University Medical Center - Sophia Children's Hospital, Rotterdam, the Netherlands.
- Department of Information and (Medical) Technology, Erasmus MC, University Medical Center, Rotterdam, the Netherlands.
- Department of Biomechanical Engineering, Delft University of Technology, Delft, the Netherlands.
- Department of Adult Intensive Care, Erasmus MC, University Medical Center, Rotterdam, the Netherlands.
ORCID-Profile: Melissa A C M Kalden
Zitieren
APA
Kalden, M. A. C. M., Goos, T. G., Kalden, N., Groenendaal, L. A., Reiss, I. K. M., Van Bommel, J., & Taal, H. R. (2025). The Physiological Effects of an Adjusted Alarm Architecture on a Neonatal Intensive Care Unit. Acta Paediatrica, 114(8), 1910-1917. https://doi.org/10.1111/apa.70054
Vancouver
Kalden MACM, Goos TG, Kalden N, Groenendaal LA, Reiss IKM, Van Bommel J, et al. The Physiological Effects of an Adjusted Alarm Architecture on a Neonatal Intensive Care Unit. Acta Paediatrica. 2025;114(8):1910-1917. doi:10.1111/apa.70054
BibTeX
@article{Kalden2025,
author = {Kalden, Melissa A C M and Goos, Tom G and Kalden, Nico and Groenendaal, Leo A and Reiss, Irwin K M and Van Bommel, Jasper and Taal, H Rob},
title = {The Physiological Effects of an Adjusted Alarm Architecture on a Neonatal Intensive Care Unit},
journal = {Acta Paediatrica},
year = {2025},
volume = {114},
number = {8},
pages = {1910--1917},
doi = {10.1111/apa.70054}
}
EUVIMED