Systematic quantification of outer retinal morphology in three different SD ‐ OCT devices in geographic atrophy
Acta Ophthalmologica · 29.07.2026
DOI: 10.1111/aos.70210 ↗Kurz gefasst
Fachartikel in der Zeitschrift Acta Ophthalmologica, online erschienen am 29.07.2026. Verfasst von Klaudia Birner und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology and Optometry, Medical University of Vienna, Vienna, Austria. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Purpose
To define interdevice differences in quantitative outer retinal biomarkers across three spectral‐domain (SD) optical coherence tomography (OCT) devices in geographic atrophy (GA).
Methods
Patients with GA prospectively underwent imaging using Spectralis (Heidelberg Engineering), Cirrus (Zeiss) and Maestro (Topcon) OCT devices. Ellipsoid zone (EZ) and outer nuclear layer (ONL) thickness, as well as EZ and retinal pigment epithelium (RPE) loss, were automatically pre‐segmented and manually corrected. Quantitative measurements were compared between OCT devices using anova or Friedman tests. Bland–Altman analysis determined mean differences and 95% limits of agreement (LoA).
Results
A total of 7080 B‐scans from 120 OCT volumes of 40 eyes were analysed. Mean patient age was 80 ± 7 years. Mean EZ thickness in the 6‐mm macular cube was 34.8 ± 4.6, 33.4 ± 4.1 and 31.0 ± 4.8 μm for Spectralis, Cirrus and Maestro, respectively, while mean ONL thickness was 77.8 ± 9.2, 78.5 ± 9.4 and 75.8 ± 9.6 μm. Interdevice differences were significant for EZ and ONL‐thickness (all p < 0.05). Mean EZ thickness bias ranged from 1.5 to 3.8 μm, with the widest LoA between Spectralis and Maestro. Mean ONL‐thickness bias ranged from −0.7 to 2.7 μm, with the widest LoA between Cirrus and Maestro. EZ‐loss and EZ/RPE‐loss‐area also differed significantly (both p < 0.01).
Conclusion
Visualization and quantification of atrophic features at the level of the photoreceptors, including EZ thickness, EZ loss and ONL thickness, are influenced by the implemented OCT machine. Consistency in the OCT device during patient monitoring allows for the highest precision and eliminates interdevice differences in quantitative outcome measures.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Ophthalmology and Optometry, Medical University of Vienna, Vienna, Austria.
ORCID-Profile: Klaudia Birner, Anna Eidenberger, Johannes Schrittwieser, Stefan Sacu, Ursula Schmidt-Erfurth
Zitieren
APA
Birner, K., Frank-Publig, S., Mai, J., Eidenberger, A., Schrittwieser, J., Leitner-Barrios, G., Barasits, M., Gumpinger, M., Reiter, G. S., Sacu, S., & Schmidt-Erfurth, U. (2026). Systematic quantification of outer retinal morphology in three different SD ‐ OCT devices in geographic atrophy. Acta Ophthalmologica. https://doi.org/10.1111/aos.70210
Vancouver
Birner K, Frank-Publig S, Mai J, Eidenberger A, Schrittwieser J, Leitner-Barrios G, et al. Systematic quantification of outer retinal morphology in three different SD ‐ OCT devices in geographic atrophy. Acta Ophthalmologica. 2026. doi:10.1111/aos.70210
BibTeX
@article{Birner2026,
author = {Birner, Klaudia and Frank-Publig, Sophie and Mai, Julia and Eidenberger, Anna and Schrittwieser, Johannes and Leitner-Barrios, Gabriel and Barasits, Magdalena and Gumpinger, Markus and Reiter, Gregor S and Sacu, Stefan and Schmidt-Erfurth, Ursula},
title = {Systematic quantification of outer retinal morphology in three different SD ‐ OCT devices in geographic atrophy},
journal = {Acta Ophthalmologica},
year = {2026},
doi = {10.1111/aos.70210}
}
EUVIMED