Self‐monitoring of the intraocular pressure with ICARE® HOME2 and a smartphone application: Feasibility and acceptability
Acta Ophthalmologica · 2025-02 · 103 (1) · 109-114
DOI: 10.1111/aos.16789 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift Acta Ophthalmologica, online erschienen am 12.11.2024. Verfasst von Laurence Quérat und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anterior Segment Disorders, Glaucoma, Neuro-Ophthalmology and Oculoplastics, St. Erik Eye Hospital, Solna, Sweden. Hauptthemen laut MEDLINE: Intraocular Pressure, Tonometry, Ocular, Smartphone, Mobile Applications. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 13 Literaturangaben.
Abstract (Originalsprache)
Purpose
To evaluate the feasibility and acceptability of self‐tonometry by glaucoma patients using iCare® Home2 (ICH2) and a smartphone application, without healthcare personnel supervision.
Methods
After downloading a smartphone application, participants learned to handle the ICH2 via an instruction film, measured their intraocular pressure (IOP) and transferred the data to the caregiver via the application. Participants' measurements were compared to healthcare personnel measurements using Goldmann applanation tonometry (GAT). Participants answered questionnaires about their experience of self‐tonometry.
Results
Twenty‐eight participants were included in the study and three of them dropped off. There was no statistically significant difference between the last measurements obtained with ICH2 by the participants and the GAT measurements for the right and the left eye, p = 0.781 and p = 0.888 respectively. The inter‐user agreement was good to excellent between the last measurements obtained with ICH2 by the participants and the GAT measurements for the right and the left eye (ICC 0.907 and 0.830 respectively). The overall impression of the participants was very positive, with 92% happy (a lot or completely) with the instruction film and 88% happy (a lot or completely) with the smartphone application. Thirty‐six percent would have preferred to have an instructor by their side.
Conclusion
Our study showed the feasibility and acceptability of self‐tonometry with iCare® Home2 and a smartphone application without the supervision of a healthcare provider in a group of motivated glaucoma patients.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Anterior Segment Disorders, Glaucoma, Neuro-Ophthalmology and Oculoplastics, St. Erik Eye Hospital, Solna, Sweden.
- Division of Eye and Vision, Department of Clinical Neuroscience, Karolinska Institute, Solna, Sweden.
ORCID-Profile: Laurence Quérat
Zitieren
APA
Quérat, L., & Chen, E. (2025). Self‐monitoring of the intraocular pressure with ICARE® HOME2 and a smartphone application: Feasibility and acceptability. Acta Ophthalmologica, 103(1), 109-114. https://doi.org/10.1111/aos.16789
Vancouver
Quérat L, Chen E. Self‐monitoring of the intraocular pressure with ICARE® HOME2 and a smartphone application: Feasibility and acceptability. Acta Ophthalmologica. 2025;103(1):109-114. doi:10.1111/aos.16789
BibTeX
@article{Qurat2025,
author = {Quérat, Laurence and Chen, Enping},
title = {Self‐monitoring of the intraocular pressure with ICARE® HOME2 and a smartphone application: Feasibility and acceptability},
journal = {Acta Ophthalmologica},
year = {2025},
volume = {103},
number = {1},
pages = {109--114},
doi = {10.1111/aos.16789}
}
EUVIMED