Comparison of perimetric 24‐2 and 30‐2 test patterns in detecting visual field defects in patients with tumours in the pituitary region

Bodil Gesslein iD, Magdalena Naumovska, Olof Neumann, Dimitrios Bizios, Boel Bengtsson iD, Peter Siesjö, Erik Uvelius, Björn Hammar, Rafi Sheikh iD

Acta Ophthalmologica · 2024-05 · 102 (3) · 326-333

DOI: 10.1111/aos.15731 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Acta Ophthalmologica, online erschienen am 14.07.2023. Verfasst von Bodil Gesslein und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Department of Clinical Sciences Lund, Lund University, Skåne University Hospital, Lund, Sweden. Hauptthemen laut MEDLINE: Visual Field Tests, Pituitary Neoplasms. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Skåne County Council Research Grants, Skåne University Hospital (SUS) Research Grants, The Cronqvist Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Automated perimetry provides a standardized method of measuring the visual field. The Humphrey Field Analyser (HFA) uses the 24‐2 test pattern to cover 24 degrees centrally or the 30‐2 test pattern to cover a slightly broader region of 30 degrees. The aim of this study was to determine whether the 24‐2 test pattern provides comparable information to the 30‐2 test pattern in detecting visual field defects in patients with tumours in the pituitary region.

Methods

A retrospective cohort study was carried out on patients with tumours in the pituitary region and radiologically confirmed compression of the visual pathway. Included patients (79 of 133) had been examined using the Humphrey 30‐2 visual field test, after which the 30‐2 test patterns were reduced into corresponding 24‐2 test patterns. The location of visual field defects, visual acuity and the perimetric parameters mean deviation (MD) and visual field index (VFI) were also recorded.

Results

No patient was classified differently when evaluated with the 24‐2 test pattern, compared to the 30‐2 test pattern. Interestingly, although the majority of patients had visual field defects located in the temporal visual field of each eye, a significant minority did not. In addition, it was found that a large proportion of patients had normal visual acuity (≥0.8).

Conclusions

The use of the HFA 24‐2 test pattern reliably detected visual field defects in patients with tumours in the pituitary region. The present study indicates that MD and VFI are not reliable parameters for evaluating visual field defects due to compression.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Visual Field TestsPituitary NeoplasmsHumansVisual FieldsRetrospective StudiesVision Disorders

Stichwörter der Autor:innen

30‐2 and 24‐2 test patternHFAperimetrypituitary adenomapituitary tumoursvisual field defects

Autor:innen und Institutionen

  • Department of Ophthalmology, Department of Clinical Sciences Lund, Lund University, Skåne University Hospital, Lund, Sweden.
  • Department of Ophthalmology, Department of Clinical Sciences Malmö, Lund University, Skåne University Hospital, Malmö, Sweden.
  • Department of Neurosurgery, Department of Clinical Sciences Lund, Lund University, Skåne University Hospital, Lund, Sweden.

ORCID-Profile: Bodil Gesslein, Boel Bengtsson, Rafi Sheikh

Förderung

  • Skåne County Council Research Grants
  • Skåne University Hospital (SUS) Research Grants
  • The Cronqvist Foundation
  • The Swedish Eye Foundation
  • The Swedish Government Grant for Clinical Research (ALF)
  • The Swedish Medical Association
  • Society of University Surgeons

Zitieren

APA

Gesslein, B., Naumovska, M., Neumann, O., Bizios, D., Bengtsson, B., Siesjö, P., Uvelius, E., Hammar, B., & Sheikh, R. (2024). Comparison of perimetric 24‐2 and 30‐2 test patterns in detecting visual field defects in patients with tumours in the pituitary region. Acta Ophthalmologica, 102(3), 326-333. https://doi.org/10.1111/aos.15731

Vancouver

Gesslein B, Naumovska M, Neumann O, Bizios D, Bengtsson B, Siesjö P, et al. Comparison of perimetric 24‐2 and 30‐2 test patterns in detecting visual field defects in patients with tumours in the pituitary region. Acta Ophthalmologica. 2024;102(3):326-333. doi:10.1111/aos.15731

BibTeX
@article{Gesslein2024,
  author = {Gesslein, Bodil and Naumovska, Magdalena and Neumann, Olof and Bizios, Dimitrios and Bengtsson, Boel and Siesjö, Peter and Uvelius, Erik and Hammar, Björn and Sheikh, Rafi},
  title = {Comparison of perimetric 24‐2 and 30‐2 test patterns in detecting visual field defects in patients with tumours in the pituitary region},
  journal = {Acta Ophthalmologica},
  year = {2024},
  volume = {102},
  number = {3},
  pages = {326--333},
  doi = {10.1111/aos.15731}
}
Kostenlos starten →