Auf einen Blick

Sechs nichtinvasive Algorithmen zur Schätzung der Stroke Work wurden bei 22 Schweinen mit LVAD verglichen. Kein Algorithmus war unter allen Bedingungen optimal.

Fragestellung

Wie gut schätzen sechs nichtinvasive Algorithmen die linksventrikuläre Stroke Work aus LVAD-Parametern unter verschiedenen hämodynamischen Bedingungen?

Methode

Vergleich von 6 Algorithmen bei 22 gesunden Deutschen-Landrasse-Schweinen mit Deltastream DP3 LVAD; nach isofluraninduzierter Herzinsuffizienz wurden 4 Interventionen geprüft.

Ergebnisse

Signifikante Haupteffekte für Algorithmus, Intervention und ihre Interaktion; 54,04 % der Variabilität erklärt. Druckkopf-Fluss zeigte konsistent hohe Korrelationen, Pumpenflussfläche war robust außer bei Dobutamin, systolischer Anteil bei Dobutamin am besten.

Einschränkungen

Die Druckkopf-Fluss-Methode benötigt semi-invasive Druckmessungen; der systolische Anteil scheiterte häufig an der mathematischen Optimierung.

Schlussfolgerung der Autor:innen

Kein einzelner Algorithmus war unter allen Bedingungen optimal; die Auswahl sollte situationsabhängig erfolgen. Die ausgeprägte interindividuelle Variabilität spricht für patientenspezifische Faktoren.

Kernaussagen

  • 6 Algorithmen wurden in 22 Schweinen mit LVAD verglichen
  • Kein Algorithmus war unter allen Bedingungen am besten
  • Patientenspezifische Faktoren schienen die Ergebnisse stark zu beeinflussen

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Fachartikel in der Zeitschrift Artificial Organs, online erschienen am 10.11.2026. Verfasst von Moriz A Habigt und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anaesthesiology, Medical Faculty, RWTH Aachen University, Aachen, Germany. Die Arbeit nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Bundesministerium für Forschung und Technologie.

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Accurate assessment of cardiac work during left ventricular assist device (LVAD) therapy is essential for patient management, device explantation timing, and adaptive control algorithm development. While invasive pressure–volume methods provide accurate work and contractility assessment, their clinical applicability is limited. Non‐invasive algorithms utilizing inherent pump parameters offer a promising alternative, yet no systematic comparison exists.

Methods

Six non‐invasive algorithms for estimating stroke work (SW) were compared in 22 healthy German Landrace pigs (48.9 ± 2.6 kg) with a Deltastream DP3 LVAD: maximal LVAD flow derivative, motor power–LV pressure relationship, pressure head–flow relationship, systolic fraction of pressure–flow area, and total/pulsatile components of pump current and flow. After isoflurane‐induced heart failure, four interventions were performed: preload reduction (inferior vena cava occlusion, n = 105), afterload increase (aortic balloon occlusion, n = 97), dobutamine infusion ( n = 21), and speed ramps ( n = 14). Linear multilevel modeling revealed significant main effects for algorithm ( χ 2 = 582.38), intervention type ( χ 2 = 73.77), and their interaction ( χ 2 = 324.80; all p < 0.001), explaining 54.04% of variability.

Results

The pressure head‐flow algorithm demonstrated consistently high correlations across interventions, though requiring semi‐invasive pressure measurements. The pump flow area algorithm performed robustly during preload, afterload, and speed ramps but failed to capture SW changes during dobutamine. The systolic fraction algorithm achieved the highest correlations during dobutamine but frequently failed mathematical optimization, limiting applicability.

Conclusion

No single algorithm performed optimally across all conditions, necessitating context‐specific selection. Pronounced inter‐individual variability underscores the importance of patient‐specific factors. These findings provide the first systematic comparison of non‐invasive SW algorithms during LVAD therapy, offering guidance for algorithm selection.

Schlagwörter

Stichwörter der Autor:innen

cardiac contractilityhemodynamic assessmentleft ventricular assist devicenon‐invasive monitoringpump parametersstroke work

Autor:innen und Institutionen

  • Department of Anaesthesiology, Medical Faculty, RWTH Aachen University, Aachen, Germany.
  • Institute of Automatic Control, RWTH Aachen University, Aachen, Germany.
  • Machine Learning Group, Cognitive Interaction Technology, Bielefeld University, Bielefeld, Germany.

ORCID-Profile: Moriz A Habigt, Marc Hein, Frederic Olbrich, Alexander Schulz, Barbara Hammer, Sebastian Billig

Förderung

  • Bundesministerium für Forschung und Technologie

Zitieren

APA

Habigt, M. A., Hein, M., Mechelinck, M., Olbrich, F., Stemmler, M., Schulz, A., Hammer, B., & Billig, S. (2026). In Vivo Comparison of Algorithms Estimating Left Ventricular Stroke Work From LVAD ‐Derived Pump Parameters. Artificial Organs. https://doi.org/10.1111/aor.70277

Vancouver

Habigt MA, Hein M, Mechelinck M, Olbrich F, Stemmler M, Schulz A, et al. In Vivo Comparison of Algorithms Estimating Left Ventricular Stroke Work From LVAD ‐Derived Pump Parameters. Artificial Organs. 2026. doi:10.1111/aor.70277

BibTeX
@article{Habigt2026,
  author = {Habigt, Moriz A and Hein, Marc and Mechelinck, Mare and Olbrich, Frederic and Stemmler, Maike and Schulz, Alexander and Hammer, Barbara and Billig, Sebastian},
  title = {In Vivo Comparison of Algorithms Estimating Left Ventricular Stroke Work From LVAD ‐Derived Pump Parameters},
  journal = {Artificial Organs},
  year = {2026},
  doi = {10.1111/aor.70277}
}
Kostenlos starten →