Large‐Eddy Simulation of the FDA Benchmark Blood Pump: Validation Against Experiments and Implications for Turbulent Flow Mechanisms
Artificial Organs · 2026
DOI: 10.1111/aor.70190 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Artificial Organs, 2026. Verfasst von Xuanming Huang und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanics and Aerospace Engineering, Southern University of Science and Technology, Shenzhen, Guangdong, China. Hauptthemen laut MEDLINE: Heart-Assist Devices, Computer Simulation, Models, Cardiovascular. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 38 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Shenzhen Science and Technology Program, The Department of Science and Technology of Guangdong Province.
Abstract (Originalsprache)
Background
This study presents a systematic validation and comparative assessment of computational fluid dynamics (CFD) strategies for centrifugal blood pump simulations using the US Food and Drug Administration benchmark model.
Methods
A scale‐resolving large eddy simulation (LES) with transient sliding‐interface (SI) coupling is evaluated and compared against Reynolds‐averaged Navier–Stokes (RANS) approaches employing both multiple reference frame and SI formulations. Numerical predictions are validated through direct comparison with particle image velocimetry measurements under two representative operating conditions.
Results
The results indicate that LES with transient rotor‐stator coupling achieves consistently improved agreement with experimental velocity fields compared with RANS‐based methods, particularly in the diffuser region where strong intermittency and wall‐bounded turbulence are present. In contrast, RANS‐based approaches exhibit noticeable discrepancies in these regions. A mesh sensitivity study and an assessment of temporal averaging effects are conducted for LES. The quality of the LES results is further quantified using three complementary metrics, demonstrating that a mesh resolution of approximately 80 million cells achieves a well‐resolved LES regime. Building on the validated scale‐resolving simulations, detailed analyses of vortical structures, turbulent kinetic energy distributions, and velocity energy spectra are performed to characterize the internal flow physics of the pump. The results reveal the central role of transient vortex dynamics in turbulence generation within centrifugal blood pumps.
Conclusions
Overall, this study demonstrates that scale‐resolving, transient simulation approaches are essential for accurately capturing the highly unsteady, turbulence‐dominated flow features in ventricular assist devices and provides practical guidance for future high‐fidelity hemodynamic and hemocompatibility studies.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Mechanics and Aerospace Engineering, Southern University of Science and Technology, Shenzhen, Guangdong, China.
- Department of Mechanical Engineering (Robotics), Guangdong Technion-Israel Institute of Technology, Shantou, Guangdong, China.
ORCID-Profile: Xuanming Huang, Chi Ding, Yujie Sun, Shidi Huang, Andrea Cioncolini, Damiano Padovani, Ju Liu
Förderung
- National Natural Science Foundation of China
- Shenzhen Science and Technology Program
- The Department of Science and Technology of Guangdong Province
- Center for Computational Science and Engineering at the Southern University of Science and Technology
Zitieren
APA
Huang, X., Ding, C., Sun, Y., Huang, S., Cioncolini, A., Padovani, D., & Liu, J. (2026). Large‐Eddy Simulation of the FDA Benchmark Blood Pump: Validation Against Experiments and Implications for Turbulent Flow Mechanisms. Artificial Organs. https://doi.org/10.1111/aor.70190
Vancouver
Huang X, Ding C, Sun Y, Huang S, Cioncolini A, Padovani D, et al. Large‐Eddy Simulation of the FDA Benchmark Blood Pump: Validation Against Experiments and Implications for Turbulent Flow Mechanisms. Artificial Organs. 2026. doi:10.1111/aor.70190
BibTeX
@article{Huang2026,
author = {Huang, Xuanming and Ding, Chi and Sun, Yujie and Huang, Shidi and Cioncolini, Andrea and Padovani, Damiano and Liu, Ju},
title = {Large‐Eddy Simulation of the FDA Benchmark Blood Pump: Validation Against Experiments and Implications for Turbulent Flow Mechanisms},
journal = {Artificial Organs},
year = {2026},
doi = {10.1111/aor.70190}
}
EUVIMED