Kurz gefasst

Fallbericht in der Zeitschrift Artificial Organs, 2026. Verfasst von Bhagyashri Surve und 4 weiteren Autor:innen. Beteiligte Einrichtung: Tata Memorial Centre, Advanced Centre for Treatment, Research & Education in Cancer, HBNI, Navi Mumbai, India. Hauptthemen laut MEDLINE: Plateletpheresis, Blood Donors, Blood Component Removal. Die Arbeit nennt 10 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

With the rising demand for platelet products, the use of apheresis‐derived single donor platelets (SDPs) has grown substantially. While plateletpheresis is generally considered safe, it is not devoid of complications. Materials and Methods This retrospective descriptive study reviews three adverse events that occurred during SDP collection at a transfusion medicine department. Root cause analysis (RCA) was employed using tools such as the 5 “Whys” technique and the Ishikawa (Fishbone) Diagram to identify underlying causes and propose preventive strategies. Case Details In the first incident, a first‐time donor experienced vasovagal symptoms triggered by a visible blood leak from the apheresis kit. The second incident involved a centrifuge alarm and mid‐procedure halt due to blood spillage caused by improper kit installation. The third incident was identified post‐procedure when the platelet concentrate was found contaminated with red blood cells from a manufacturing defect. All donors recovered without lasting health effects; however, such technical or procedural failures can erode donor confidence, reduce willingness for repeat donations, and impact retention of valuable apheresis donors.

Conclusion

The events in this case series highlight that adverse outcomes can stem from both equipment‐related faults and human errors. Strengthening the voluntary apheresis donor base in India requires prioritizing donor safety, ensuring positive donation experiences, comprehensive staff training, adherence to standard operating procedures (SOPs), and fostering a supportive, trust‐based environment.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PlateletpheresisBlood DonorsBlood Component RemovalHumansRetrospective StudiesSyncope, VasovagalBlood DonationFemaleMalePlatelet TransfusionEquipment Failure

Stichwörter der Autor:innen

Ishikawa diagramdonor retentiononcology centerplateletpheresisroot cause analysistechnical malfunctionthe 5 “whys” technique

Autor:innen und Institutionen

  • Tata Memorial Centre, Advanced Centre for Treatment, Research & Education in Cancer, HBNI, Navi Mumbai, India.

ORCID-Profile: Chitra Dewanji, Aditi Khanna

Zitieren

APA

Surve, B., Dewanji, C., Khanna, A., Saha, S., & Ojha, S. (2026). When Technology Falters: Ensuring Donor Safety and System Reliability Amid Apheresis Technical Challenges. Artificial Organs. https://doi.org/10.1111/aor.70097

Vancouver

Surve B, Dewanji C, Khanna A, Saha S, Ojha S. When Technology Falters: Ensuring Donor Safety and System Reliability Amid Apheresis Technical Challenges. Artificial Organs. 2026. doi:10.1111/aor.70097

BibTeX
@article{Surve2026,
  author = {Surve, Bhagyashri and Dewanji, Chitra and Khanna, Aditi and Saha, Suryatapa and Ojha, Shashank},
  title = {When Technology Falters: Ensuring Donor Safety and System Reliability Amid Apheresis Technical Challenges},
  journal = {Artificial Organs},
  year = {2026},
  doi = {10.1111/aor.70097}
}
Kostenlos starten →