EUVIMEDEUVIMED entdecken
StudienprotokollOpen Access · CC BY-NC-ND 4.0

Development of a Protocol for Long‐Term Ex Vivo Normothermic Machine Perfusion of Rodent Livers

Mark Ly iD, Daniel Babekuhl, Anita Niu iD, Paul Yousif, Chuanmin Wang, Ngee-Soon Lau iD, Mark D Gorrell, Geoffrey W McCaughan, Michael Crawford, Carlo Pulitano iD

Artificial Organs · 2025

DOI: 10.1111/aor.15024 ↗

Kurz gefasst

Studienprotokoll in der Zeitschrift Artificial Organs, 2025. Verfasst von Mark Ly und 9 weiteren Autor:innen. Beteiligte Einrichtung: Australian National Liver Transplantation Unit, Royal Prince Alfred Hospital, Sydney, New South Wales, Australia. Hauptthemen laut MEDLINE: Perfusion, Liver, Organ Preservation, Liver Transplantation. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Royal Prince Alfred Transplant Institute.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Long‐term normothermic machine perfusion (LT‐NMP) enables the assessment and optimization of livers for days and, potentially, weeks. However, models of LT‐NMP have only been described for human and pig livers, which are resource intensive and impractical for laboratory research. Cost‐effective small animal models of LT‐NMP are needed for future research. This study aimed to develop a system for LT‐NMP of rat livers for up to 72 h.

Methods

This study was performed in two stages: the development phase ( n = 20) and validation phase ( n = 5). The perfusion system included an organ reservoir, pump, heat exchanger, long‐term oxygenator, and dialysis. Hormonal and nutritional support were continuously infused. During the validation phase, five consecutive grafts were perfused using our protocol. At 72 h postreperfusion, grafts were assessed for viability, which was based on hemodynamic stability, mitochondrial function, bile production, and metabolic activity.

Results

Rodent livers were supported up to 107 h using our LT‐NMP protocol. All grafts in the validation phase remained viable at 72 h ( n = 5/5). The median oxygen consumption and bile production at 72 h were 0.079 mLO 2 /min/g‐liver and 8.6 uL/h/g‐liver, respectively. All grafts had a systemic vascular resistance less than 0.25 mmHg/mL/min. Metabolic activity, defined as lactate clearance, glucose production, or response to glucagon, was observed in all grafts (5/5).

Conclusions

This is the first study to report LT‐NMP of rodent livers up to 5 days. Using our protocol, rat livers could reliably be supported until 72 h. This model provides a greater opportunity to investigate novel therapeutics to assess, optimize, and regenerate liver grafts for transplantation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PerfusionLiverOrgan PreservationLiver TransplantationAnimalsRatsMaleOxygen Consumption

Stichwörter der Autor:innen

liver transplantationlong‐term normothermic machine perfusionmachine perfusionorgan preservationrat

Autor:innen und Institutionen

  • Australian National Liver Transplantation Unit, Royal Prince Alfred Hospital, Sydney, New South Wales, Australia.
  • Centre for Organ Assessment Repair and Optimisation, Royal Prince Alfred Hospital, Sydney, New South Wales, Australia.
  • Centenary Institute, Faculty of Medicine and Health, The University of Sydney, Sydney, New South Wales, Australia.
  • Faculty of Medicine and Health, The University of Sydney, Sydney, New South Wales, Australia.
  • Faculty of Engineering, The University of Sydney, Sydney, New South Wales, Australia.

ORCID-Profile: Mark Ly, Anita Niu, Ngee-Soon Lau, Carlo Pulitano

Förderung

  • Royal Prince Alfred Transplant Institute

Zitieren

APA

Ly, M., Babekuhl, D., Niu, A., Yousif, P., Wang, C., Lau, N. S., Gorrell, M. D., McCaughan, G. W., Crawford, M., & Pulitano, C. (2025). Development of a Protocol for Long‐Term Ex Vivo Normothermic Machine Perfusion of Rodent Livers. Artificial Organs. https://doi.org/10.1111/aor.15024

Vancouver

Ly M, Babekuhl D, Niu A, Yousif P, Wang C, Lau NS, et al. Development of a Protocol for Long‐Term Ex Vivo Normothermic Machine Perfusion of Rodent Livers. Artificial Organs. 2025. doi:10.1111/aor.15024

BibTeX
@article{Ly2025,
  author = {Ly, Mark and Babekuhl, Daniel and Niu, Anita and Yousif, Paul and Wang, Chuanmin and Lau, Ngee-Soon and Gorrell, Mark D and McCaughan, Geoffrey W and Crawford, Michael and Pulitano, Carlo},
  title = {Development of a Protocol for Long‐Term Ex Vivo Normothermic Machine Perfusion of Rodent Livers},
  journal = {Artificial Organs},
  year = {2025},
  doi = {10.1111/aor.15024}
}
Kostenlos starten →