Normothermic Machine Perfusion vs. Static Cold Storage in Liver Transplantation: A Systematic Review and Meta‐Analysis
Artificial Organs · 2025-06 · 49 (6) · 945-954
DOI: 10.1111/aor.14960 ↗Kurz gefasst
Systematische Übersicht in der Zeitschrift Artificial Organs, online erschienen am 30.01.2025. Verfasst von Patricia Viana und 6 weiteren Autor:innen. Beteiligte Einrichtung: University of Extreme South of Santa Catarina, Criciuma, Brazil. Hauptthemen laut MEDLINE: Liver Transplantation, Organ Preservation, Perfusion. Die Arbeit wurde bisher 16-mal zitiert (Crossref) und nennt 55 Literaturangaben.
Abstract (Originalsprache)
Background
Normothermic machine perfusion (NMP) represents an alternative to prolong liver preservation and reduce organ discard rates. We performed an updated systematic review and meta‐analysis to compare NMP with static cold storage (SCS) in liver transplantation.
Methods
MEDLINE, Embase, and Cochrane were searched for randomized controlled trials (RCTs) or observational studies. Risk ratios (RR) and mean differences were calculated. p < 0.05 was considered significant. A random‐effects model was applied for all outcomes. PROSPERO ID: CRD42023486184.
Results
We included 1295 patients from 5 RCTs and 6 observational studies from 2016 to 2023. 592 (45.7%) underwent NMP. A subgroup RCT analysis favored NMP for non‐anastomotic strictures (RR 0.4; 95% CI 0.2, 0.9), postreperfusion syndrome (RR 0.4; 95% CI 0.27, 0.56), and early allograft dysfunction (RR 0.6; 95% CI 0.4, 0.9). NMP favored higher organ utilization rates (RR 1.1; 95% CI 1.02, 1.18). No significant differences between NMP and SCS were observed in graft survival or patient survival at 12 months, primary non‐function, serious adverse events, overall biliary complications, AST, or bilirubin levels peak within the first 7 days, ICU or hospital length of stay.
Conclusion
Our findings suggest that NMP is associated with lower non‐anastomotic biliary stricture rates, postreperfusion syndrome, early allograft dysfunction, and higher organ utilization in the RCT subgroup analysis, without increasing adverse events.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- University of Extreme South of Santa Catarina, Criciuma, Brazil.
- Universidad Peruana Cayetano Heredia, Lima, Peru.
- Univeristat Internacional de Catalunya, Barcelona, Spain.
- Souza Marques Medical School, Rio de Janeiro, Brazil.
- Division of Organ Transplantation, Department of Surgery, University of Massachusetts, Worcester, Massachusetts, USA.
- Division of Transplant Surgery, Department of Surgery, University of Texas Medical Branch, Galveston, Texas, USA.
ORCID-Profile: Patricia Viana, Maria M R Mora, Thamiris D D Cabral, Paulo N Martins, Michael Kueht, Isabella Faria
Zitieren
APA
Viana, P., Castillo-Flores, S., Mora, M. M. R., Cabral, T. D. D., Martins, P. N., Kueht, M., & Faria, I. (2025). Normothermic Machine Perfusion vs. Static Cold Storage in Liver Transplantation: A Systematic Review and Meta‐Analysis. Artificial Organs, 49(6), 945-954. https://doi.org/10.1111/aor.14960
Vancouver
Viana P, Castillo-Flores S, Mora MMR, Cabral TDD, Martins PN, Kueht M, et al. Normothermic Machine Perfusion vs. Static Cold Storage in Liver Transplantation: A Systematic Review and Meta‐Analysis. Artificial Organs. 2025;49(6):945-954. doi:10.1111/aor.14960
BibTeX
@article{Viana2025,
author = {Viana, Patricia and Castillo-Flores, Samy and Mora, Maria M R and Cabral, Thamiris D D and Martins, Paulo N and Kueht, Michael and Faria, Isabella},
title = {Normothermic Machine Perfusion vs. Static Cold Storage in Liver Transplantation: A Systematic Review and Meta‐Analysis},
journal = {Artificial Organs},
year = {2025},
volume = {49},
number = {6},
pages = {945--954},
doi = {10.1111/aor.14960}
}