Real‐time physiological environment emulation for the Istanbul heart ventricular assist device via acausal cardiovascular modeling
Artificial Organs · 2025
DOI: 10.1111/aor.14903 ↗Kurz gefasst
Fachartikel in der Zeitschrift Artificial Organs, 2025. Verfasst von Hammad Ur Rahman und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, Koç University, Istanbul, Turkey. Hauptthemen laut MEDLINE: Heart-Assist Devices, Models, Cardiovascular, Heart Failure. Die Arbeit nennt 25 Literaturangaben. Gefördert von: Türkiye Bilimsel ve Teknolojik Araştırma Kurumu.
Abstract (Originalsprache)
Background and Objectives The cost and complexity associated with animal testing are significantly reduced by using mock circulatory loops prior. Novel mock circulatory loops allow us to test biomedical devices preclinically due to their flexibility, scalability, and cost‐effectiveness. The presented work describes the development of a hardware‐in‐the‐loop platform to emulate human physiology for the Istanbul Heart (iHeart‐II) LVAD.
Methods
A closed‐loop system is developed whereby the effect of the LVAD on the heart and vice versa can be studied. An acausal model of the cardiovascular system is calibrated to emulate advanced‐stage heart failure. A new prototype of the iHeart‐II LVAD is connected between two air‐actuated chambers emulating the left ventricle and aortic chambers with PID controllers tracking numerically modeled pressures from the in silico model. A lead–lag compensator is used to maintain fluid level. Controllers are tuned using nonlinear Hammerstein‐Weiner models identified using open‐loop data. The iHeart‐II LVAD is operated at various speeds in its operational range, and the resulting hemodynamics are visualized in real time.
Results
Hemodynamic variables, such as LVAD flow rate, aortic, left ventricular, and pulse pressure, demonstrate trends similar to clinical observations. The iHeart‐II LVAD achieves hemodynamic normalization at ~3500 rpm for the emulated condition.
Conclusions
A novel evaluation methodology is adopted to study the performance of the iHeart LVAD under advanced‐stage heart failure emulation. The models and controllers used in the platform are readily replicable to facilitate VAD research, pedagogy, design, and development.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Mechanical Engineering, Koç University, Istanbul, Turkey.
- School of Medicine, Cardiovascular Surgery Department, Başkent University, Istanbul, Turkey.
ORCID-Profile: Hammad Ur Rahman, Khunsha Mehmood, Ismail Lazoglu
Förderung
- Türkiye Bilimsel ve Teknolojik Araştırma Kurumu
Zitieren
APA
Rahman, H. U., Mehmood, K., Abdulhamid, F., Lazoglu, I., Bakuy, V., & Küçükaksu, D. S. (2025). Real‐time physiological environment emulation for the Istanbul heart ventricular assist device via acausal cardiovascular modeling. Artificial Organs. https://doi.org/10.1111/aor.14903
Vancouver
Rahman HU, Mehmood K, Abdulhamid F, Lazoglu I, Bakuy V, Küçükaksu DS. Real‐time physiological environment emulation for the Istanbul heart ventricular assist device via acausal cardiovascular modeling. Artificial Organs. 2025. doi:10.1111/aor.14903
BibTeX
@article{Rahman2025,
author = {Rahman, Hammad Ur and Mehmood, Khunsha and Abdulhamid, Farouk and Lazoglu, Ismail and Bakuy, Vedat and Küçükaksu, Deniz Süha},
title = {Real‐time physiological environment emulation for the Istanbul heart ventricular assist device via acausal cardiovascular modeling},
journal = {Artificial Organs},
year = {2025},
doi = {10.1111/aor.14903}
}