A comparative in vitro study of distinct and novel stent geometries on mechanical performances of poly‐L‐lactic acid cardiovascular stents
Artificial Organs · 2025
DOI: 10.1111/aor.14893 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Artificial Organs, 2025. Verfasst von Jinwoo Kim und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Bionanosystem Engineering, Graduate School, Jeonbuk National University, Jeonju, Republic of Korea. Hauptthemen laut MEDLINE: Polyesters, Stents, Absorbable Implants. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 42 Literaturangaben. Gefördert von: Korea Medical Device Development Fund.
Abstract (Originalsprache)
Background
Poly‐L‐lactic acid (PLLA) is one of the representative polymeric materials serving as bioresorbable stents (BRS) for cardiovascular disease due to its proper biodegradation, high biocompatibility, and adequate mechanical properties among polymer candidates for BRS. However, PLLA BRS as cardiovascular stents also have limitations because their mechanical properties including low radial strength and high elastic recoil are inferior to those of metallic‐based BRS stents.
Methods
In the study, we developed and manufactured distinct and novel types of stent geometries for investigating mechanical properties of thin‐walled PLLA BRS (110 μm) for cardiovascular applications. Five key mechanical tests, including radial strength, crimping profile, flexibility, elastic recoil, and foreshortening were performed through a comprehensive analysis. In addition, we applied the finite element method for further validation and insight of mechanical behaviors of the PLLA BRS.
Results
Results revealed that Model 2 had advantages in high flexibility as well as radial strengths, which would be a proper option for complex and acutely curved lesions. Model 3 would be an optimum selection for stent placement in mild target site due to its strength in minimum elastic recoil. Even though Model 4 showed the highest radial strength, finite element simulation showed that the geometry caused higher maximum stress than that of Model 2 and Model 3 during the crimping process. Model 1 showed the most vulnerable geometry among the tested models in both in vitro and finite element analysis.
Conclusion
Such data may suggest potential guidance in regard to understanding the mechanical behaviors of PLLA BRS as not only applicable cardiovascular but also peripheral and intracranial stents.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Bionanosystem Engineering, Graduate School, Jeonbuk National University, Jeonju, Republic of Korea.
- Division of Mechanical Design Engineering, Jeonbuk National University, Jeonju, Republic of Korea.
- Innovative Mechanobio Active Materials Based Medical Device Demonstration Center, Jeonbuk National University, Jeonju, Republic of Korea.
- Department of Bionanotechnology and Bioconvergence Engineering, Graduate School, Jeonbuk National University, Jeonju, Republic of Korea.
- CGBio Co. Ltd., Seongnam, Republic of Korea.
ORCID-Profile: Chan Hee Park
Förderung
- Korea Medical Device Development Fund
Zitieren
APA
Kim, J., Lee, J. H., Choi, E. A., Lee, H. J., Oh, J., Byeon, D. H., & Park, C. H. (2025). A comparative in vitro study of distinct and novel stent geometries on mechanical performances of poly‐L‐lactic acid cardiovascular stents. Artificial Organs. https://doi.org/10.1111/aor.14893
Vancouver
Kim J, Lee JH, Choi EA, Lee HJ, Oh J, Byeon DH, et al. A comparative in vitro study of distinct and novel stent geometries on mechanical performances of poly‐L‐lactic acid cardiovascular stents. Artificial Organs. 2025. doi:10.1111/aor.14893
BibTeX
@article{Kim2025,
author = {Kim, Jinwoo and Lee, Jung Ho and Choi, Eun Ae and Lee, Hyeon Ji and Oh, Jin and Byeon, Dae-Heung and Park, Chan Hee},
title = {A comparative in vitro study of distinct and novel stent geometries on mechanical performances of poly‐L‐lactic acid cardiovascular stents},
journal = {Artificial Organs},
year = {2025},
doi = {10.1111/aor.14893}
}
EUVIMED