Kurz gefasst

Fachartikel in der Zeitschrift Artificial Organs, online erschienen am 27.08.2024. Verfasst von Benjamin J Schürmann und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, University Hospital RWTH Aachen University, Aachen, Germany. Hauptthemen laut MEDLINE: Hemolysis, Erythrocytes. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deutsche Forschungsgemeinschaft.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Hemolysis in mechanical circulatory support systems is currently determined quantitatively. To also locally resolve hemolysis, we are developing a fluorescent hemolysis detection method. This requires a translucent two‐phase blood analog fluid combined with particle image velocimetry, an optical flow field measurement. The blood analog fluid is composed of red blood cell surrogates. However, producing surrogates in sufficient volume is a challenge. We therefore present a high‐volume and high‐concentration production for our surrogates: ghost cells, hemoglobin‐depleted erythrocytes.

Methods

In the ghost cell production, the hemoglobin is removed by a repeated controlled osmolar lysis. We have varied the solution mixture, centrifugation time, and centrifugation force in order to increase production efficiency. The production is characterized by measurements of output volume, hematocrit, transparency, and rheology of the blood analog fluid.

Results

The volume of produced ghost cells was significantly increased, and reproducibility was improved. An average production of 389 mL of ghost cells were achieved per day. Those ghost cells diluted in plasma have a rheology similar to blood while being permeable to light.

Conclusion

The volume of ghost cells produced is sufficient for optical measurements as particle image velocimetry in mechanical circulatory support systems. This makes further work on experimental measurements for a locally resolved hemolysis detection possible.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HemolysisErythrocytesHumansRheologyHematocritHemoglobinsCentrifugationHeart-Assist Devices

Stichwörter der Autor:innen

fluorescent hemolysis detectionlocally resolved thrombosis detectionparticle image velocimetryresealed ghost cellstranslucent two‐phase blood analog fluid

Substanzen

Hemoglobins

Autor:innen und Institutionen

  • Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, University Hospital RWTH Aachen University, Aachen, Germany.
  • Institute of Applied Medical Engineering, University Hospital RWTH Aachen University, Aachen, Germany.

ORCID-Profile: Benjamin J Schürmann, Pia Creutz, Thomas Schmitz-Rode, Ulrich Steinseifer, Johanna C Clauser

Förderung

  • Deutsche Forschungsgemeinschaft

Zitieren

APA

Schürmann, B. J., Creutz, P., Schmitz-Rode, T., Steinseifer, U., & Clauser, J. C. (2025). Ghost cells as a two‐phase blood analog fluid: high‐volume and high‐concentration production. Artificial Organs, 49(1), 31-38. https://doi.org/10.1111/aor.14846

Vancouver

Schürmann BJ, Creutz P, Schmitz-Rode T, Steinseifer U, Clauser JC. Ghost cells as a two‐phase blood analog fluid: high‐volume and high‐concentration production. Artificial Organs. 2025;49(1):31-38. doi:10.1111/aor.14846

BibTeX
@article{Schrmann2025,
  author = {Schürmann, Benjamin J and Creutz, Pia and Schmitz-Rode, Thomas and Steinseifer, Ulrich and Clauser, Johanna C},
  title = {Ghost cells as a two‐phase blood analog fluid: high‐volume and high‐concentration production},
  journal = {Artificial Organs},
  year = {2025},
  volume = {49},
  number = {1},
  pages = {31--38},
  doi = {10.1111/aor.14846}
}