In vitro study on device‐induced damage to blood cellular components and degradation of von Willebrand factor in a CentriMag pump‐assisted circulation
Artificial Organs · 2024
DOI: 10.1111/aor.14766 ↗Kurz gefasst
Fachartikel in der Zeitschrift Artificial Organs, 2024. Verfasst von Shigang Wang und 5 weiteren Autor:innen. Beteiligte Einrichtung: Artificial Organ Lab, Department of Surgery, University of Maryland School of Medicine, Baltimore, Maryland, USA. Hauptthemen laut MEDLINE: von Willebrand Factor, Stress, Mechanical, Heart-Assist Devices. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NHLBI NIH HHS, National Heart, Lung, and Blood Institute.
Abstract (Originalsprache)
Background
High mechanical shear stress (HMSS) generated by blood pumps during mechanical circulatory support induces blood damage (or function alteration) not only of blood cell components but also of plasma proteins.
Methods
In the present study, fresh, healthy human blood was used to prime a blood circuit assisted by a CentriMag centrifugal pump at a flow rate of 4.5 L/min under three pump pressure heads (75, 150, and 350 mm Hg) for 4 h. Blood samples were collected for analyses of plasma‐free hemoglobin (PFH), von Willebrand factor (VWF) degradation and platelet glycoprotein (GP) IIb/IIIa receptor shedding.
Results
The extent of all investigated aspects of blood damage increased with increasing cross‐pump pressure and duration. Loss of high‐molecular‐weight multimers (HMWM)‐VWF in Loop 2 and Loop 3 significantly increased after 2 h. PFH, loss of HMWM‐VWF, and platelet GPIIb/IIIa receptor shedding showed a good linear correlation with mean shear stress corresponding to the three pump pressure heads.
Conclusions
HMSS could damage red blood cells, cause pathological VWF degradation, and induce platelet activation and platelet receptor shedding. Different blood components can be damaged to different degrees by HMSS; VWF and VWF‐enhanced platelet activation may be more susceptible to HMSS.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Artificial Organ Lab, Department of Surgery, University of Maryland School of Medicine, Baltimore, Maryland, USA.
- Fischell Department of Bioengineering, A. James Clark School of Engineering, University of Maryland, College Park, Maryland, USA.
ORCID-Profile: Zhongjun J Wu
Förderung
- NHLBI NIH HHS
- National Heart, Lung, and Blood Institute
Zitieren
APA
Wang, S., Sun, W., Han, D., Clark, K. P., Griffith, B. P., & Wu, Z. J. (2024). In vitro study on device‐induced damage to blood cellular components and degradation of von Willebrand factor in a CentriMag pump‐assisted circulation. Artificial Organs. https://doi.org/10.1111/aor.14766
Vancouver
Wang S, Sun W, Han D, Clark KP, Griffith BP, Wu ZJ. In vitro study on device‐induced damage to blood cellular components and degradation of von Willebrand factor in a CentriMag pump‐assisted circulation. Artificial Organs. 2024. doi:10.1111/aor.14766
BibTeX
@article{Wang2024,
author = {Wang, Shigang and Sun, Wenji and Han, Dong and Clark, Kiersten P and Griffith, Bartley P and Wu, Zhongjun J},
title = {In vitro study on device‐induced damage to blood cellular components and degradation of von Willebrand factor in a CentriMag pump‐assisted circulation},
journal = {Artificial Organs},
year = {2024},
doi = {10.1111/aor.14766}
}
EUVIMED