Organ blood flow assessment with the ModulHeart cardiorenal support device
Artificial Organs · 2024-01 · 48 (1) · 61-69
DOI: 10.1111/aor.14655 ↗Kurz gefasst
Fachartikel in der Zeitschrift Artificial Organs, online erschienen am 03.10.2023. Verfasst von Gabriel Georges und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of cardiac surgery, Quebec Heart and Lung Institute, Quebec, Quebec, Canada. Hauptthemen laut MEDLINE: Heart Failure, Heart-Assist Devices. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 26 Literaturangaben. Gefördert von: Puzzle Medical Devices Inc..
Abstract (Originalsprache)
Background
ModulHeart (Puzzle Medical Devices Inc) is a novel percutaneous flow entrainment pump anchored in the descending aorta. The current study evaluates the hemodynamic effect of ModulHeart support and its impact on cerebral, myocardial, and renal blood flow.
Methods
ModulHeart was implanted in the descending aorta of four healthy calves. A ramp protocol (2000 RPM increments) was performed with the pump operating at five different speeds from 14 000 to 22 000 RPM. For each speed, pressures proximal and distal to the pump, and right heart catheterization measurements were recorded. Stable‐isotope labeled microspheres were injected in the left ventricle to evaluate organ perfusion.
Results
Thermodilution cardiac output increased by 23% at 22 000 RPM. Greater pump speeds resulted in greater pump gradients, up to 10 mm Hg in mean arterial pressure at 22 000 RPM, without significant reduction of proximal perfusion pressures. Arterial pulse pressure remained stable at all speeds. ModulHeart was not associated with a reduction in cerebral or myocardial blood flow at any speed. Renal cortical and medullary blood flow increased by up to 50% and 40%, respectively.
Conclusion
The ModulHeart device implanted in the descending aorta of healthy calves resulted in significant arterial pressure gradients and preserved pulse pressure. Greater pump speeds translated into greater increases in renal blood flow, with no decrease in cerebral or myocardial perfusion.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of cardiac surgery, Quebec Heart and Lung Institute, Quebec, Quebec, Canada.
- Puzzle Medical Devices Inc., Montreal, Quebec, Canada.
- Department of cardiology, Centre Hospitalier de l'Université de Montréal, Montreal, Quebec, Canada.
- Cardiovascular Research Foundation, New York, New York, USA.
- Department of cardiology, Gagnon Cardiovascular Institute, Morristown Medical Center, Morristown, New Jersey, USA.
ORCID-Profile: Philippe Généreux
Förderung
- Puzzle Medical Devices Inc.
Zitieren
APA
Georges, G., Trudeau, F., Potvin, J., Burkhoff, D., & Généreux, P. (2024). Organ blood flow assessment with the ModulHeart cardiorenal support device. Artificial Organs, 48(1), 61-69. https://doi.org/10.1111/aor.14655
Vancouver
Georges G, Trudeau F, Potvin J, Burkhoff D, Généreux P. Organ blood flow assessment with the ModulHeart cardiorenal support device. Artificial Organs. 2024;48(1):61-69. doi:10.1111/aor.14655
BibTeX
@article{Georges2024,
author = {Georges, Gabriel and Trudeau, François and Potvin, Jeannot and Burkhoff, Daniel and Généreux, Philippe},
title = {Organ blood flow assessment with the ModulHeart cardiorenal support device},
journal = {Artificial Organs},
year = {2024},
volume = {48},
number = {1},
pages = {61--69},
doi = {10.1111/aor.14655}
}
EUVIMED