Effects of adding the second drainage cannula in severely hypoxemic patients supported with VV ECMO due to COVID ‐19‐associated ARDS
Artificial Organs · 2023
DOI: 10.1111/aor.14591 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift Artificial Organs, 2023. Verfasst von Konstanty Szułdrzyński und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anaesthesiology and Intensive Care, National Medical Institute of the Ministry of Interior and Administration, Warsaw, Poland. Hauptthemen laut MEDLINE: COVID-19, Extracorporeal Membrane Oxygenation, Respiratory Distress Syndrome. Die Arbeit nennt 37 Literaturangaben.
Abstract (Originalsprache)
Background
Extracorporeal membrane oxygenation (ECMO) is a recognized method of support in patients with severe and refractory acute respiratory distress syndrome (ARDS) caused by SARS‐CoV‐2 infection. While veno‐venous (VV) ECMO is the most common type, some patients with severe hypoxemia may require modifications to the ECMO circuit. In this study, we aimed to investigate the effects of adding a second drainage cannula to the circuit in patients with refractory hypoxemia, on their gas exchange, mechanical ventilation, ECMO settings, and clinical outcomes.
Methods
We conducted an observational retrospective study based on a single‐center institutional registry including all consecutive cases of COVID‐19 patients requiring ECMO admitted to the Centre of Extracorporeal Therapies in Warsaw between March 1, 2020 and March 1, 2022. We selected patients who had an additional drainage cannula inserted. Changes in ECMO and ventilator settings, blood oxygenation, and hemodynamic parameters, as well as clinical outcomes were assessed.
Results
Of 138 VV ECMO patients, 12 (9%) patients met the inclusion criteria. Ten patients (83%) were men, and mean age was 42.2 ± 6.8. An addition of drainage cannula resulted in a significant raise in ECMO blood flow (4.77 ± 0.44 to 5.94 ± 0.81 [L/min]; p = 0.001), and the ratio of ECMO blood flow to ECMO pump rotations per minute (RPM), whereas the raise in ECMO RPM alone was not statistically significant (3432 ± 258 to 3673 ± 340 [1/min]; p = 0.064). We observed a significant drop in ventilator FiO 2 and a raise in PaO 2 to FiO 2 ratio, while blood lactates did not change significantly. Nine patients died in hospital, one was referred to lung transplantation center, two were discharged uneventfully.
Conclusions
The use of an additional drainage cannula in severe ARDS associated with COVID‐19 allows for an increased ECMO blood flow and improved oxygenation. However, we observed no further improvement in lung‐protective ventilation and poor survival.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Anaesthesiology and Intensive Care, National Medical Institute of the Ministry of Interior and Administration, Warsaw, Poland.
- Thoracic Research Centre, Collegium Medicum Nicolaus Copernicus University, Innovative Medical Forum, Bydgoszcz, Poland.
- Department of Cardiac Surgery and Transplantology, National Medical Institute of the Ministry of Interior and Administration, Warsaw, Poland.
- Cardio-Thoracic Surgery Department, Heart and Vascular Centre, Maastricht University Medical Centre, Cardiovascular Research Institute Maastricht (CARIM), Maastricht, The Netherlands.
- Department of Anesthesia and Intensive Care, Istituto Mediterraneo per i trapianti e Terapie ad alta specializzazione (IRCCS-ISMETT), Palermo, Italy.
- National Medical Institute of the Ministry of the Interior and Administration, Warsaw, Poland.
ORCID-Profile: Konstanty Szułdrzyński, Mariusz Kowalewski, Miłosz Jankowski, Kamil Chudziński, Gennaro Martucci, Roberto Lorusso, Piotr Suwalski
Zitieren
APA
Szułdrzyński, K., Kowalewski, M., Jankowski, M., Staromłyński, J., Prokop, J., Pasierski, M., Chudziński, K., Drobiński, D., Martucci, G., Lorusso, R., Wierzba, W., Zaczyński, A., Król, Z., & Suwalski, P. (2023). Effects of adding the second drainage cannula in severely hypoxemic patients supported with VV ECMO due to COVID ‐19‐associated ARDS. Artificial Organs. https://doi.org/10.1111/aor.14591
Vancouver
Szułdrzyński K, Kowalewski M, Jankowski M, Staromłyński J, Prokop J, Pasierski M, et al. Effects of adding the second drainage cannula in severely hypoxemic patients supported with VV ECMO due to COVID ‐19‐associated ARDS. Artificial Organs. 2023. doi:10.1111/aor.14591
BibTeX
@article{Szudrzyski2023,
author = {Szułdrzyński, Konstanty and Kowalewski, Mariusz and Jankowski, Miłosz and Staromłyński, Jakub and Prokop, Joanna and Pasierski, Michał and Chudziński, Kamil and Drobiński, Dominik and Martucci, Gennaro and Lorusso, Roberto and Wierzba, Waldemar and Zaczyński, Artur and Król, Zbigniew and Suwalski, Piotr},
title = {Effects of adding the second drainage cannula in severely hypoxemic patients supported with VV ECMO due to COVID ‐19‐associated ARDS},
journal = {Artificial Organs},
year = {2023},
doi = {10.1111/aor.14591}
}
EUVIMED