EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Comparing left ventricular assist device inflow cannula angle between median sternotomy and thoracotomy using 3D reconstructions

Madeleine Pearman iD, Sam Emmanuel iD, Paul Jansz, Alasdair Watson, Mark Connellan, Arjun Iyer, Sumita Barua, Christopher Simon Hayward iD

Artificial Organs · 2023

DOI: 10.1111/aor.14492 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Artificial Organs, 2023. Verfasst von Madeleine Pearman und 7 weiteren Autor:innen. Beteiligte Einrichtung: St Vincent's Hospital, Sydney, Darlinghurst, New South Wales, Australia. Hauptthemen laut MEDLINE: Heart-Assist Devices, Heart Failure. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Left ventricular assist device (LVAD) implantation via thoracotomy has many potential advantages compared to conventional sternotomy, including improved inflow cannula (IFC) positioning. We compared the difference in IFC angles, postoperative, and long‐term outcomes for patients with LVADs implanted via thoracotomy and sternotomy.

Methods

A single‐center, retrospective analysis of 14 patients who underwent thoracotomy implantation was performed and matched with 28 patients who underwent sternotomy LVAD implantations for a total of 42 patients. Inclusion required a minimum LVAD support duration of 30 days and excluded concomitant procedures. A postoperative CT‐chest was used to measure the angle the between the IFC and mitral valve in two‐dimensions and results were compared with three‐dimensional reconstruction using the same CT chest. Outcome data were extracted from medical records.

Results

There was no significant difference in gender, INTERMACS score, BMI, or age between the two groups. Median cardiopulmonary bypass time was longer in the thoracotomy group compared to the sternotomy group, 107 min (86–122) versus 76 min (56–93), p < 0.01. 3D reconstructions revealed less deviation of the IFC away from the mitral valve in devices implanted via thoracotomy compared to sternotomy, median (IQR) angle 16.3° (13.9°–21.0°) versus 23.2° (17.9°–26.4°), p < 0.01. Rates of pump thrombosis, stroke, and gastrointestinal bleeding were not significantly different.

Conclusions

Devices implanted via thoracotomy demonstrated less deviation away from mitral valve. However, there was no difference in morbidity between the two approaches. 3D reconstruction of the heart is an innovative technique to measure angulation and is clinically advantageous when compared to 2D imaging.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart-Assist DevicesHeart FailureHumansSternotomyThoracotomyCannulaRetrospective StudiesImaging, Three-Dimensional

Stichwörter der Autor:innen

device angulation continuous flow pumpsdevice positioningthrombosisventricular assist devices

Autor:innen und Institutionen

  • St Vincent's Hospital, Sydney, Darlinghurst, New South Wales, Australia.
  • School of Medicine, University of Notre Dame, Sydney, Chippendale, New South Wales, Australia.
  • School of Medicine, University of New South Wales, Sydney, Kensington, New South Wales, Australia.
  • Victor Chang Cardiac Research Institute, Sydney, New South Wales, Australia.

ORCID-Profile: Madeleine Pearman, Sam Emmanuel, Christopher Simon Hayward

Zitieren

APA

Pearman, M., Emmanuel, S., Jansz, P., Watson, A., Connellan, M., Iyer, A., Barua, S., & Hayward, C. S. (2023). Comparing left ventricular assist device inflow cannula angle between median sternotomy and thoracotomy using 3D reconstructions. Artificial Organs. https://doi.org/10.1111/aor.14492

Vancouver

Pearman M, Emmanuel S, Jansz P, Watson A, Connellan M, Iyer A, et al. Comparing left ventricular assist device inflow cannula angle between median sternotomy and thoracotomy using 3D reconstructions. Artificial Organs. 2023. doi:10.1111/aor.14492

BibTeX
@article{Pearman2023,
  author = {Pearman, Madeleine and Emmanuel, Sam and Jansz, Paul and Watson, Alasdair and Connellan, Mark and Iyer, Arjun and Barua, Sumita and Hayward, Christopher Simon},